Implantable collamer lens versus small incision lenticule extraction for high myopia correction: A systematic review and meta-analysis

Abstract

Purpose

To compare the efficacy, safety, predictability and visual quality between implantable collamer lens (ICL) implantation and small incision lenticule extraction (SMILE) for high myopia correction in adults.

Methods

A systematic review and meta-analysis was conducted. A comprehensive literature search was done based on databases including PubMed, Science Direct, Embase, and the Cochrane Central Register of Controlled Trials. The efficacy index, safety index, changes in Snellen lines of corrected distance visual acuity (CDVA), predictability (difference between post-operative and attempted spherical equivalent error, SER), incidence of halos, and change in higher-order aberrations (HOAs) were compared. Mean difference (MD) and 95% confidence interval (CI) was used to estimate continuous outcomes, risk ratio (RR) and 95%CI was used to estimate categorical outcomes.

Results

Five observational studies involving 555 eyes were included in this review. Studies’ sample sizes (eyes) ranged from 76 to 197. Subjects’ refraction ranged from -6 diopter (D) to -12D. Study duration of most researches were 6 months or 12 months. Compared to SMILE, ICL implantation showed better efficacy index (MD=0.09, 95%CI:0.01 to 0.16) and better safety index (MD=0.08, 95%CI: 0.00 to 0.16). Compared with SMILE, more ICL-treated eyes gained one or more Snellen lines of CDVA (RR=1.54, 95%CI:1.28 to 1.86), more gained two or more lines (RR=2.09, 95%CI:1.40 to 3.13), less lost one or more lines (RR=0.17, 95%CI:0.05 to 0.63). There was no difference in predictability between two treatments, RRs of predictability of within ±0.5D and ±1D were 1.13 (95%CI: 0.94 to 1.36) and 1.00 (95%CI: 0.98 to 1.02). Compared with SMILE, ICL implantation came with a higher risk of halos [RR=1.79, 95%CI: 1.48 to 2.16] and less increase in total HOAs (MD=-0.23, 95%CI: -0.42 to -0.03).

Conclusion

Compared with SMILE, ICL implantation showed a higher risk of halos, but equal performance on SER control, and better performance on efficacy index, safety index, CDVA improvement and HOAs control. Overall, ICL implantation might be a better choice for high myopia correction in adults.

Peer Review reports

Introduction

Nowadays, it is well accepted to use small incision lenticule extraction (SMILE) to correct low-to-moderate myopia [1, 2]. For high myopia correction, SMILE also brought satisfying prognosis [3, 4], however, SMILE has its inherent limitations such as thick ablation depths, limited ablation zones and increased aberrations, SMILE could increase the risk of dry eye [5], myopic regression [6], haze [7] and corneal ectasia [8].

Implantable collamer lens (ICL) implantation was another promising way for high myopia correction [9], the postoperative spherical equivalent error (SER) was reported to be predictable [10, 11] and stable [12, 13]. ICL implantation broadened the scope of target population, such as patients with thin cornea [14] and even keratoconus [15]. A few studies compared ICL implantation with SMILE for high myopia correction, conclusions from different studies were controversial. Siedlecki J [16] and coworkers reported ICL implantation yielded better uncorrected distance visual acuity, better refractive accuracy, and fewer higher-order aberrations (HOAs) than SMILE. Moshirfar M [17] and coworkers thought SMILE might be comparable to ICL for high myopia correction, similarly, Wei R [18] reported both treatments showed compared performance for high myopia correction.

In this review, we aim to make a strengthened comparison between ICL implantation and SMILE for high myopia correction in adults. The efficacy index, safety index, changes in Snellen lines of corrected distance visual acuity (CDVA), predictability, incidence of halos, and change in higher-order aberrations (HOAs) would be compared between two treatments.

Methods and Materials

Inclusion criteria

Studies were included under the following consideration: 1) Subjects with high myopia (SER should be equal or greater than -6 diopter, D); 2) Subjects’ age≥18 years old; 3) The intervention measures must include ICL implantation and SMILE.

Exclusion criteria

Studies were excluded in any of the following condition: 1) Studies of case report, letter, comment or review; 2) Studies used only ICL implantation or only SMILE; 3) Studies that included low to moderate myopic patients (SER of -0.5D to -6D) or other kinds of patients.

Databases and Search strategy

We searched PubMed, Science Direct, and the Cochrane Central Register of Controlled Trials (from inception to 15 August, 2021) for studies published in English, the detailed search strategies were shown in the end of the manuscript.

Outcomes

According to the scale of extracted data, the following outcomes were quantitatively assessed: The efficacy index, safety index, changes in Snellen lines of CDVA, predictability, incidence of halos, and change in HOAs. Besides, the following outcomes were qualitatively described: endothelial cell loss, complications including cataract and dry eye, visual quality including objective scatter index (OSI) and modulation transfer function cut-off frequency (MTFcut-off) value.

Data Extraction

The following information measured at last-follow up time of each study was extracted: First author, publication year, subjects’ mean/median age, sample size (number of eyes), study design, mean follow-up duration, subjects’ SER range and type of ICL. We extracted the data of outcomes for analysis using a pre-designed data form. Briefly, for categorical data, such as the number of halos, the number of events were extracted. For continuous data, such as the efficacy index, we extracted the mean value and standard deviation (SD).

Data synthesis and statistical analysis

Meta-analysis was performed using either a fixed-effects model or a random-effects model according to the heterogeneity across included studies. The heterogeneity was assessed by a Q-test and the I2 statistic. The I2 statistic describes the percentage of variability caused by heterogeneity rather than by chance. An I2 of ≤50% indicates a relatively small heterogeneity across studies, subsequently a fixed-effects model would be used, otherwise a random-effects model would be used [19]. We used the mean differences (MDs) and 95% confidence intervals (CIs) to make comparison of continuous outcomes between ICL implantation and SMILE, we used risk ratios (RRs) and 95% CIs to estimate categorical outcomes. We used Egger’s test to determine publication bias. The significance level was set to be 0.05, two-tailed. All statistical analysis was done using the open-source R program (Version 4.0.0).

Results

Paper selection

The paper selection process was shown in Figure 1. Initially, a total of 293 articles were identified from PubMed, Science direct and Cochrane Central Register of Controlled Trials. 62 duplicates were removed, then 181 publications were further excluded by title and abstract. 23 reviews, letters or case reports were excluded. 22 articles were further excluded due to the following reasons: study in Czech (n=1), studies where only ICL implantation was performed (n=2), studies where subjects were children (n=2), studies recruited subjects of low or moderate myopia, or emmetropia (n=16), studies of iris-fixated Artiflex lens (n=1). Finally, 5 studies [13, 16, 18, 20, 21](involving 555 eyes) were included in this meta-analysis.

Fig. 1
figure 1

Flow chart of paper selection

Characteristics of the included studies

The characteristics of the five included studies were shown in Table 1. All studies were observational studies, and were published between year 2019 to 2021. All subjects were high myopic adults (age≥18), the sample sizes (number of eyes) ranged from 76 to 197. Subjects’ refraction ranged from -6D to -12D. Three studies used EVO Visian ICL (Visian ICL V4c) and two studies used Visian ICL without knowing the model. The median follow-up duration of five studies ranged from three months to 60 months.

Table 1 Characteristics of included studies

Efficacy and safety index

All five studies evaluated the efficacy index (Figure 2). Overall, ICL implantation showed a statistically better efficacy index (MD=0.09, 95%CI:0.01 to 0.16) and a better safety index (MD=0.08, 95%CI: 0.00, 0.16) than SMILE. Random-effects model were used for meta-analysis due to a large heterogeneity across studies (I2>50%).

Fig. 2
figure 2

Forest plot of comparison on efficacy index and safety index

CDVA

Three studies [16, 18, 20] assessed change in Snellen lines of CDVA (Fig. 3), the heterogeneity across three studies was small (I2=50%), so a fixed-effects model was applied. Compared with SMILE-treated eyes, more ICL-treated eyes gained one or more lines of CDVA (RR=1.54, 95%CI:1.28 to 1.86), more ICL-treated eyes gained two or more lines of CDVA (RR=2.09, 95%CI:1.40 to 3.13), less ICL-treated eyes lost one or more lines of CDVA (RR=0.17, 95%CI: 0.05 to 0.63). No eyes lost two or more Snellen lines in each group.

Fig. 3
figure 3

Forest plot of comparison on gaining or loss of corrected distance visual acuity (A: gaining one or more lines, B: gaining two or more lines, C: losing one or more lines, D: losing two or more lines)

Predictability

Four studies (except for study of Qin Q [21]) reported predictability of both treatments, difference of post-operative SER and attempted SER was used to evaluate the predictability. The I2 of predictability of within ± 0.5D and within ± 1D were 80% and 0% respectively, thus a random-effects model and a fixed-effects model were used to conduct meta-analysis respectively. Forest plot (Fig. 4) showed no statistical difference in predictability of within ±0.5D between two treatments (RR=1.13, 95%CI: 0.94 to 1.36) , there was no statistical difference in predictability of within ±1D either (RR=1.00, 95%CI: 0.98 to 1.02).

Fig. 4
figure 4

Forest plot of comparison on predictability

Halos

Three studies [13, 16, 18] evaluated halos after treatment, ICL implantation showed statistically higher risk of halos than SMILE (Fig. 5), the RR was 1.79 (95%CI: 1.48 to 2.16), there was no heterogeneity across included studies (I2=0%).

Fig. 5
figure 5

Forest plot of comparison on halos

Increase of total HOAs

Two studies assessed change of total HOAs, the heterogeneity across two studies was large (I2=97%), thus a random-effects model was applied. ICL-treated eyes showed a smaller increase of total HOAs (Fig. 6) than SMILE-treated eyes (MD=-0.23, 95%CI: -0.42, -0.03).

Fig. 6
figure 6

Forest plot of comparison on total higher-order aberrations

Other outcomes

For other outcomes including cornea endothelial cell loss, complications including cataract and dry eye, visual quality including objective scatter index (OSI) and modulation transfer function cut-off frequency (MTFcut-off) value, no pooled MD or pooled RR was calculated due to limited data. Qin Q and coworkers reported no significant endothelial cell density loss in either ICL-treated eyes or SMILE-treated eyes. Niu LL [13] and coworkers reported no statistical difference between ICL implantation and SMILE in change of either OSI or MTFcut-off before and after surgery. None study reported occurrence of cataract or dry eye in each treatment group.

Publication bias test

Publication bias was checked using Egger’s test, no publication bias was found by Egger’s test (p>0.05).

Discussion

High myopia, predisposed by genetic and environmental factors [22–24], poses challenge to clinical treatment, clinicians have to balance the benefit and risk since no choice is perfect. In this meta-analysis, we compared two popular options (SMILE and ICL implantation) for high myopia correction to help doctors make better decision. Efficacy index, calculated from visual acuity, is the most important outcome reflecting the efficacy of treatments. By this review, ICL implantation showed a better efficacy index than SMILE, which indicated that ICL implantation might be more efficient than SMILE for high myopia correction. ICL implantation also showed a better performance on safety index than SMILE. Besides, ICL implantation tended to come with a lower risk of visual acuity loss than SMILE, only 1.09% (2/182) of ICL-treated eyes lost one or more lines of CDVA, while the percentage reached 7.46% (15/201) in SMILE-treated eyes.

The predictability on SER reflects the accuracy of treatments, both ICL implantation and SMILE should achieve good accuracy according to their design principles. In this meta-analysis, both treatments showed good and equal performance on predictability, 93.67% (207/221) ICL-treated eyes and 83.19% (198/238) SMILE-treated eyes achieved a predictability of within ±0.5D, 100% (221/221) and 99.58% (237/238) achieved a predictability of within ±1D. Many previous studies reported similar findings, usually more than 95% ICL-treated eyes were within ±1.00 D of the intended refraction [11, 25–27]. A large retrospective study included 722 SMILE-treated high myopic eyes, 98% were within ±1.00 D of the intended refraction [28].

This meta-analysis showed ICL implantation leads to a smaller change in total HOAs than SMILE, this is reasonable since theoretically ICL implantation does not cause damage to the physiological structure of the cornea itself, thus eyeballs are able to preserve the adjustment ability, ensuring that subjects could obtain a more ideal visual quality. On the contrary, SMILE causes damage to the corneal surface morphology, and thus leads to change of HOAs. Evidence from previous studies supported the theory, this meta-analysis showed a smaller increase in total HOAs in ICL-treated eyes compared with SMILE-treated eyes, many other studies reported similar findings, usually, there was no significant increase of HOAs ICL-treated eyes [29, 30], while for SMILE-treated high myopic eyes, significant increase of total HOAs were commonly reported [31–34].

Contrast sensitivity, OSI, and MTFcut-off, are also important reflections of visual quality, however, no quantitative conclusion was drawn in this meta-analysis because few studies assessed these parameters. None of the included study performed contrast sensitivity test, Igarashi A [35] and coworkers reported that ICL implantation improved the contrast sensitivity while SMILE decreased the contrast sensitivity, Shin JY [36] and coworkers reported that ICL implantation induced fewer ocular and corneal HOAs, which resulted in a better contrast sensitivity at mesopic levels. Current evidence [36] showed no difference on OSI between ICL implantation and SMILE. But in terms of MTFcut-off value, evidence was contradicted, Qin Q [21] and coworkers reported that the postoperative MTFcut-off value of ICL-treated eyes was higher than SMILE-treated eyes, while Niu LL [13] and coworker found no significant difference between two treatments. However, given that either study of Qin Q [21] or study of Niu LL [13] was observational studies, which meaning the preoperative MTFcut-off value of two treatments might not be balanced, thus assessing the change in MTFcut-off value and change in OSI might be more meaningful, Niu LL [13] and coworkers reported no significant change before and after surgery, either for MTFcut-off value or OSI.

Complications are also important reflections of treatments’ safety, endothelial cell density loss was thought to be a main backward of ICL implantation [37]. In this meta-analysis, Qin Q’s study [21] reported no significant decrease before and after surgery either in the ICL group or SMILE group, the current evidence is not enough to assess the safety of both treatments. Halos were reported to be the leading complication of ICL implantation, the prevalence rates ranged from 15.2% [38, 39] to 93.5% [16, 18]. By this meta-analysis, up to 60.81% (104/171) ICL-treated eyes and 35.56% (64/180) SMILE-treated eyes perceived halos, in study of Wei R [18], the halos rates in ICL group and SMILE group were even as high as 93.5% and 54.4%, the reason might that the follow-up time was short (six months), as halos were commonly seen in the early period after ICL and SMILE surgery. Although no cataract formation was reported in either ICL-treated eyes or SMILE-treated eyes in five studies, it doesn’t represent that cataract is not a concern since the follow-up duration was usually one year or less except for study of Siedlecki J [16] (about two-year follow-up), Guber I39 and coworkers reported the lens opacity rate could reach 40.9% in 133 high myopic eyes 10 years after ICL implantation, 18 eyes of which underwent phacoemulsification.

In conclusion, both ICL implantation and SMILE had satisfying and equal performance on refraction control. ICL implantation came with a higher risk of halos, but was better than SMILE on efficacy, safety, CDVA improvement and total HOAs control. Overall, for high myopia correction, ICL implantation might be a better choice than SMILE. However, the conclusion came from observational studies with relatively short-term follow-up, evidence from randomized controlled trials and long-term studies is still needed.

Search strategy

  1. 1.

    PubMed, Science direct, Embase

((posterior chamber phakic intraocular lens) OR piol OR (implantable collamer lens) OR ICL OR ticl OR V4 OR V4C OR STAAR) AND ((small-incision lenticule extraction) OR smile OR (cornea refractive surgery)) AND (myopia OR (refractive errors) OR refraction)

  1. 2.

    Cochrane Central Register of Controlled Trials

#1posterior chamber phakic intraocular lens

#2piol

#3implantable collamer lens

#4ICL

#5ticl

#6V4

#7V4C

#8STAAR

#9#1 or #2 or #3 or #4 or #5 or #6 or #7 or #8

#10small-incision lenticule extraction

#11smile

#12 cornea refractive surgery

#13#10 or #11 or #12

#14myopia

#15refractive errors

#16refraction

#17 #14 or #15 or #16

#31#9 AND #13 AND #17

Availability of data and materials

Data would be available upon reasonable request

References

  1. Dishler JG, Slade S, Seifert S, Schallhorn SC. Small-Incision Lenticule Extraction (SMILE) for the Correction of Myopia with Astigmatism: Outcomes of the United States Food and Drug Administration Premarket Approval Clinical Trial. Ophthalmology. 2020;127(8):1020–34.

    Article  Google Scholar 

  2. Randleman JB. Small Incision Lenticule Extraction (SMILE): What Now? What Next? Ophthalmology. 2020;127(8):1035–6.

    Article  Google Scholar 

  3. Yin Y, Lu Y, Xiang A, et al. Comparison of the optical quality after SMILE and FS-LASIK for high myopia by OQAS and iTrace analyzer: a one-year retrospective study. Bmc Ophthalmol. 2021;21(1):292.

    Article  Google Scholar 

  4. Qian Y, Chen X, Naidu RK, Zhou X. Comparison of efficacy and visual outcomes after SMILE and FS-LASIK for the correction of high myopia with the sum of myopia and astigmatism from -10.00 to -14.00 dioptres. Acta Ophthalmol. 2020;98(2):e161–72.

    Article  Google Scholar 

  5. Denoyer A, Landman E, Trinh L, et al. Dry eye disease after refractive surgery: comparative outcomes of small incision lenticule extraction versus LASIK. Ophthalmology. 2015;122(4):669–76.

    Article  Google Scholar 

  6. Wang S, Dai J, Chu R, Zhou X. Long-term outcome of epi-LASIK for high myopia. Eur J Ophthalmol. 2012;22(Suppl 7):S98–105.

    Article  Google Scholar 

  7. Rosman M, Alio JL, Ortiz D, Perez-Santonja JJ. Comparison of LASIK and photorefractive keratectomy for myopia from -10.00 to -18.00 diopters 10 years after surgery. J Refract Surg. 2010;26(3):168–76.

    Article  Google Scholar 

  8. Randleman JB, Russell B, Ward MA, et al. Risk factors and prognosis for corneal ectasia after LASIK. Ophthalmology. 2003;110(2):267–75.

    Article  Google Scholar 

  9. Sanders DR, Brown DC, Martin RG, et al. Implantable contact lens for moderate to high myopia: phase 1 FDA clinical study with 6 month follow-up. J Cataract Refract Surg. 1998;24(5):607–11.

    Article  CAS  Google Scholar 

  10. Kocova H, Vlkova E, Michalcova L, Motyka O. Implantation of posterior chamber phakic intraocular lens for myopia and hyperopia - long-term clinical outcomes. J Fr Ophtalmol. 2017;40(3):215–23.

    Article  CAS  Google Scholar 

  11. Yan Z, Miao H, Zhao F, et al. Two-Year Outcomes of Visian Implantable Collamer Lens with a Central Hole for Correcting High Myopia. J Ophthalmol. 2018;2018:8678352.

    PubMed  PubMed Central  Google Scholar 

  12. Kamiya K, Shimizu K, Igarashi A, Komatsu M. Comparison of Collamer toric implantable [corrected] contact lens implantation and wavefront-guided laser in situ keratomileusis for high myopic astigmatism. J Cataract Refract Surg. 2008;34(10):1687–93.

    Article  Google Scholar 

  13. Niu L, Miao H, Tian M, et al. One-year visual outcomes and optical quality of femtosecond laser small incision lenticule extraction and Visian Implantable Collamer Lens (ICL V4c) implantation for high myopia. Acta Ophthalmol. 2020.

  14. Bhikoo R, Rayner S, Gray T. Toric implantable collamer lens for patients with moderate to severe myopic astigmatism: 12-month follow-up. Clin Exp Ophthalmol. 2010;38(5):467–74.

    Article  Google Scholar 

  15. Abdelmassih Y, El-Khoury S, Chelala E, et al. Toric ICL Implantation After Sequential Intracorneal Ring Segments Implantation and Corneal Cross-linking in Keratoconus: 2-Year Follow-up. J Refract Surg. 2017;33(9):610–6.

    Article  Google Scholar 

  16. Siedlecki J, Schmelter V, Mayer WJ, et al. SMILE Versus Implantable Collamer Lens Implantation for High Myopia: A Matched Comparative Study. J Refract Surg. 2020;36(3):150–9.

    Article  Google Scholar 

  17. Moshirfar M, Somani AN, Motlagh MN, et al. Comparison of FDA-Reported Visual and Refractive Outcomes of the Toric ICL Lens, SMILE, and Topography-Guided LASIK for the Correction of Myopia and Myopic Astigmatism. J Refract Surg. 2019;35(11):699–706.

    Article  Google Scholar 

  18. Wei R, Li M, Zhang H, et al. Comparison of objective and subjective visual quality early after implantable collamer lens V4c (ICL V4c) and small incision lenticule extraction (SMILE) for high myopia correction. Acta Ophthalmol. 2020.

  19. Cao K, Friedman DS, Jin S, et al. Multifocal versus monofocal intraocular lenses for age-related cataract patients: a system review and meta-analysis based on randomized controlled trials. Surv Ophthalmol. 2019;64(5):647–58.

    Article  Google Scholar 

  20. Jiang Z, Wang H, Luo DQ, Chen J. Optical and visual quality comparison of implantable collamer lens and femtosecond laser assisted laser in situ keratomileusis for high myopia correction. Int J Ophthalmol. 2021;14(5):737–43.

    Article  Google Scholar 

  21. Qin Q, Bao L, Yang L, et al. Comparison of visual quality after EVO-ICL implantation and SMILE to select the appropriate surgical method for high myopia. Bmc Ophthalmol. 2019;19(1):21.

    Article  Google Scholar 

  22. Cai XB, Zheng YH, Chen DF, et al. Expanding the Phenotypic and Genotypic Landscape of Nonsyndromic High Myopia: A Cross-Sectional Study in 731 Chinese Patients. Invest Ophthalmol Vis Sci. 2019;60(12):4052–62.

    Article  CAS  Google Scholar 

  23. Cai XB, Shen SR, Chen DF, et al. An overview of myopia genetics. Exp Eye Res. 2019;188:107778.

    Article  CAS  Google Scholar 

  24. Jin ZB, Wu J, Huang XF, et al. Trio-based exome sequencing arrests de novo mutations in early-onset high myopia. Proc Natl Acad Sci U S A. 2017;114(16):4219–24.

    Article  CAS  Google Scholar 

  25. Li K, Wang Z, Zhang D, et al. Visual Outcomes and Corneal Biomechanics after Visian V4c Phakic Intraocular Lens Implantation in subclinical keratoconus. J Cataract Refract Surg. 2020.

  26. Kamiya K, Shimizu K, Kobashi H, et al. Clinical outcomes of posterior chamber toric phakic intraocular lens implantation for the correction of high myopic astigmatism in eyes with keratoconus: 6-month follow-up. Graefes Arch Clin Exp Ophthalmol. 2011;249(7):1073–80.

    Article  Google Scholar 

  27. Chaitanya SR, Anitha V, Ravindran M, et al. Safety and efficacy of toric implantable collamer lens V4c model - A retrospective South Indian study. Indian J Ophthalmol. 2020;68(12):3006–11.

    Article  Google Scholar 

  28. Hansen RS, Lyhne N, Grauslund J, Vestergaard AH. Small-incision lenticule extraction (SMILE): outcomes of 722 eyes treated for myopia and myopic astigmatism. Graefes Arch Clin Exp Ophthalmol. 2016;254(2):399–405.

    Article  Google Scholar 

  29. Wan T, Yin H, Wu Z, Yang Y. Comparative Study of Implantable Collamer Lens Implantation in Treating Four Degrees of Myopia: Six-Month Observation of Visual Results, Higher-Order Aberrations, and Amplitude of Accommodation. Curr Eye Res. 2019:1–8.

  30. Hashemian SJ, Farrokhi H, Foroutan A, et al. Ocular higher-order aberrations changes after implantable collamer lens implantation for high myopic astigmatism. J Curr Ophthalmol. 2018;30(2):136–41.

    Article  Google Scholar 

  31. Lee H, Yong KD, Reinstein DZ, et al. Comparing corneal higher-order aberrations in corneal wavefront-guided transepithelial photorefractive keratectomy versus small-incision lenticule extraction. J Cataract Refract Surg. 2018;44(6):725–33.

    Article  Google Scholar 

  32. Qin B, Li M, Chen X, et al. Early visual outcomes and optical quality after femtosecond laser small-incision lenticule extraction for myopia and myopic astigmatism correction of over -10 dioptres. Acta Ophthalmol. 2018;96(3):e341–6.

    Article  Google Scholar 

  33. Miraftab M, Hashemi H, Aghamirsalim M, et al. Matched comparison of corneal higher order aberrations induced by SMILE to femtosecond assisted LASIK and to PRK in correcting moderate and high myopia: 3.00mm vs. 6.00mm. Bmc Ophthalmol 2021;21(1):216.

  34. Xia F, Qin B, Shang J, et al. Four-Year Outcomes of Small Incision Lenticule Extraction for Extreme High Myopia and Myopic Astigmatism. Front Med (Lausanne). 2020;7:575779.

    Article  Google Scholar 

  35. Igarashi A, Kamiya K, Shimizu K, Komatsu M. Visual performance after implantable collamer lens implantation and wavefront-guided laser in situ keratomileusis for high myopia. Am J Ophthalmol. 2009;148(1):164–70.

    Article  Google Scholar 

  36. Shin JY, Ahn H, Seo KY, et al. Comparison of higher order aberrations after implantable Collamer Lens implantation and wavefront-guided LASEK in high myopia. J Refract Surg. 2012;28(2):106–11.

    Article  Google Scholar 

  37. Fuller DJ, Carper SW, Clay L, et al. Polyamine regulation of heat-shock-induced spermidine N1-acetyltransferase activity. Biochem J. 1990;267(3):601–5.

    Article  CAS  Google Scholar 

  38. Bamashmus MA, Al-Arabi AH, Alawad MA. Visual outcomes and patient satisfaction after implantable collamer lens and Toric implantable collamer lens correction for moderate to high myopia and myopic astigmatism. Saudi Med J. 2013;34(9):913–9.

    PubMed  Google Scholar 

  39. Guber I, Mouvet V, Bergin C, et al. Clinical Outcomes and Cataract Formation Rates in Eyes 10 Years After Posterior Phakic Lens Implantation for Myopia. Jama Ophthalmol. 2016;134(5):487–94.

    Article  Google Scholar 

Download references

Acknowledgements

None

Funding

None

Author information

Authors and Affiliations

Authors

Contributions

Concept (Xiu Hua Wan, Ningli Wang, Zi-Bing Jin), literature search (Kai Cao, Shanshan Jin), data extraction (Kai Cao, Shanshan Jin), data analysis (Kai Cao), manuscript writing (Kai Cao, Xiu Hua Wan, Ningli Wang, Zi-Bing Jin, Jingshang Zhang, Jinda Wang, Shuying Chen), English polishing (Mayinuer Yusufu).

Corresponding author

Correspondence to Xiu Hua Wan.

Ethics declarations

Ethics approval and consent to participate

Not applicable

Competing interests

There is no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cao, K., Zhang, J., Wang, J. et al. Implantable collamer lens versus small incision lenticule extraction for high myopia correction: A systematic review and meta-analysis. BMC Ophthalmol 21, 450 (2021). https://doi.org/10.1186/s12886-021-02206-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-021-02206-9

Keywords