Xen45 gel stent ab interno trimming for ostium occlusion: case report

Abstract

Background

Despite the XEN gel stent low-risk profile, various complications following the implant have been described, including internal and external occlusion, with a consequent postoperative rise in intraocular pressure (IOP). In this case report we aimed to present a XEN45 stent internal occlusion successfully treated by trimming in situ its proximal end with a 25 G vitreous scissors using a bimanual technique.

Case presentation

A 63-year-old male patient affected by primary open angle glaucoma (IOP = 25 mmHg) and a full thickness macular hole in his right eye, underwent ab-interno Xen gel stent implantation and, 1 month later, a 25 G vitrectomy surgery. Despite a significant IOP reduction after stent implant, 6 days after vitrectomy, IOP increased (25 mmHg) and the conjunctival bleb flattened following occlusion of stent internal ostium by a clot of presumed fibrinous material. The Nd:YAG laser failed to remove the clot, so that we decided to snip a small bit of the proximal end of the Xen tube (about 0,5 mm length) with a 25 G vitreous scissors, using a bimanual technique. In the postoperative day 1 and month 1, the IOP was 8 mmHg and 12 mmHg, respectively. The Anterior Segment OCT confirmed a functional, layered, filtering bleb, and the normal appearance and patency of the XEN proximal segment. No side effects from the intervention were observed.

Conclusions

Ab interno trimming with vitreous scissors of the occluded proximal end of the XEN stent may represent a safe, rapid and efficient method to restore aqueous humor subconjunctival drainage.

Peer Review reports

Background

Glaucoma is the leading cause of irreversible vision loss worldwide. In 2013, the number of affected people was estimated to be 64.3 million, increasing to 111.8 million in 2040 [1].

The only modifiable risk factor is the intraocular pressure (IOP); hence its lowering is considered the therapeutic approach for glaucoma [2]. This goal can be performed by medications, usually preferred as a first-line therapy, laser techniques and surgery [3]. Trabeculectomy remains the standard surgical procedure, although it may lead to significant complications [4]. Although it was reported a higher rate of complications in trabeculectomy versus the mini glaucoma device implantation, some device–related late complications due to shunt dislocation or extrusion have been described [5]. Minimally-invasive glaucoma surgeries (MIGS) have been developed to provide safer and less traumatic surgical interventions, aiming at curtailing the need for topical medications [6].

The Xen45 (Allergan Plc, Dublin, Ireland) is a tubular implant 6-mm-long and with an inner lumen of 45-μm, made of hydrophilic collagen, comprised of cross-linked porcine gelatin, which drains aqueous from the anterior chamber (AC) to the subconjunctival space (SS), bypassing the resistance of the dysfunctional trabecular meshwork [7]. The implant is injected ab interno through a clear corneal incision with a pre-loaded, single-use injector with a 27-gauge needle. The ideal stent placement should leave 2.0 mm of exposed implant in the subconjunctival space (preferentially in a more superficial layer than the sub-Tenon space), 1.0 mm in the AC, and 3.0 mm tunneled through sclera. The lumen size and length of the implant are designed to provide approximately 8 mm pressure resistance according to the Hagen-Poiseuille equation, so as to confer protection against hypotony [7].

Anterior segment optical coherence tomography (AS-OCT) enables tomographic imaging not only of the filtering bleb, but also the visualization of the implant itself, its subconjunctival and intrascleral pathway, and even the intracameral segment [8].

Several studies demonstrated the XEN gel stent efficacy in lowering IOP [9, 10].However, despite its low-risk profile, various complications following Xen implant have been described, including internal and external occlusion, with a consequent postoperative rise in IOP [9, 10].

There are several methods of unblocking the end of glaucoma stents occluded by fibrin clots such as Nd:YAG laser lysis [11],removal by ILM forceps [12], and Intracameral tissue plasminogen activator [13].Although the accidental or intentional trimming of the Xen stent external ostium has been previously described [14,15,16], the ab interno cutting of tube end has been described only in eyes with Baerveldt GDD implantation [17].

In this report we describe a case of Xen internal ostium obstruction by presumed fibrinous material, which occurred after vitreoretinal surgery, and its management.

Case presentation

History

A 63-year-old man was referred to our clinic for primary open glaucoma in both eyes, with an intraocular pressure of 25 mmHg in the right eye despite topic medications (Dorzolamide, Timolol, Travoprost). At the initial visit the anterior segment of the pseudophakic right eye was normal and the best corrected visual acuity was 20/200. The fundus examination showed peripapillary atrophy, glaucomatous neuroretinal rim loss (cup/disk ratio 0,7) and a full thickness macular hole in his right eye. The patient underwent an ab-interno gel stent implantation in the upper nasal quadrant (Xen45, Allergan) after a subconjunctival injection of mitomycin-C (0,1 ml 0,02%). After hydrating the incisions, 0.1 ml of 1% cefuroxime was injected in the anterior chamber and 4 mg/1 ml of dexamethasone phosphate subtenon (Decadron®, Farmaceutici Caber SpA, Italy). Post-surgery care included antibiotic prophylaxis and topical corticoids in decreasing dosage during 1 month. The IOP was 6 mmHg and 10 mmHg in post-operative day 1 and post-operative day 15, respectively. One month after the implantation, the patient underwent retinal surgery in his right eye. The macular hole was successfully closed with a 25-G pars plana vitrectomy (Constellation vitrectomy machine, Alcon Laboratories, Inc., Fort Worth, TX, USA), using the inverted ILM flap technique [18].

The surgery was uneventful and subconjunctival dexamethasone was used to minimize inflammation. Topical medication with combined antibiotic and steroid drops was prescribed for 1 week.

Six days after vitreous surgery the IOP spiked to 25 mmHg, and the anterior segment examination (Takagi TD10 Eye Cam on a Takagi 700GL Slit lamp) showed a flattening of the conjunctival bleb and a translucent clot covering the internal ostium of the XEN. (Fig. 1) At the AS-OCT (AngioVue®, Optovue, Fremont, CA, USA) the conjunctival bleb appeared flat and non-functional, and the clot stood out as hyperreflective material which we presumed to be fibrin, occluding the internal end of the stent (Fig. 2).

Fig. 1
figure 1

Biomicroscopy image showing: A The Xen45 anterior chamber portion (about 1 mm) obstructed by a translucent clot of fibrinous material surrounding the internal end (white circle); B A segment of the Xen45 is visible under a flat conjunctival bleb

Fig. 2
figure 2

The AS-OCT shows the fibrinous material surrounding the inner end and occluding the lumen of the Xen45; a non-filtering, non-functional, flat conjunctival bleb is also visible

Lysis of the fibrin clot was attempted with YAG-laser (1,1 mJ of power), but the tube remained occluded and the IOP elevated. Therefore, an ab interno revision of the gel stent was performed.

Surgical procedure

A surgical procedure was performed according to the following steps after obtaining written informed consent from the patient.

  • After application of topical anesthesia with the usual sterile conditions in the operating room, one paracentesis site was created at 90° with respect to the tube (approx. at 10 o’clock) by using a 15° Knife (Stab 15° Safety Knife, Surgistar, California)

  • Trypan blue (Vision Blue, DORC international, BV Zuidland, Netherlands) was injected into the anterior chamber to stain the clot and to verify the extension of tube occlusion.

  • Viscoelastic was introduced in the anterior chamber and a careful fibrin clot removal was attempted with a 25-gauge inner limiting membrane (ILM) forceps (Revolution DSP 25+ Serrated Forceps, Alcon-Grieshaber, Fribourg, Switzerland); however, the clot was strongly adherent to the tube end, and it did not clear.

  • A second paracentesis site was created at 90° with respect to the Xen tube position (approx.at 4 o’clock), and a 25 G straight vitreous scissors (Revolution DSP 25G Curved Scissors, Alcon-Grieshaber, Fribourg, Switzerland) was inserted to snip the proximal end of the Xen tube (a small segment of about 0,5 mm length), flush to the ILM forceps grabbing distally the tube. The flexible nature of the stent required a two-handed technique in order to both immobilize and truncate the end of the stent. (Fig. 3)

  • The viscoelastic and the excised tube fragment were then removed, and the AC was irrigated with balanced salt solution (BSS) to induce subconjunctival bleb formation. The drainage efficiency was further verified by trypan blue injection into the AC.

  • Finally, the corneal incisions were closed by hydrosuture.

Fig. 3
figure 3

Digital reproduction of the two-handed technique, with a 25-gauge inner limiting membrane (ILM) forceps (A) and a 25 G straight vitreous scissors (B), used in order to both immobilize and truncate the end of the stent

In the postoperative day 1, the IOP was 8 mmHg; the anterior segment examination showed an open diffuse filtering bleb, a normal appearing new internal ostium, and a sustained blue staining of the stent [19]. (Fig. 4) The AS-OCT confirmed a functional, layered, filtering bleb, and the normal appearance and patency of the stent. (Fig. 5) No side effects from the intervention were observed.

Fig. 4
figure 4

Anterior segment image acquired on postoperative day1 shows the XEN gel stent visible in the anterior chamber (A) and an open diffuse filtering bleb (B). The tube is stained with blue, confirming the open drainage pathway from the anterior chamber.

Fig. 5.
figure 5

AS-OCT image evidences the patency of the internal ostium of the tube (A) and a functional, layered, filtering conjunctival bleb (B)

Topical medication with combined antibiotic and steroid drops was prescribed for 2 weeks. One month after surgery the patient showed IOP =12 mmHg, a visual acuity of 20/40, a patent internal ostium, and an open diffuse filtering bleb.

Discussion and conclusions

Despite several studies have demonstrated the effectiveness and safety of the Xen45, a variety of postoperative complications have been reported [9, 10].In particular, early lumen occlusion can occur in 4% of Xen stent procedures for primary open-angle glaucoma (blood, fibrin, air bubble or iris) [20], or later due to continued inflammation [21].

Vitreoretinal surgery is considered a possible cause of AC fibrin formation [22]: in our patient a presumed fibrin clot formed at the internal end of the stent 1 week after macular hole surgery.

Different methods of unblocking the proximal end of glaucoma stents occluded by fibrin/blood clots have been described: Nd:YAG laser lysis [11], removal by ILM forceps [12], and Intracameral tissue plasminogen activator [13].

Despite the Nd:YAG laser was effective in a case of XEN occlusion, in our case it was unsuccessful. One-hand removal by ILM forceps was described to be successful in a case of blood clot internal occlusion [12], but was ineffective in our case, probably due to the high adhesiveness of the fibrin clot to the tube end. Intracameral tissue plasminogen activator has been used successfully in eyes with valved glaucoma drainage implants to clear or prevent tube occlusion by fibrin/blood clots, but its use has never been reported in eyes with Xen stent blockage.

Although unintentional Xen stent truncation of external end has been observed during needling procedure for in-growing fibrotic tissue [14,15,16], and intentional truncation has been performed with YAG laser in a case of threatened conjunctival extrusion [23], ab interno trimming of tube end has been described only in eyes with Baerveldt GDD implantation [24].

Indeed, doubts and concerns may arise from manipulation and surgical trimming, with consequent shortage, of the stent tube.

Undoubtedly, even cautious clamping of the internal end may retract the stent into the anterior chamber, and attention must be paid to limit its sliding. Shortage of the tube in the range of 1 mm seems to be well tolerated: In our patient, the IOP dropped to 8 mmHg 1 day after trimming, to stabilize at 12 mmHg after 1 month. Similarly, in reported cases of Xen external amputation of the subconjunctival portion, the shortage length ranged from 0.7 to 1,1 mm, and did not seem to impair the expected normal resistance (set at 8 mmHg) of the draining device [14,15,16].

As previously reported, the XEN 45 implant was designed to limit secondary hypotony by virtue of its length and width according to the Hagen-Poiseuille equation [7], Resistance to flow is directly proportional to the length and inversely proportional to the radius of the tube to the fourth power [24]. Thus, a shorter tube provides a smaller pressure differential, or less resistance to flow. Contrarily, Jansonius postulated that there is no physical basis for a relationship between tube length and intraocular pressure after glaucoma drainage implant surgery [25]. Considering this aspect was referred to glaucoma tubes, further studies are needed to evaluate the applicability of Hagen-Poiseuille law to Xen gel stent.

However, once the bleb has matured and subconjunctival fibrosis has occurred, the resultant IOP will ultimately be limited by subconjunctival resistance [24].

Further study will be needed to determine the long-term effects of shorter Xen45 tube on IOP control and wound healing response of the bleb.

In conclusion, to our knowledge, this is the first case reporting ab interno trimming of the proximal end of Xen45 stent occluded by fibrinous material. The post-operative course and IOP decrease were satisfactory, and the AS-OCT imaging confirmed the bleb had restored, and the tube shunt had completely cleared of obstruction. Our bimanual method appears to be minimally invasive, rapid and efficient to reestablish the drainage flow, and may be easily applied to similar cases, when other methods have failed.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AC:

Anterior Chamber

AS-OCT:

Anterior Segment – Optical Coherence Tomography

BSS:

Balanced Salt Solution

ILM:

Inner Limiting Membrane

IOP:

Intraocular Pressure

MIGS:

Minimally-invasive glaucoma surgeries

SS:

Subconjunctival Space

References

  1. Tham YC, Li X, Wong TY, Quigley HA, Aung T, Cheng CY. Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis. Ophthalmology. 2014;121:2081–90.

    Article  Google Scholar 

  2. European Glaucoma Society Terminology and Guidelines for Glaucoma. 4th edition—part 1 supported by the EGS foundation. Br J Ophthalmol. 2017;101:1–72.

    Google Scholar 

  3. European Glaucoma Society. Terminology and guidelines for Glaucoma: treatment principles and options (chapter 3). Br J Ophthalmol. 2017;101:130–95.

    Article  Google Scholar 

  4. Koike KJ, Chang PT. Trabeculectomy: a brief history and review of current trends. Int Ophthalmol Clin. 2018;58:117–33.

    Article  Google Scholar 

  5. Tognetto D, Cecchini P, D’Aloisio R, Vattovani O, Turco G. Scanning electron microscopy evaluation of an EX-PRESS mini glaucoma shunt after explantation. J Glaucoma. 2017;26:e1–4.

    Article  Google Scholar 

  6. Lavia C, Dallorto L, Maule M, Ceccarelli M, Fea AM. Minimally-invasive glaucoma surgeries (MIGS) for open angle glaucoma: a systematic review and meta-analysis. PLoS One. 2017. https://doi.org/10.1371/journal.pone.0183142.

  7. Lewis RA. Ab interno approach to the subconjunctival space using a collagen glaucoma stent. J Cataract Refract Surg. 2014;40:1301–6.

    Article  Google Scholar 

  8. Rigo J, Castany M, Banderas S, Pujol O, Amilburu M, Dou A. Possible intraluminal obstruction of the XEN45 gel stent observed with anterior segment optical coherence tomography. J Glaucoma. 2019;28:1095–101.

    Article  Google Scholar 

  9. Chatzara A, Chronopoulou I, Theodossiadis G, Theodossiadis P, Chatziralli I. XEN implant for Glaucoma treatment: a review of the literature. Semin Ophthalmol. 2019;34:93–7.

    Article  Google Scholar 

  10. Fea AM, Durr GM, Marolo P, Malinverni L, Economou MA, Ahmed I. Xen® gel stent: a comprehensive review on its use as a treatment option for refractory glaucoma. Clin Ophthalmol. 2020;14:1805–32.

    Article  CAS  Google Scholar 

  11. Scantling-Birch Y, Merzougui W, Lindfield D. Early postoperative lumen blockage of ab-interno gel stent (XEN) cleared with Nd:YAG laser. Indian J Ophthalmol. 2020;68:524.

    Article  Google Scholar 

  12. Ferreira NP, Pinto LA, Marques-Neves C. XEN gel stent internal Ostium occlusion: Ab-Interno revision. J Glaucoma. 2017;26:e150–2.

    Article  Google Scholar 

  13. Zalta AH, Sweeney CP, Zalta AK, Kaufman AH. Intracameral tissue plasminogen activator use in a large series of eyes with valved glaucoma drainage implants. Arch Ophthalmol. 2002;120:1487–93.

    Article  Google Scholar 

  14. Olivari S, Cutolo CA, Negri L, et al. XEN implant fracture during needling procedure. J Glaucoma. 2019;28:1086–9.

    Article  Google Scholar 

  15. Bustros Y, Chaudhary A, Salinas L, Mansouri K. Cutting the subconjunctival fragment of the XEN gel implant during needling procedure. Eur J Ophthalmol. 2020;30:NP11–5.

    Article  Google Scholar 

  16. Linton E, Au L. Technique of Xen implant revision surgery and the surgical outcomes: a retrospective interventional case series. Ophthalmol Ther. 2020;9:149–57.

    Article  Google Scholar 

  17. Baciu P, Moster SJ, Lee RK. Rapid intraocular tube trimming for Glaucoma drainage implant devices. J Glaucoma. 2017;26:e174–6.

    Article  Google Scholar 

  18. Shroff D, Gupt P, Atri N, Gupta C, Shroff C. Inverted internal limiting membrane (ILM) flap technique for macular hole closure: patient selection and special considerations. Clin Ophthalmol. 2019;13:671–8.

    Article  Google Scholar 

  19. Ling J, Bell NP, Feldman RM. Sustainable staining of the Xen gel stent. Investig Ophthalmol Vis Sci. 2018;59:2015.

    Article  Google Scholar 

  20. Karimi A, Lindfield D, Turnbull A, et al. A multi-Centre interventional case series of 259 ab-interno Xen gel implants for glaucoma, with and without combined cataract surgery. Eye. 2019;33:469–77.

    Article  Google Scholar 

  21. Gillmann K, Mansouri K, Bravetti GE, Mermoud A. Chronic intraocular inflammation as a risk factor for XEN gel stent occlusion: a case of microscopic examination of a fibrin-obstructed XEN stent. J Glaucoma. 2018;27:739–41.

    Article  Google Scholar 

  22. Cunha LP, Costa-Cunha LVF, Costa CF, Bastos RR, dos Prata TS, Monteiro MLR. Predictive factors for anterior chamber fibrin formation after vitreoretinal surgery. Rev Bras Oftalmol. 2014;73:93–7.

    Article  Google Scholar 

  23. Lee GA, Shah P. YAG laser Xen stent truncation. Ophthalmol Glaucoma. 2020;3:359.

    Article  Google Scholar 

  24. Sheybani A, Reitsamer H, Ahmed IIK. Fluid dynamics of a novel micro-fistula implant for the surgical treatment of glaucoma. Investig Ophthalmol Vis Sci. 2015;56:4789–95.

    Article  CAS  Google Scholar 

  25. Jansonius NM. A relationship between tube length and intraocular pressure after glaucoma drainage implant surgery cannot be explained by Poiseuille’s law. Acta Ophthalmol. 2014;92:2014.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

Author information

Authors and Affiliations

Authors

Contributions

FT, MF wrote the manuscript. PG, LM, PN, GD revised the manuscript. All authors have read and approved the manuscript.

Corresponding author

Correspondence to Maurizio Fossarello.

Ethics declarations

Ethics approval and consent to participate

Since only normal clinical practice was described in this case report, formal ethical approval by the Independent Review Board was not required in accordance with the policy of our Institution.

Consent for publication

Written consent for the publication of this case report, any related information and images was obtained from the patient.

Competing interests

The Authors have no financial or proprietary interest in products, methods, or material used in this case report.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tatti, F., Gentile, P., Mangoni, L. et al. Xen45 gel stent ab interno trimming for ostium occlusion: case report. BMC Ophthalmol 21, 446 (2021). https://doi.org/10.1186/s12886-021-02207-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-021-02207-8

Keywords