Development of an implantable collamer lens sizing model: a retrospective study using ANTERION swept-source optical coherence tomography and a literature review

Abstract

Background

Optimal sizing for phakic intraocular lens (EVO-ICL with KS-AquaPort) implantation plays an important role in preventing postoperative complications. We aimed to formulate optimal lens sizing using ocular biometric parameters measured with a Heidelberg anterior segment optical coherence tomography (AS-OCT) device.

Methods

We retrospectively analyzed 892 eyes of 471 healthy subjects treated with an intraocular collamer lens (ICL) and assigned them to either the development (80%) or validation (20%) set. We built vault prediction models using the development set via classic linear regression methods as well as partial least squares and least absolute shrinkage and selection operator (LASSO) regression techniques. We evaluated prediction abilities based on the Bayesian information criterion (BIC) to select the best prediction model. The performance was measured using Pearson’s correlation coefficient and the mean squared error (MAE) between the achieved and predicted results.

Results

Measurements of aqueous depth (AQD), anterior chamber volume, anterior chamber angle (ACA) distance, spur-to-spur distance, crystalline lens thickness (LT), and white-to-white distance from ANTERION were highly associated with the ICL vault. The LASSO model using the AQD, ACA distance, and LT showed the best BIC results for postoperative ICL vault prediction. In the validation dataset, the LASSO model showed the strongest correlation (r = 0.582, P < 0.001) and the lowest MAE (104.7 μm).

Conclusion

This is the first study to develop a postoperative ICL vault prediction and lens-sizing model based on the ANTERION. As the measurements from ANTERION and other AS-OCT devices are not interchangeable, ANTERION may be used for optimal ICL sizing using our formula. Because our model was developed based on the East Asian population, further studies are needed to explore the role of this prediction model in different populations.

Peer Review reports

Background

The first phakic intraocular lens (IOL) is implanted in the anterior chamber angle (ACA), which causes endothelial decompensation and glaucoma [1]. Iris-claw (iris-fixated) IOLs located in the anterior chamber without endothelial touching have been used to correct refractive errors, but significant endothelial cell loss in some patients has been reported [2]. Posterior chamber phakic IOLs, which are located behind the iris, have been introduced most recently to minimize the effect of IOL on corneal and anterior chamber angles. The implantable collamer lens (EVO Visian implantable collamer lens [ICL] with KS-AquaPort, STAAR Surgical, USA) has been widely used for phakic intraocular lens implantation [3]. This surgical procedure is currently acknowledged as a safe and effective method for vision correction for a wide range of refractive errors. It almost preserves the cornea and accommodation function of the crystalline lens after surgery. Therefore, ICL is an important surgical option for high levels of myopia, hyperopia, and astigmatism, which cannot be corrected by corneal laser ablation. In this surgical procedure, an ICL is implanted in the ciliary sulcus in the posterior chamber, far from the corneal endothelium [4]. Thus, irreversible damage to the corneal endothelium caused by IOL may be minimized, and IOL implantation is reversible and replaceable with another IOL.

The postoperative vault, which is the distance between the IOL and crystalline lens, is an important factor in selecting the optimal IOL size to reduce complications [5]. The consensus is that the ideal vault, which is the gap between the IOL and crystalline lens, should be approximately 500 μm to prevent postoperative complications. A very low vault may be associated with subcapsular cataract [6]. In addition, a vault that is too high significantly reduces the ACA opening. Too large IOL may cause iridocorneal touch and ACA block [7]. Consequently, it may increase the intraocular pressure (IOP) and reduce the number of corneal endothelial cells [8]. The abnormal position of the ICL is also associated with an improper high vault [9]. Because the postoperative vault depends on the anatomy of the anterior chamber, space between the iris and crystalline lens, and size of the ICL, it is essential to select an appropriate ICL size before surgery. The postoperative vault is often much higher or lower than intended, and reoperation is sometimes needed to avoid vision-related complications [10].

Although the manufacturer provides ICL sizing based on corneal size and anterior chamber depth (ACD), directly measuring the anatomical space into which the ICL is located is more accurate in determining the optimal ICL size [11]. Anterior segment optical coherence tomography (AS-OCT) was recently developed to measure the anatomy of the anterior segment of the eye more accurately. CASIA2 (Tomey, Nagoya, Japan) and ANTERION (Heidelberg Engineering GmbH, Heidelberg, Germany) are swept-source AS-OCT devices that provide rapid image acquisition and adequate visualization depth. ANTERION can perform a wider and deeper scan than other AS-OCT devices [12]. It uses a 1300 nm infrared light source to capture the anterior segments with a high resolution (< 10 μm). The embedded software measures the anterior chamber dimensions in six evenly spaced radial scans over 12 angle locations. The infrared camera captures horizontal cross-sectional images measuring the entire anterior segment and laterals and performs en face imaging of the subject’s eye (Fig. 1). Furthermore, previous studies have reported that it outperforms CASIA2 and IOLMaster in terms of intra-device repeatability [13],[14]. Therefore, the use of ANTERION for clinical purposes is on the rise. However, CASIA2 has been widely used for ICL implantation surgery with several lens-sizing formulas [15], whereas no studies have been conducted on lens sizing using biometry from ANTERION.

Fig. 1
figure 1

Examples of the ANTERION result pages for anterior segment measurements

Previous studies have shown that the structures of the posterior chamber, especially the ciliary sulcus, measured by ultrasound biomicroscopy (UBM) or ultrasound scanners, might provide better vault predictability [16]. UBM has the advantage of visualizing the structures behind the iris, which cannot be observed in AS-OCT images [17]. However, measurement using UBM is uncomfortable because of the placement of an eyecup between the lids and is very time-consuming in the clinic [4]. These difficulties may lead to a low repeatability. Therefore, AS-OCT measurement, which has a higher scanning speed and better resolution of the anterior chamber, has been the gold standard method for capturing the structures of the anterior segment for ICL surgery in many clinics.

According to a recent study, biometry measurements from ANTERION and CASIA2 are not interchangeable [13]. Therefore, a new ICL sizing formula is needed for clinics that use ANTERION to achieve more accurate ICL sizing and postoperative vault prediction. In this study, we developed an ICL sizing formula for the ANTERION AS-OCT device. We evaluated the effect of biometry measurements from ANTERION and performed regression analysis to calculate the optimal ICL size on the achieved vault.

Materials and methods

Dataset

We retrospectively collected the preoperative and postoperative ocular measurement data from the B&VIIT Eye Center (Seoul, South Korea). This study was approved by the institutional review board of the Korean National Institute for Bioethics Policy (No. 2021-3387-001). The patients underwent refractive surgery with posterior phakic intraocular lens implantation using ICL (V4c and V5 models, EVO Visian ICL with KS-AquaPort) from February 2021 to September 2021. The inclusion criteria for this study were as follows: age between 18 and 50 years, stable refraction, − 0.50 to − 20.00 diopters of hyperopia or myopia with astigmatism of 5.50 D or less, and availability of the preoperative scanning results of ANTERION, CASIA2, and the ICL vault at 1 month post-surgery. AS-OCT was performed under dark-light conditions using blackout curtains. To ensure a non-accommodative state, the patient was asked to stay far away during the examination. One trained observer marked the scleral spurs in each image during the examination. We also measured the anterior chamber width (ACW) and crystalline lens rise (CLR) using CASIA2 to compare the developed formula with the NK formula. It should be noted that CLR is currently not measured using ANTERION (Fig. 1).

The workflow for data management is shown in Fig. 2. Subjects with a history of ocular surgery, corneal disease, glaucoma, uveitis, or retinal disease were excluded from the study. Patients with missing data were excluded from this study. This study included 892 eyes of 471 healthy subjects treated with ICLs. To design the retrospective development of the model and prospective validation via chronological splitting, we assigned patients who visited before June 2021 (80%, n = 696 eyes of 368 subjects) to the development dataset and those who visited after July 2021 (20%, n = 196 eyes of 103 subjects) to the test validation dataset.

Fig. 2
figure 2

Workflow for data management for the development of regression models using ANTERION for ICL sizing

This study aimed to develop a formula for postoperative vault prediction to select the optimal ICL size. Because there are currently four commercial ICL sizes (12.1, 12.6, 13.2, and 13.7 mm), surgeons should select the one that achieves the best outcome. The ICL vault prediction formulas were developed using the development dataset via various linear regression techniques, including ICL size and other measurements from ANTERION. Postoperative vault was measured using ANTERION. We set a postoperative ICL vault of 500 μm as the best result for obtaining the optimal ICL size. Therefore, in the validation stage, we selected the ICL size that yielded the predicted result of the ICL vault closest to 500 μm.

Phakic ICL implantation surgeries were performed using standard methods as described in previous studies [5],[18]. The lens was implanted in the posterior chamber via a 3.0-mm-sized temporal clear corneal incision. Expert surgeons (IHR, JKK, and ISL) with an average experience of 10 years determined the size of the ICL by considering the manufacturer’s nomogram (based on ACD and white-to-white distance [WTW]), an in-house developed nomogram [5], and ocular measurements of the ACA and spur-to-spur (STS) distances from ANTERION. The in-house nomogram mainly used ACW, ACD, ICL power, CLR, angle-to-angle distance (ATA), pupil size, and WTW to calculate the postoperative vaults. Before surgery, the surgeons determined the lens size for each patient based on postoperative IOL vault predictions, with a target size of 500 μm. All experts were board-certified ophthalmologists with an average experience of 5 years in ICL surgery.

Regression techniques

We constructed six ICL vault prediction models: STAAR nomogram-based, KS formula-based, forward stepwise selection, backward stepwise selection, partial least squares (PLS), and least absolute shrinkage and selection operator (LASSO) regression models. Multivariable linear regression models (STAAR nomogram-based, KS formula-based, forward stepwise selection, and backward stepwise selection models) were fitted using the least-squares approach in a standard manner, provided by the statistical software SPSS Statistics v.24 (IBM Corp., Armonk, NY, USA). PLS and LASSO are well-known advanced regression methods for dimension reduction and robust prediction in biomedical fields [19]. These techniques are fundamentally based on familiar expressions used for multivariate linear regression. The PLS technique analyzes new vector components by combining predictors. A previous study reported that it performed well in ICL sizing using Visante OCT (Carl Zeiss, Germany) [20]. LASSO leads to a sparse regression solution for the coefficients corresponding to the most important predictors [21]. LASSO shows better identification of predictors than other regression methods [22] and outperforms machine learning techniques in predicting clinical outcomes in ophthalmology [23]. We used SPSS 23.0 (SPSS Inc., Chicago, USA) for classic statistical analysis, as well as MATLAB R2021a (MathWorks Inc., Natick, MA, USA) for PLS (“plsregress” function) and LASSO (“lasso” function) techniques.

The input variables from preoperative ANTERION measurements included central corneal thickness (CCT), aqueous depth (AQD), anterior chamber volume, ACA distance, STS distance, lens thickness (LT), pupil diameter, and WTW. We also analyzed age, sex, spherical equivalent, mean keratometry (K), ICL power, ICL type (toric or non-toric lens), and ICL size to build the ICL vault prediction model.

Model selection

After the training process with the entire development dataset, we evaluated the prediction abilities based on the Bayesian information criterion (BIC) to select the best prediction model. BIC is a reliable indicator for comparing the effectiveness of prediction models and has been widely used in prediction model selection when different numbers of input variables are included [22]. The BIC penalizes the number of variables to avoid an unstable or overfitting regression model. In this study, BIC is expressed as follows [24]:

$$BIC=k \text{log}\left(n\right)-2\text{log}\left(\widehat{L}\right)= k \text{log}\left(n\right)+n\text{log}\left(\frac{RSS}{n}\right)$$

where \(k\) is the number of predictors, \(n\) is the number of samples in the validation set, \(\widehat{L}\) is the likelihood of the model, and \(RSS\) is the residual sum of squares of the regression result. The best model yielded the lowest BIC value and was the most effective model without overfitting. In this study, we determined the best model based on the BIC values in the validation.

We also compared our method with the NK formula based on ACW and CLR. Since the participants in this study had a different distribution of CLR than those in the original study (CLR value in our development set from Table 1 = − 0.077 ± 0.138 mm; CLR value in the Japanese population = 0.039 ± 0.183 mm), we applied the coefficient values of the NK formula optimized to the Korean population. Modified NK formula developed for the previous study [5] is expressed as follows:

$$\begin{array}{l}{\rm{Optimal\ ICL\ size (mm) = 8}}{\rm{.16 + 0}}{\rm{.36}} \times \\{\rm{ACW(mm) + 1}}{\rm{.03}} \times {\rm{CLR (mm)}}\end{array}$$

Pearson’s correlation coefficients and mean squared error (MAE) between the achieved and predicted vault values were used to evaluate the regression models. All statistical tests were performed in a two-sided manner, with the significance level set at a P-value < 0.050.

Table 1 Preoperative demographics and postoperative ICL vaults of the study participants

Results

The demographics and measurements of the study participants in the development and validation datasets are shown in Table 1. Except for ACA distance (P = 0.044) and STS distance (P = 0.006), there were no significant differences between the development and validation datasets. There was only one case using an ICL with a size of 13.7 mm in the development dataset and none in the validation set. The mean achieved ICL vault values at 1 month post-surgery were 513.94 ± 162.76 μm in the development dataset and 514.01 ± 148.06 μm in the validation dataset. According to the retrospective chart review, all surgeries were uneventful and there were no vision-threatening complications.

Table 2 shows the descriptive statistics of the preoperative variables and ICL size to predict postoperative ICL vaults. The coefficients of the descriptive statistics were calculated using correlation coefficients and standard multivariable regression. The predictors, including AQD (r = 0.441, P < 0.001), LT (r = − 0.418, P < 0.001), and ICL size (r = 0.392, P < 0.001), showed a very strong correlation with the postoperative vault. The anterior chamber volume (ACV), ACA distance, STS distance, and WTW were also significantly correlated. In the multivariable regression model using all predictors, ICL size, AQD, LT, and ACA distance had an impact of more than 10% on the ICL vault calculation, with statistical significance.

Table 2 Correlation analysis and multivariable linear regression results for postoperative vault prediction using ANTERION

Table 3 presents the prediction results of the regression models developed using the development and validation datasets. In the self-validation of the development dataset, PLS showed the best prediction performance using age, sex, toric lens, mean K, ICL size, ICL power, CCT, AQD, ACV, LT, ACA distance, pupil diameter, and WTW (13 predictors) as well as the strongest correlation (r = 0.664, P < 0.001) and lowest MAE (96.9 μm). When we investigated the BIC to consider the effectiveness of the vault prediction models in the development dataset, the LASSO model using ICL size, AQD, ACA distance, and LT yielded the best results. The final LASSO model used fewer predictors, with a smaller loss in prediction than the PLS model. In the validation dataset, the LASSO model showed the strongest correlation (r = 0.582, P < 0.001) and the lowest MAE (104.7 μm). It also had the best effectiveness for vault prediction according to BIC among the regression models.

Table 3 Multiple regression analysis for postoperative ICL vault prediction in the development dataset

Finally, we selected the LASSO model as the best prediction model for optimal ICL sizing, because it showed the lowest BIC value in both the development (5769.9) and validation (1894.9) sets. The smallest number of predictors in the LASSO model contributed to the best effectiveness for vault prediction. The predicted postoperative vault was calculated according to the LASSO model, as follows:

$$\begin{array}{l}{\rm{Postoperative\ ICL\ vault (}}\mu m)\\= {\rm{ - 1052}}{\rm{.26 + 129}}{\rm{.94}} \times {\rm{AQD (mm) - 134}}{\rm{.54}} \times \\{\rm{ACA\ distance(mm) - 217}}{\rm{.53}} \times {\rm{LT(mm) + }}\\{\rm{283}}{\rm{.62}} \times {\rm{ICL size (mm)}}\end{array}$$

As we assumed that the ideal ICL vault was 500 μm, the optimal ICL size was calculated as follows:

$$\begin{array}{l}{\rm{Optimal\ ICL\ size (mm) = 5}}{\rm{.472 - 0}}{\rm{.458}} \times {\rm{AQD(mm) + }}\\{\rm{0}}{\rm{.474}} \times {\rm{ACA\ distance (mm) + 0}}{\rm{.767}} \times {\rm{LT (mm)}}\end{array}$$

The surgeon selected the ICL lens size from among the four commercial sizes (12.1, 12.6, 13.2, and 13.7 mm) closest to the calculated optimal ICL size. It should be noted that the use of a large lens with a size of 13.7 mm was extremely rare in the development dataset for these equations. We developed a simple web-based calculator application using a final equation (https://soo9028.github.io/iol-prediction-webpage/).

Figure 3 shows the validation results for the LASSO model. In the Bland-Altman analysis, the average difference between the real achieved and LASSO-predicted vault values was 4.9 μm, and its standard deviation was 120.8 μm. When we analyzed the distribution of the accuracies according to the predicted vault, the LASSO model showed significantly better prediction performance than the STAAR nomogram-based model (P < 0.001 in the χ2 test).

Fig. 3
figure 3

Distribution of the achieved vault against the predicted vault using the LASSO formula in the validation dataset. (A) Bland-Altman plot. (B) Vault errors between the proposed LASSO formula and the STAAR nomogram

The developed model was compared with the achieved results and the NK formula in the validation data (Fig. 4). The correlation coefficient between the achieved vaults and the developed LASSO was 0.582 (P < 0.001). The correlation coefficient between the LASSO and NK formulas was 0.768 (P < 0.001). Figure 5 shows that the distributions of the achieved vault (514.01 ± 148.06), NK formula vault prediction (511.52 ± 81.57), and developed LASSO formula vault prediction (516.57 ± 98.37) values showed no significant difference between each other. In addition, the MAEs showed no difference between the two formulas (P = 0.220).

Fig. 4
figure 4

Scatter plots showing the postoperative ICL vault prediction results in the validation dataset. (A) Distribution of the achieved vault against the LASSO predicted vault. (B) Distribution of the LASSO formula against the modified NK formula

Fig. 5
figure 5

Comparison between the LASSO formula based on ANTERION and the modified NK formula based on CASIA2. (A) Comparison between the achieved vault and prediction results. (B) Mean absolute error (MAE) comparison between the LASSO and modified NK formulas

Discussion

Appropriate IOL sizing with accurate postoperative vault prediction is necessary to achieve safer ICL surgeries and better clinical outcomes. We introduced the first formula to estimate the postoperative ICL vault for the novel AS-OCT device, ANTERION, which has not been used for ICL surgery. Compared to previous studies that focused on regression and feature extraction, our study contributes to the adoption of LASSO and BIC to select a better prediction model for external validation. In addition, we developed a web-based calculator (https://soo9028.github.io/iol-prediction-webpage/) for better accessibility to the developed formula. Our method was almost equivalent to the NK formula based on CASIA2, which has been widely used with reliable performance in predicting postoperative ICL results. As ANTERIORN does not provide CLR values (Fig. 1), the LASSO algorithm selected LT as a significant variable to estimate the impact of the crystalline lens on IOL position. The final formula showed better results than the STAAR nomogram, which used limited predictors, including WTW and ACD. ANTERION provided more significant metrics for estimating the posterior chamber space for ICL implantation than WTW measured at the ocular surface.

Table 4 presents a literature review of ICL size and postoperative vault predictions. Several regression approaches have been developed to achieve vault prediction and optimal ICL sizing. Some of these approaches use UBM to measure the anterior segment [11], whereas others are based on AS-OCT measurements, mainly CASIA2 [15]. UBM (or ultrasound scanner) is a non-invasive technique that uses high-frequency ultrasound transducers (50–100 MHz) for imaging of the anterior segment including iris, crystalline lens, and ciliary body [25]. In a study using UBM, the regression model included ACD, sulcus-to-sulcus [26], and sulcus-to-sulcus lens rise [11]. Recently, a digital ultrasound scanner (ArcScan Inc., Morrison, Colorado) measured the inner diameter of the ciliary body and predicted the postoperative vault more accurately than the UBM [16]. The CASIA2 provides reproducible measurements of the cornea, ACA, iris, and CLR. It has been used to diagnose angle-closure [27] and measure corneal and anterior chamber structures [28]. In a study using CASIA2, the NK formula included ACW and CLR to predict vault size [29]. The ATA from CASIA2 was used as a reliable variable in the KS formula [30]. Recently, the ATA-based formula was revised for more accurate vault prediction at IOL sizes of 12.1, 13.2, and 13.7 mm in a validation study [31]. Most studies have used classic linear regression techniques [29], and PLS has been adopted to achieve more accurate predictions [20]. Recent studies have employed machine-learning approaches using large datasets and many input variables [18]. Although large machine-learning-based models have been applied, too many variables are required to obtain a modest performance gain [32]. A previous study emphasized the ethnic differences and developed a LASSO-based model using MS-39 (Costruzione Strumenti Oftalmici, Florence, Italy) for Caucasian eyes [33]. In this study, we predicted postoperative ICL vault using LASSO regression based on the measurement data from ANTERION. Our model included only three variables (AQD, ACA distance, and LT) and showed the best prediction performance and effectiveness.

Table 4 Summary of ICL sizing and postoperative vault prediction using anterior segment imaging domains

We developed a new ICL sizing formula based on ANTERION. Using our formula, ANTERION can be used for optimal ICL sizing in the clinic. Because the measurements from two swept-source AS-OCT devices, ANTERION and CASIA2, are not interchangeable [13], the direct use of the anatomic feature values from ANTERION in the formulas for CASIA2 is limited and does not guarantee accurate prediction. According to the literature, ANTERION has several advantages over other AS-OCT devices [13]. It can visualize the ciliary muscle and entire crystalline LT with a high resolution of < 10 μm and can also measure the axial length of the eye. The posterior lens surface cannot be visualized using other AS-OCT techniques, such as CASIA2 and RTVue (Optovue Inc., Fremont, California, USA). ANTERION uses the image averaging technique, which is a new feature of AS-OCT devices, to improve the signal-to-noise ratio to measure wider and deeper structures of the anterior segments [34]. It also showed high repeatability for anatomical measurements of the anterior segments. As expected, the axial measurements (AQD and LT) and width (ACA distance) were selected for the final LASSO regression model, similar to the NK formula using CASIA2 [15]. The difference is that our formula contains ACD and LT without using a crystalline lens rise or crystalline lens vault (LV). In this study, we used only measurements captured using a fully automated process. We did not collect LV measurements because they vary depending on the light conditions and study population [5],[35]. According to our results, LT is also highly correlated with postoperative ICL vault, and the combination of LT and other measurements showed better predictive performance than the STAAR nomogram-based and KS formula-based methods, which do not contain LT information.

The simplest linear model, LASSO, outperformed the complex regression approaches, which usually achieved good results in other studies. This may be due to the occurrence of overfitting to the development set in complex regression models using many variables. This result indicates that none of the predictors improved the performance in predicting postoperative ICL vaults. Previous studies have also shown that LASSO with fewer predictors outperforms other complex regressions [22],[23]. Because LASSO assigns zero to most predictors with low impact, it is considered a good feature selection technique [19]. In our study, because the predictors from ANTERION had many intercorrelated associations and different variations, the greater capacity of other complex regressions with more variables seems to lead to overfitting. The standard regression provided insightful results regarding the impact of each variable on vault calculation (Table 2). However, standard linear regression assumes that all predictors are independent of each other [36]. Therefore, it has a multicollinearity problem and works worse than the LASSO model. LASSO can control the multicollinearity between predictors by reducing predictors, and can finally avoid an overfitting problem [37]. The BIC metric successfully evaluated both the regression performance and the effectiveness of ICL vault prediction by penalizing the number of predictors. BIC found that the LASSO-based formula is the most robust model for predicting the ICL vault because it avoids overfitting.

This study had several limitations. First, it was conducted in a single center involving the East Asian population. There may be differences in the anatomical features of the anterior segment between different ethnic groups [38]. In addition, there was no external validation to evaluate the performance of the ICL sizing. Therefore, surgeons in other centers should carefully apply the developed formula, particularly in different ethnic populations. Second, the use of a large lens with a size of 13.7 mm was extremely rare in our study. Therefore, we could not confirm whether our formula can be used for the largest ICL size. Third, we did not consider other ocular measurements such as axial length or corneal parameters. ANTERION provides axial length measurements, but we did not include them to focus on the anterior chamber structure. Anterior chamber shapes have also been associated with factors measured by other modalities in previous studies [39],[40]. Fourth, we did not consider ICL power in this study. According to a previous study, postoperative refraction was associated with the postoperative IOL vault level [41]. This means that our formula can be integrated into the calculation of the ICL refractive power.

Conclusion

To our knowledge, this is the first study to develop a postoperative ICL vault prediction model based on the ANTERION AS-OCT. It can be used to optimize the ICL size to avoid postoperative complications. In the present study, the LASSO regression model using AQD, ACA distance, and LT showed better performance in estimating ICL vault and ICL sizing than other regression models. Because our model was developed based on the East Asian population, further studies are needed to explore the role of this prediction model in different populations.

Data Availability

The datasets generated and/or analyzed during the current study are not publicly available due to personal privacy concerns but are available from the corresponding author on reasonable request. The developed calculator is available at https://soo9028.github.io/iol-prediction-webpage/.

Abbreviations

ACA:

Anterior chamber angle

ACD:

Anterior chamber depth

ACV:

Anterior chamber volume

ACW:

Anterior chamber width

AQD:

Aqueous depth

AS-OCT:

Anterior segment optical coherence tomography

ATA:

Angle-to-angle distance

BIC:

Bayesian information criterion

CCT:

Central corneal thickness

CLR:

Crystalline lens rise

ICL:

Intraocular collamer lens

IOL:

Intraocular lens

LASSO:

Least absolute shrinkage and selection operator

LT:

Lens thickness

LV:

Lens vault

MAE:

Mean squared error

PLS:

Partial least squares

STS:

Spur-to-spur

UBM:

Ultrasound biomicroscopy

WTW:

White-to-white distance

References

  1. Chen L-J, Chang Y-J, Kuo JC, Rajagopal R, Azar DT. Metaanalysis of cataract development after phakic intraocular lens surgery. J Cataract Refractive Surg. 2008;34:1181–200.

    Article  Google Scholar 

  2. Galvis V, Villamil JF, Acuña MF, Camacho PA, Merayo-Lloves J, Tello A, et al. Long-term endothelial cell loss with the iris-claw intraocular phakic lenses (Artisan®). Graefes Arch Clin Exp Ophthalmol. 2019;257:2775–87.

    Article  PubMed  Google Scholar 

  3. Alfonso JF, Fernández-Vega L, Lisa C, Fernandes P, González-Meijome J, Montés-Micó R. Long-term evaluation of the central vault after phakic Collamer® lens (ICL) implantation using OCT. Graefes Arch Clin Exp Ophthalmol. 2012;250:1807–12.

    Article  PubMed  Google Scholar 

  4. Wan T, Yin H, Yang Y, Wu F, Wu Z, Yang Y. Comparative study of anterior segment measurements using 3 different instruments in myopic patients after ICL implantation. BMC Ophthalmol. 2019;19:182.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Kang EM, Ryu IH, Lee G, Kim JK, Lee IS, Jeon GH, et al. Development of a web-based ensemble machine learning application to select the optimal size of posterior Chamber Phakic intraocular Lens. Volume 10. Trans Vis Sci Tech. The Association for Research in Vision and Ophthalmology; 2021. pp. 5–5.

  6. Fernandes P, González-Méijome JM, Madrid-Costa D, Ferrer-Blasco T, Jorge J, Montés-Micó R. Implantable collamer posterior chamber intraocular lenses: a review of potential complications. J Refract Surg. 2011;27:765–76.

    Article  PubMed  Google Scholar 

  7. Choi H, Kim T, Kim SJ, Sa BG, Ryu IH, Lee IS, et al. Predicting Postoperative Anterior Chamber Angle for Phakic intraocular Lens Implantation using Preoperative Anterior Segment Metrics. Translational Vis Sci Technol. 2023;12:10.

    Article  Google Scholar 

  8. Yang W, Zhao J, Sun L, Zhao J, Niu L, Wang X, et al. Four-year observation of the changes in corneal endothelium cell density and correlated factors after Implantable Collamer Lens V4c implantation. Br J Ophthalmol BMJ Publishing Group Ltd. 2021;105:625–30.

    Google Scholar 

  9. Shi M, Kong J, Li X, Yan Q, Zhang J. Observing implantable collamer lens dislocation by panoramic ultrasound biomicroscopy. Eye Nat Publishing Group. 2015;29:499–504.

    CAS  Google Scholar 

  10. Lee D-H, Choi S-H, Chung E-S, Chung T-Y. Correlation between preoperative biometry and posterior Chamber Phakic Visian Implantable Collamer Lens Vaulting. Ophthalmology. 2012;119:272–7.

    Article  PubMed  Google Scholar 

  11. Kojima T, Yokoyama S, Ito M, Horai R, Hara S, Tomoaki N, et al. Optimization of an Implantable Collamer Lens sizing Method using high-frequency Ultrasound Biomicroscopy. Am J Ophthalmol Elsevier. 2012;153:632–637e1.

    Article  Google Scholar 

  12. Montés-Micó R, Tañá-Rivero P, Aguilar-Córcoles S, Ruiz-Santos M, Rodríguez-Carrillo MD, Ruiz-Mesa R. Angle-to-angle and spur-to-spur distance analysis with high-resolution optical coherence tomography. Eye and Vision. 2020;7:42.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Cheng S-M, Zhang J-S, Li T-T, Wu Z-T, Wang P, Yu A-Y. Repeatability and agreement of two swept-source Optical Coherence Tomographers for Anterior Segment parameter measurements. J Glaucoma. 2022;31:602–8.

    Article  PubMed  Google Scholar 

  14. Schiano-Lomoriello D, Hoffer KJ, Abicca I, Savini G. Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices. Sci Rep Nature Publishing Group. 2021;11:983.

    CAS  Google Scholar 

  15. Nakamura T, Isogai N, Kojima T, Yoshida Y, Sugiyama Y. Optimization of implantable collamer lens sizing based on swept-source anterior segment optical coherence tomography. J Cataract Refractive Surg. 2020;46:742–8.

    Article  Google Scholar 

  16. Reinstein DZ, Archer TJ, Vida RS, Piparia V, Potter JG. New Sizing Parameters and Model for Predicting Postoperative Vault for the Implantable Collamer Lens posterior Chamber Phakic intraocular Lens. J Refractive Surg SLACK Incorporated. 2022;38:272–9.

    Article  Google Scholar 

  17. Ishikawa H. Anterior segment imaging for glaucoma: OCT or UBM? Br J Ophthalmol. 2007;91:1420–1.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Kamiya K, Ryu IH, Yoo TK, Kim JS, Lee IS, Kim JK, et al. Prediction of phakic intraocular Lens Vault using machine learning of Anterior Segment Optical Coherence Tomography Metrics. Am J Ophthalmol. 2021;226:90–9.

    Article  PubMed  Google Scholar 

  19. Datta S, Le-Rademacher J, Datta S. Predicting patient survival from microarray data by accelerated failure time modeling using partial least squares and LASSO. Biometrics. 2007;63:259–71.

    Article  CAS  PubMed  Google Scholar 

  20. Oleszko A, Marek J, Muzyka-Wozniak M. Application of a partial least squares regression algorithm for posterior Chamber Phakic intraocular Lens Sizing and Postoperative Vault Prediction. J Refract Surg. 2020;36:606–12.

    Article  PubMed  Google Scholar 

  21. Tibshirani R. Regression shrinkage and selection via the lasso.Journal of the Royal Statistical Society Series B (Methodological). 1996;267–88.

  22. Oh E, Yoo TK, Park E-C. Diabetic retinopathy risk prediction for fundus examination using sparse learning: a cross-sectional study. BMC Med Inform Decis Mak. 2013;13:106.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Rohm M, Tresp V, Müller M, Kern C, Manakov I, Weiss M, et al. Predicting Visual Acuity by using machine learning in patients treated for Neovascular Age-Related Macular Degeneration. Ophthalmology. 2018;125:1028–36.

    Article  PubMed  Google Scholar 

  24. Shi P, Tsai C-L. Regression model selection—a residual likelihood approach. J Royal Stat Society: Ser B (Statistical Methodology). 2002;64:237–52.

    Article  Google Scholar 

  25. Pavlin CJ, Harasiewicz K, Sherar MD, Foster FS. Clinical use of Ultrasound Biomicroscopy. Ophthalmology. 1991;98:287–95.

    Article  CAS  PubMed  Google Scholar 

  26. Dougherty PJ, Rivera RP, Schneider D, Lane SS, Brown D, Vukich J. Improving accuracy of phakic intraocular lens sizing using high-frequency ultrasound biomicroscopy. J Cataract Refractive Surg. 2011;37:13–8.

    Article  Google Scholar 

  27. Wang X, Chen X, Tang Y, Wang J, Chen Y, Sun X. Morphologic Features of Crystalline Lens in Patients with Primary Angle Closure Disease Observed by CASIA 2 Optical Coherence Tomography. Investigative Ophthalmology & Visual Science. 2020;61:40.

  28. Schröder S, Langenbucher A, Schrecker J. Comparison of corneal elevation and pachymetry measurements made by two state of the art corneal tomographers with different measurement principles. PLOS ONE Public Library of Science. 2019;14:e0223770.

    Article  Google Scholar 

  29. Nakamura T, Isogai N, Kojima T, Yoshida Y, Sugiyama Y. Implantable Collamer Lens sizing Method based on swept-source Anterior Segment Optical Coherence Tomography. Am J Ophthalmol. 2018;187:99–107.

    Article  PubMed  Google Scholar 

  30. Igarashi A, Shimizu K, Kato S, Kamiya K. Predictability of the vault after posterior chamber phakic intraocular lens implantation using anterior segment optical coherence tomography. J Cataract Refractive Surg. 2019;45:1099–104.

    Article  Google Scholar 

  31. Igarashi A, Shimizu K, Kato S. Assessment of the Vault after Implantable Collamer Lens Implantation using the KS Formula. J Refractive Surg SLACK Incorporated. 2021;37:636–41.

    Article  Google Scholar 

  32. Shen Y, Wang L, Jian W, Shang J, Wang X, Ju L et al. Big-data and artificial-intelligence-assisted vault prediction and EVO-ICL size selection for myopia correction.Br J Ophthalmol. 2021;bjophthalmol-2021-319618.

  33. Rocamora L, Orlando JI, Lwowski C, Kohnen T, Mertens E, Van Keer K. Postoperative vault prediction for phakic implantable collamer lens surgery: LASSO formulas. J Cataract Refractive Surg. 2023;49:126.

    Article  Google Scholar 

  34. Pardeshi AA, Song AE, Lazkani N, Xie X, Huang A, Xu BY. Intradevice repeatability and interdevice agreement of ocular biometric measurements: a comparison of two swept-source anterior segment OCT Devices. Translational Vis Sci Technol. 2020;9:14.

    Article  Google Scholar 

  35. Tañá-Rivero P, Ruiz-Mesa R, Aguilar-Córcoles S, Tello-Elordi C, Ramos-Alzamora M, Montés-Micó R. Lens-vault analysis and its correlation with other biometric parameters using swept-source OCT. J Optom. 2022;15:88–99.

    Article  PubMed  Google Scholar 

  36. Murata H, Asaoka R, Fujino Y, Matsuura M, Hirasawa K, Shimada S et al. Comparing the usefulness of a new algorithm to measure visual field using the variational Bayes linear regression in glaucoma patients, in comparison to the Swedish interactive thresholding algorithm. British Journal of Ophthalmology. BMJ Publishing Group Ltd; 2022;106:660–6.

  37. Shi J, He Q, Pan Y, Zhang X, Li M, Chen S. Estimation of Appendicular Skeletal Muscle Mass for Women Aged 60–70 Years Using a Machine Learning Approach.Journal of the American Medical Directors Association. 2022;23:1985.e1-1985.e7.

  38. Chan SM, Svitova TF, Lin MC. Accounting for Ethnicity-Related Differences in Ocular Surface Integrity as a Step Toward Understanding Contact Lens Discomfort. Eye & Contact Lens: Science & Clinical Practice. 2017;43:23–31.

  39. Hosny M, Ali ó JL, Claramonte P, Attia WH. P érez-SJJ. Relationship between Anterior Chamber depth, refractive state, corneal diameter, and axial length. J Refractive Surg SLACK Incorporated. 2000;16:336–40.

    Article  CAS  Google Scholar 

  40. Yoo TK, Ryu IH, Kim JK, Lee IS, Kim HK. A deep learning approach for detection of shallow anterior chamber depth based on the hidden features of fundus photographs. Comput Methods Programs Biomed. 2022;219:106735.

    Article  PubMed  Google Scholar 

  41. Serra P, Sánchez Trancón Á, Torrado Sierra O, Baptista A, Cerpa Manito S. Posterior Chamber Phakic intraocular lenses for the correction of myopia: factors influencing the postoperative refraction. Optics. Volume 2. Multidisciplinary Digital Publishing Institute; 2021. pp. 292–305.

Download references

Acknowledgements

Not applicable.

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

Study conceptualization: TK, IHR, and TKY. Data collection: BYL and HJC. Statistical analysis plan and critical revision of methodology: IHR, JKK, ISL, and TKY. Data analysis: TK, SJK, and BGS. Drafting the manuscript: TK, SJK, and TKY. HK, EH, and TKY revised the manuscript. Critical revision of the manuscript and final approval: All authors. TK and SJK contributed equally to the study.

Corresponding author

Correspondence to Tae Keun Yoo.

Ethics declarations

Ethics approval and consent to participate

This retrospective study was approved by the institutional review board of the Korean National Institute for Bioethics Policy (KNIBP; No. 2021-3387-001). All participants provided written informed consent prior to the preoperative examinations. All methods were performed in accordance with the Declaration of Helsinki and the KNIBP guidelines.

Consent for publication

Not applicable.

Competing Interests

Ik Hee Ryu, Jin Kuk Kim, and Tae Keun Yoo are executives of VISUWORKS, Inc., which is a Korean Artificial Intelligence startup that provides medical machine learning solutions. Jin Kuk Kim is an executive of the Korea Intelligent Medical Industry Association. They received salaries or stocks as part of their standard compensation packages. The remaining authors declare no conflicts of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kim, T., Kim, S.J., Lee, B.Y. et al. Development of an implantable collamer lens sizing model: a retrospective study using ANTERION swept-source optical coherence tomography and a literature review. BMC Ophthalmol 23, 59 (2023). https://doi.org/10.1186/s12886-023-02814-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-023-02814-7

Keywords