Optical bench evaluation of the effect of pupil size in new generation monofocal intraocular lenses

Abstract

Background

A new generation of enhanced monofocal IOLs has been introduced to slightly increase the depth of focus as compared to standard monofocal IOLs. The purpose of this study is to evaluate the effect of pupil size on the through-focus optical performance of three new enhanced monofocal IOLs, designed to improve the range of vision as compared to standard monofocal IOLs.

Methods

Optical bench testing in white light was performed for different pupils, using an average cornea eye. Distance image quality was evaluated using Modulation Transfer Function (MTF) measurements. Through-focus Visual Acuity (VA) was simulated from these measurements (sVA). Three enhanced monofocal IOLs (ICB00, ISOPure, and RayOne-EMV) and three standard monofocal IOLs: two aspheric (ZCB00 and SN60WF) and one spherical (AAB00) were included.

Results

The enhanced monofocal IOLs provided an improvement in the intermediate sVA as compared to standard monofocal IOLs. For ICB00, the improvement was independent of the pupil size, while for the ISOPure and RayOne-EMV, the intermediate sVA improved with increased pupil size. Similar to the spherical monofocal IOL, the ISOPure and RayOne-EMV showed a strong correlation between improvement in intermediate sVA and reduction of distance sVA and MTF, and increasing pupil size. ICB00 provided the same distance sVA as the aspheric monofocal IOLs and the lowest variability in MTF with pupil size.

Conclusion

Optical bench results showed that the ISOPure and RayOne-EMV provide similar performance to a spherical monofocal IOL, with a strong pupil dependency for distance and intermediate vision. The other enhanced monofocal IOL, ICB00, provided a sustained improvement in simulated intermediate VA and maintained distance image quality comparable to that of the standard aspheric monofocal IOLs, even for larger pupils.

Peer Review reports

What was known

  • Distance MTF is reduced with larger pupil size in eyes with imperfect correction of SA.

  • New generation enhanced monofocal IOLs can increase the depth of focus.

What this paper adds

  • The effect of the pupil size on the optical image quality and simulated visual acuity of different monofocal IOLs depends on the optical technology and the correction of corneal spherical aberration.

  • The new generation monofocal IOL, designed to improve the depth of focus while fully correcting corneal spherical aberration, shows a pupil-independent improvement in simulated intermediate visual acuity and maintained distance visual acuity compared to conventional monofocal IOLs.

Introduction

The standard of care in cataract surgery is monofocal intraocular lenses (IOLs). The current monofocal lenses comprise a variety of different optical concepts, which can be characterized by the optical design, that determines for example the amount of Spherical Aberration (SA) that is being induced by the optic, the spectral filtering properties, the chromatic dispersion of the material, and, of course, the geometrical design (e.g., sharp optic edge, haptic design, and angulation). The amount of SA is of particular interest, as it interacts with the SA of the patient’s cornea and thus determines the achievable optical resolution and depth of focus. There are mainly three different IOL concepts on the market based on SA. The first group is conventional spherical IOLs, which feature a spherical optical surface and therefore induce a positive amount of SA by design. The second group of IOLs uses aspheric optical surfaces with zero asphericity to eliminate this intrinsic SA of the IOL. These IOLs are referred to as “aberration neutral IOLs”. The third group of IOLs employ aspheric surfaces as well, but are designed to induce a negative amount of SA in order to (partially or fully) compensate for the cornea’s SA, referred to as “aberration correcting IOLs”. This third group of lenses was described by Holladay in 2002 [1] and has been demonstrated to reduce ocular SA to near-zero and significantly improve the contrast sensitivity at distance compared to standard spherical monofocal IOLs[2,3,4,5].

Recently, a new generation of enhanced monofocal IOLs has been introduced to slightly increase the depth of focus as compared to standard monofocal IOLs; these include the TECNIS Eyhance, model ICB00 (Johnson & Johnson Vision, Santa Ana, USA)[6], the RayOne EMV (Rayner Intraocular Lenses Limited, Worthing, UK), and ISOPure 1.2.3 (PhysIOL S.A, Liege, Belgium)[7]. In addition, currently, there are numerous IOL designs available in the market that use refractive technologies to extend the depth of focus. These refractive technologies are designed to improve uncorrected reading vision without the typical side effects (e.g. halo, starbursts, and glare) of diffractive concepts. However, although refractive designs might bring some advantages to reduce photic phenomena, they are also associated with more pupil dependency, and distance image quality reduction, than diffractive designs.

In this study, we used optical bench testing to evaluate the effect of pupil size on through-focus optical performance and distance image quality of three new-generation enhanced monofocal IOLs. The results were compared to those of standard monofocal IOLs with different levels of SA and chromatic aberration.

Methods

A total of six IOL models, with different optical designs and materials, were evaluated as listed in Table 1. The TECNIS Eyhance, model ICB00, and the ISOPure 1.2.3 use non-diffractive technologies and are defined by polynomial-based surfaces that are designed to increase depth of focus with minimum compromises at distance. The RayOne EMV is an aspheric monofocal IOL with positive spherical aberration designed to provide improved intermediate vision when implanted with 1.0D offset (monovision). These three IOL models were selected to provide a wide representation of the new category of enhanced monofocal IOLs, with different optical designs and materials. Additionally, to illustrate the expected performance of the standard monofocal IOLs, two aspheric and one spherical IOL were also included in the analysis: two aspheric (TECNIS Monofocal, model ZCB00, and the Acrysof IQ Monofocal, model SN60WF) designed to fully or partially compensate corneal spherical aberration and one spherical IOL (the Sensar 1-pc, model AAB00).

Modulation transfer function (MTF) and the phase transfer function (PTF) were measured in water at room temperature with white light and for a wide range of pupil diameters (2, 3, 4, and 5 mm). The IOLs were measured in a custom made physical model eye [8] that follows the a ISO 11979-2:2014 model eye 2. This eye model was selected because it resembles the spherical aberration of the average cornea (0.28 micrometers for a 6 mm entrance pupil as reported by Wang et al. [9]) as well as the average chromatic aberration of the human cornea (~ 1D for chromatic aberration between 450 and 650 nm). In order to derive a clinically meaningful metric for the image quality, the MTF and PTF measurements were used to calculate the simulated binocular visual acuity (sVA) using weighted Optical Transfer Function (wOTF) metric and the methodology described by Alarcon et al. [10]. Because the coefficients to convert wOTF to simulated VA were only provided for a 3 mm pupil only, sVA for other pupil sizes (2, 4 and 5 mm) was provided relative to the standard aspheric monofocal IOL, model ZCB00. This was required to balance the impact of the coefficients for the different pupil sizes.

To simulate the effect of different lighting conditions, the best focus position (0D) was found for an average photopic pupil of 3 mm was maintained during the measurement at the other different pupil sizes.

Table 1 Intraocular lens models under evaluation

Results

Simulated visual acuity

Figure 1 shows the defocus curves based on simulated binocular VA for the 6 different lenses for an average photopic pupil of 3 mm. As Fig. 1 illustrates, RayOne EMV provides a range of vision comparable to that of a standard monofocal IOL (AAB00, ZCB00 or SN60WF), while ICB00 and ISOpure provide an improved range of vision, especially around − 1.5D.

Fig. 1
figure 1

Simulated binocular VA defocus curves of different monofocal IOLs for 3 mm pupil

The effect of pupil size on sVA at far (0 D) and intermediate (-1.5 D) defocus points, relative to the standard aspheric monofocal IOL ZCB00, is shown in Figs. 2 and 3 respectively. As expected, the standard aspheric monofocal SN60WF, which partially compensates for corneal spherical aberration, has similar sVA performance as ZCB00 with an average difference below 0.05 logMAR for intermediate and distance sVA. The standard spherical monofocal AAB00 and the new generation enhanced monofocals, ISOPure and RayOne EMV, showed a similar correlation between a drop in far sVA and improvement in intermediate sVA for larger pupils. The enhanced monofocal ICB00 provided the lowest pupil variation, with no difference with respect to ZCB00 at far, and 0.1 logMAR improvement at intermediate for all pupil sizes.

Fig. 2
figure 2

Simulated VA at far (0D) with respect to the aspheric monofocal IOL ZCB00 from 5 to 2 mm pupil. Negative values indicate worse VA than the aspheric monofocal IOL.

Fig. 3
figure 3

Simulated VA at intermediate (-1.5D) with respect to the aspheric monofocal IOL ZCB00 from 5 to 2 mm pupil. Positive values indicate better VA than the aspheric monofocal IOL.

Distance image quality

Figure 4 provides the ranges of the MTF at 50 c/mm between 2 and 5 mm pupils for the 6 different IOL models. The difference between the minimum and the maximum MTF provides an estimation of the variability in distance image contrast with the pupil size. This variation between the maximum and minimum MTF values of the three standard monofocal IOLs, ZCB00, SN60WF and AAB00, was directly related to the level of corneal spherical aberration correction provided by each of the designs (see Table 1). The spherical monofocal IOL, model AAB00, and the new generation enhanced IOLs (ISOpure and RayOne EMV) provided the highest pupil dependency in distance image quality, with a relative drop of more than 50% from the maximum to the minimum MTF level. The aspheric monofocal IOL, model ZCB00, and the new generation enhanced monofocal ICB00 provided the lowest variability with pupil size and highest minimum MTF values.

Fig. 4
figure 4

Minimum (Min) and maximum (Max) MTF at 50 c/mm at far in the pupil range between 2 and 5 mm. Minimum MTF for all IOLs was for 5 mm pupil. Maximum MTF was found for different pupil sizes depending on the IOL.

As illustrated in Fig. 5A, models AAB00, ISOPure and RayOne EMV also provided the lowest MTF levels for a standard mesopic pupil (5 mm), with large losses in image contrast specially for lower spatial frequencies. For a standard photopic pupil of 3 mm (Fig. 5B), all IOLs provided similar MTF values except for model ZCB00, which has the highest MTF levels at all spatial frequencies.

Fig. 5
figure 5

Through frequency MTF curves at far (0D) for 5 mm (A) and 3 mm (B) pupil

Discussion

Optical bench results provide a good estimation of the clinical performance of intraocular lenses when performed in clinically relevant conditions (cornea that resembles the spherical and chromatic aberration of the real cornea and white light) and using clinically relevant pre-clinical metrics[10, 13,14,15]. The additional value of optical bench studies, compared to clinical investigations, is that they provide a direct comparison under the exact same conditions between different IOL models without the intrinsic variability associated with clinical studies. In this study, the effect of the pupil size was evaluated for 3 standard monofocal IOLs (2 aspheric aberration correcting models and one spherical model) and 3 new generation monofocal IOLs with different levels of spherical aberration and different chromatic performance.

While SA is the most important monochromatic aberration in the human eye, chromatic aberrations also play an important role for optical performance, especially in the pseudophakic eye, and there are significant differences in the chromatic dispersion of current materials used for IOLs[11, 16]. To account for this, measurements were performed in white light, using an eye model with the same average spherical and chromatic aberration of the average cornea.

The results of this optical bench study show that the new generation monofocal IOL, model ICB00, provided an extended range of focus and an improvement in simulated intermediate VA relative to standard aspheric monofocal IOL, while maintaining the same distance sVA. This is in agreement with the outcomes of the meta-analysis study of Wan et al. [17] summarizing the results of 12 different clinical studies comparing ICB00 to standard monofocal IOLs that found that ICB00 provides an improvement in unaided intermediate vision with similar distance performance. Other clinical studies have shown a significant improvement in intermediate VA in patients implanted with the ICB00 IOL compared to the aspheric monofocal lens ZCB00 but no significant differences at distance [18,19,20,21,22,23,24,25]. Cinar et al. compared clinical results of the SN60WF with the ICB00 and found superior VA for intermediate vision for the ICB00 and no significant differences at distance and near VA [26]. This consistent clinical performance of the ICB00 can be explained by the pupil independent design of this IOL model.

Stodulka et al. [7] reported clinical results on the ISOPure IOL. Although no control lens was used on this study, the mean binocular DCIVA of the ISOPure IOL was 0.2 logMAR, which is between 0.1 and 0.05 logMAR lower than the DCIVA reported for the ICB00 in different studies [18,19,20] and in good agreement with pre-clinical simulations. In contrast, the distance VA was slightly better for the ISOPure (-0.09 ± 0.06 logMAR) compared to the results reported for the ICB00 (ranging between − 0.06 ± 0.01 logMAR [19] and 0.03 ± 0.09 logMAR [23]).

Models ICB00 and ZCB00 provided consistently high MTF levels as expected, as both models fully compensate for the SA of the average cornea, and the effect of correcting spherical aberration is especially evident with large pupils [6]. Vega et al. [27] demonstrated that the ICB00 and ZCB00 showed comparable MTF levels and pupil independent optical performance, which supports our findings. This is in alignment with the clinical results of Mencucci et al. that did not find significant differences between patients implanted with ICB00 and ZCB00 when using objective image quality metrics like the Strehl ratio and the MTF cut-off frequency [18]. Additionally, Mencucci et al. [18] and Auffarth et al. [19] also measured the same contrast sensitivity in ICB00 and the ZCB00 patients.

The other new generation monofocal IOLs, ISOPure and RayOne EMV, provided MTF values at 5 mm comparable to the spherical monofocal IOL AAB00. Labuz et al. [12] compared four different refractive EDOF IOLs to the ZCB00 in an optical-bench study, including ICB00, ISOPure, RayOne-EMV, and the AE2UV/ZOE (an aspheric IOL with increased aberrations in the central part of the lens). They also found lower MTF values at 4.5 mm for the ISOPure and RayOne EMV than ICB00. The image performance for distance vision and the halo/glare pattern of the ICB00 and the AE2UV/ZOE was found to be comparable to that of the ZCB00. The ICB00 showed best overall area under the simulated defocus curve, followed by the AE2UV/ZOE and the ISOPure. The halo/glare pattern for the ISOPure and the RayOne EMV were found to be significantly larger than for the ZCB00. Although halo and scatter were not evaluated in our study, previous optical bench studies have also shown no differences between ICB00 and ZCB00 [6, 27].

The effect of pupil size on distance image quality is affected by the metric used and the light conditions. Monochromatic light should not be used to compare the performance of different intraocular lenses made of different materials, as it underestimates the effect of chromatic aberration on the image quality [28]. Additionally, although a single spatial frequency of MTF can provide a relative estimation of the contrast sensitivity expected in the patients, it does not correlate with perceived visual acuity and therefore is not a good metric to predict through-focus performance. There are numerous integral metrics that were developed to either correlate with visual acuity, contrast sensitivity, depth of focus or other visual performance scores [29,30,31,32]. Some of them incorporate the estimated neural transfer function, and others can be differentiated between monochromatic and polychromatic metrics. Additionally, the coefficients to convert the metrics to visual acuity were found for an average pupil size of 3 mm using different intraocular lens designs (monofocal, multifocal and extended depth of focus IOLs) made in different materials. Therefore, to evaluate other pupil sizes, these coefficients may not provide a good estimation of the expected visual acuity. For that reason, in our study absolute values of simulated visual acuity were only provided for 3 mm pupil, and for other pupil sizes a relative comparison was performed. The choice of the proper metric and the pupil dependency of the coefficients are therefore important and must be mentioned as a potential limitation of the current study. Another limitation of this study is that measurements were collected for an eye model with the average corneal spherical aberration. However, corneal spherical aberration is different for every eye and the outcomes of the IOL will change with corneal spherical aberration of the patient. Future studies could investigate the impact of different levels of corneal spherical aberration (or other aberrations). In addition, future studies should also consider the impact that other factors, like tilt and decentration of the IOL, can have in the performance of these designs.

Conclusion

New generation monofocal IOLs provide an improvement in depth of focus as compared to standard monofocal IOLs. The effect of pupil size on through focus performance and distance image quality depends on the IOL design and the level of correction of corneal SA. Distance MTF was mainly reduced for larger pupil sizes in monofocal IOLs that do not fully compensate for corneal spherical aberration, such as the new generation monofocal IOL ISOPure and RayOne EMV and the spherical monofocal IOL AAB00. On the other hand, designs that fully compensate for corneal SA such as model ICB00 and ZCB00, showed the lowest variability with the pupil for both distance (simulated visual acuity and MTF) and intermediate sVA.

Data availability

All the information and figures available for publication is contained in this manuscript. Raw data won’t be disclosured. For more data/information related to the data, contact the corresponding author: Aixa Alarcon (aalarco4@its.jnj.com).

References

  1. Holladay JT, Piers PA, Koranyi G, van der Mooren M, Norrby NES. A new intraocular lens design to reduce spherical aberration of pseudophakic eyes. J Refract Surg. 2002;18:683–91. https://doi.org/10.3928/1081-597X-20021101-04.

    Article  PubMed  Google Scholar 

  2. Bellucci R, Scialdone A, Buratto L, Morselli S, Chierego C, Criscuoli A, Moretti G, Piers P. Visual acuity and contrast sensitivity comparison between Tecnis and AcrySof SA60AT intraocular lenses: a multicenter randomized study. J Cataract Refract Surg. 2005;31:712–7. https://doi.org/10.1016/j.jcrs.2004.08.049.

    Article  PubMed  Google Scholar 

  3. Denoyer A, Le Lez M-L, Majzoub S, Pisella P-J. Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation: effect of contrast sensitivity and wavefront aberration improvements on the quality of daily vision. J Cataract Refract Surg. 2007;33:210–6. https://doi.org/10.1016/j.jcrs.2006.10.035.

    Article  PubMed  Google Scholar 

  4. Tzelikis PF, Akaishi L, Trindade FC, Boteon JE. Spherical aberration and contrast sensitivity in eyes implanted with aspheric and spherical intraocular lenses: a comparative study. Am J Ophthalmol. 2008;145:827–33. https://doi.org/10.1016/j.ajo.2007.12.023.

    Article  PubMed  Google Scholar 

  5. Mester U, Dillinger P, Anterist N. Impact of a modified optic design on visual function: clinical comparative study. J Cataract Refract Surg. 2003;29:652–60. https://doi.org/10.1016/S0886-3350(02)01983-1.

    Article  PubMed  Google Scholar 

  6. Alarcon A, Cánovas C, Koopman B, Weeber H, Auffarth GU, Piers PA. Enhancing the Intermediate Vision of Monofocal intraocular lenses using a higher order Aspheric Optic. J Refract Surg. 2020;36:520–7. https://doi.org/10.3928/1081597X-20200612-01.

    Article  PubMed  Google Scholar 

  7. Stodulka P, Slovak M. (2021)Visual Performance of a Polynomial Extended Depth of Focus Intraocular Lens. OJOph; 11:214–228. https://doi.org/10.4236/ojoph.2021.113017

  8. Norrby S, Piers P, Campbell C, van der Mooren M. (2007) Model eyes for evaluation of intraocular lenses. Appl Opt 2007; 46:6595–6605

  9. Wang L, Dai E, Koch DD, Nathoo A. A optical aberrations of the human anterior cornea. J Cataract Refract Surg. 2003;29(8):1514–21. https://doi.org/10.1016/S0886-3350(03)00467-X.

    Article  PubMed  Google Scholar 

  10. Alarcon A, Canovas C, Rosen R, Weeber H, Tsai L, Hileman K, Piers P. Preclinical metrics to predict through-focus visual acuity for pseudophakic patients. Biomed Opt Express. 2016;7:1877–88. https://doi.org/10.1364/BOE.7.001877.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Eppig T, Rawer A, Hoffmann P, Langenbucher A, Schröder S. (2020) On the Chromatic Dispersion of Hydrophobic and Hydrophilic Intraocular Lenses. Optom Vis Sci 2020; 97:305–313. https://doi.org/10.1097/OPX.0000000000001495

  12. Łabuz G, Son H-S, Naujokaitis T, Yildirim TM, Khoramnia R, Auffarth GU. Laboratory Investigation of Preclinical Visual-Quality Metrics and Halo-Size in enhanced Monofocal intraocular lenses. Ophthalmol therapy. 2021;10(4):1093–104. https://doi.org/10.1007/s40123-021-00411-9.

    Article  Google Scholar 

  13. Weeber HA, Piers PA. Theoretical performance of intraocular lenses correcting both spherical and chromatic aberration. J Refract Surg. 2012;28:48–52. https://doi.org/10.3928/1081597X-20111103-01.

    Article  PubMed  Google Scholar 

  14. Weeber HA, Cánovas C, Alarcón A, Piers PA. Laboratory-Measured MTF of IOLs and clinical performance. J Refract Surg. 2016;32:211–2. https://doi.org/10.3928/1081597X-20160125-01.

    Article  PubMed  Google Scholar 

  15. Piers PA, Chang DH, Alarcón A, Cánovas C. Clinically relevant Interpretations of Optical Bench Measurement of intraocular lenses. J Refract Surg. 2017;33:64. https://doi.org/10.3928/1081597X-20161018-04.

    Article  PubMed  Google Scholar 

  16. Nakajima M, Hiraoka T, Yamamoto T, Takagi S, Hirohara Y, Oshika T, Mihashi T. Differences of Longitudinal Chromatic Aberration (LCA) between eyes with intraocular lenses from different manufacturers. PLoS ONE. 2016;11:e0156227. https://doi.org/10.1371/journal.pone.0156227.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  17. Wan KH, Au AC, Kua WN, Ng AL, Cheng GP, Lam NM, Chow VW. Enhanced Monofocal Versus Conventional Monofocal intraocular Lens in cataract surgery: a Meta-analysis. J Refract Surg. 2022;38(8):538–46. https://doi.org/10.3928/1081597X-20220707-01.

    Article  PubMed  Google Scholar 

  18. Mencucci R, Cennamo M, Venturi D, Vignapiano R, Favuzza E. Visual outcome, optical quality, and patient satisfaction with a new monofocal IOL, enhanced for intermediate vision: preliminary results. J Cataract Refract Surg. 2020;46:378–87. https://doi.org/10.1097/j.jcrs.0000000000000061.

    Article  PubMed  Google Scholar 

  19. Auffarth GU, Gerl M, Tsai L, Janakiraman DP, Jackson B, Alarcon A, Dick HB. Clinical evaluation of a new monofocal IOL with enhanced intermediate function in patients with cataract. J Cataract Refract Surg. 2021;47:184–91. https://doi.org/10.1097/j.jcrs.0000000000000399.

    Article  PubMed  Google Scholar 

  20. de Luis Eguileor B, Martínez-Indart L, Alday NM, Egüén CS, Sánchez CC. (2020) Differences in intermediate vision: monofocal intraocular lenses vs. monofocal extended depth of focus intraocular lenses. Archivos de la Sociedad Española de Oftalmología (English Edition) 1;95(11):523-7. https://doi.org/10.1016/j.oftale.2020.06.001

  21. Kang KH, Song MY, Kim KY, Hwang K-Y, Kwon Y-A, Koh K. Visual performance and Optical Quality after Implantation of a New Generation Monofocal intraocular Lens. Korean J Ophthalmol. 2021;35:112–9. https://doi.org/10.3341/kjo.2020.1115.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Huh J, Eom Y, Yang SK, Choi Y, Kim HM, Song JS. A comparison of clinical outcomes and optical performance between monofocal and new monofocal with enhanced intermediate function intraocular lenses: a case-control study. BMC Ophthalmol. 2021;21:365. https://doi.org/10.1186/s12886-021-02124-w.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Lopes D, Loureiro T, Carreira R, Rodrigues Barros S, Cardoso JN, Campos P, Machado I, Campos N. Comparative evaluation of visual outcomes after bilateral implantation of an advanced or conventional monofocal intraocular lens. Eur J Ophthalmol. 2021;2021:1120672121995343. https://doi.org/10.1177/1120672121995343.

    Article  Google Scholar 

  24. Corbelli E, Iuliano L, Bandello F, Fasce F. Comparative analysis of visual outcome with three intraocular lenses: monofocal, enhanced monofocal, and extended range of vision. J Cataract Refract Surg. 2021;48(1):67–74. https://doi.org/10.1097/j.jcrs.0000000000000706.

    Article  Google Scholar 

  25. Lee JH, Moon SY, Chung HS, Park SY, Lee H, Kim JY, Tchah H. (2021) Clinical Outcomes of New Monofocal Intraocular Lenses with Enhanced Intermediate Function Compared with Extended Depth of Focus Intraocular Lenses. J Cataract Refract Surg. https://doi.org/10.1097/j.jcrs.0000000000000710

  26. Cinar E, Bolu H, Erbakan G, Yuce B, Aslan F, Fece M, Emre S. Vision outcomes with a new monofocal IOL. Int Ophthalmol. 2021;41:491–8. https://doi.org/10.1007/s10792-020-01599-8.

    Article  PubMed  Google Scholar 

  27. Vega F, Millán MS, Gil MA, Garzón N. Optical performance of a Monofocal intraocular Lens designed to extend depth of Focus. J Refract Surg. 2020;36:625–32. https://doi.org/10.3928/1081597X-20200710-01.

    Article  PubMed  Google Scholar 

  28. Zhao H, Mainster MA. The effect of chromatic dispersion on pseudophakic optical performance. Br J Ophthalmol. 2007;91:1225–9.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Cheng X, Bradley A, Thibos LN. Predicting subjective judgment of best focus with objective image quality metrics. J Vis. 2004;4:310–21. https://doi.org/10.1167/4.4.7.

    Article  PubMed  Google Scholar 

  30. Michael R, Guevara O, La Paz M, Alvarez de Toledo J, Barraquer RI. Neural contrast sensitivity calculated from measured total contrast sensitivity and modulation transfer function. Acta Ophthalmol. 2011;89:278–83. https://doi.org/10.1111/j.1755-3768.2009.01665.x.

    Article  PubMed  Google Scholar 

  31. Ravikumar S, Thibos LN, Bradley A. Calculation of retinal image quality for polychromatic light. J Opt Soc Am A Opt Image Sci Vis. 2008;25:2395–407. https://doi.org/10.1364/JOSAA.25.002395.

    Article  PubMed  Google Scholar 

  32. Leube A, Schilling T, Ohlendorf A, Kern D, Ochakovski AG, Fischer MD, Wahl S. Individual neural transfer function affects the prediction of subjective depth of focus. Sci Rep. 2018;8:1919. https://doi.org/10.1038/s41598-018-20344-x.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This investigation was supported by Johnson and Johnson Vision.

Author information

Authors and Affiliations

Authors

Contributions

AA: conception and design, acquisition of data, analysis and interpretation of data, drafting and revising the manuscript and final approval.

CC: conception and design, analysis and interpretation of data, drafting and revising the manuscript and final approval.

BK: acquisition of data, analysis of data, drafting and revising the manuscript.

MP, DK and PP: analysis and interpretation of data, revising the manuscript and final approval.

Corresponding author

Correspondence to Aixa Alarcon.

Ethics declarations

Ethics Approval and Consent to Participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

AA, CC, BK and PP: employees of Johnson and Johnson Vision.

MVP and DDK : consultant of Johnson and Johnson Vision.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Alarcon, A., Canovas, C., Koopman, B. et al. Optical bench evaluation of the effect of pupil size in new generation monofocal intraocular lenses. BMC Ophthalmol 23, 112 (2023). https://doi.org/10.1186/s12886-023-02839-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-023-02839-y

Keywords