Study | Number of eyes / patients | Technical Failure Rate [n] (Device) | ||
---|---|---|---|---|
PCI | LCOR | SS-OCT | ||
Srivannaboon et al. 2015 [24] | 100/100 | 5 (IOLMaster 500) | Â | 0 (IOLMaster 700) |
Shammas et al. 2016 [32] | 107/66 | 13 (IOLMaster 500) | 14 (Lenstar LS900) | 2 (Argos) |
Kurian et al. 2016 [33] | 100/100 | Â | 21 (Lenstar LS900) | 4 (IOLMaster 700) |
Akman et al. 2016 [34] | 188/101 | 17 (IOLMaster 500) | Â | 0 (IOLMaster 700) |
McAlinden et al. 2016 [13] | 377/210 | 136 (IOLMaster 500) | 51 (Aladdin) | 0 (OA-2000) |
Jung et al. 2017 [35] | 101/54 | 7 (Galilei G6) | Â | 1 (IOLMaster 700) |
Arriola-Villalobos et al. 2017 [36] | 80/80 | Â | 8 (Lenstar LS900) | 0 (IOLMaster 700) |
Higashiyama et al. 2018 [37] | 55/55 | 7 (IOLMaster 500) | Â | 1 (Argos) |
Lee and Kim 2018 [38] | 175/175 | 25 (IOLMaster 500) | Â | 2 (IOLMaster 700) |
An et al. 2019 [39] | 431/431 | 66 (IOLMaster 500) | Â | 10 (Argos) |
Huang et al. 2019 [40] | 171/119 | 33 (IOLMaster 500) | Â | 1 (Argos) 5 (IOLMaster 700) 5 (OA-2000) |
Yang et al. 2019 [41] | 146/83 | 17 (IOLMaster 500) | Â | 3 (Argos) 3 (IOLMaster 700) |
El Chebab et al. 2019 [42] | 129/129 | Â | 1 (Lenstar LS900) | 1 (IOLMaster 700) |
Cummings et al. 2020 [31] | 299/NA | Â | 4 (Lenstar LS900) | 0 (Argos) |