Technical failure rates for biometry between swept-source and older-generation optical coherence methods: a review and meta-analysis

Abstract

Purpose

Precise ocular measurements are fundamental for achieving excellent target refraction following both cataract surgery and refractive lens exchange. Biometry devices with swept-source optical coherence tomography (SS-OCT) employ longer wavelengths (1055–1300 nm) in order to have better penetration through opaque lenses than those with partial coherence interferometry (PCI) or low-coherence optical reflectometry (LCOR) methods. However, to date a pooled analysis showing the technical failure rate (TFR) between the methods has not been published. The aim of this study was to compare the TFR in SS-OCT and in PCI/LCOR biometry.

Methods

PubMed and Scopus were used to search the medical literature as of Feb 1, 2022. The following keywords were used in various combinations: optical biometry, partial coherence interferometry, low-coherence optical reflectometry, swept-source optical coherence tomography. Only clinical studies referring to patients undergoing routine cataract surgery, and employing at least two (PCI or LCOR vs. SS-OCT) optical methods for optical biometry in the same cohort of patients were included.

Results

Fourteen studies were included in the final analysis, which presented results of 2,459 eyes of at least 1,853 patients. The overall TFR of all included studies was 5.47% (95% confidence interval [CI]: 3.66–8.08%; overall I2 = 91.49%). The TFR was significantly different among the three methods (p < 0.001): 15.72% for PCI (95% CI: 10.73–22.46%; I2 = 99.62%), 6.88% for LCOR (95% CI: 3.26–13.92%; I2 = 86.44%), and 1.51% for SS-OCT (95% CI: 0.94–2.41%; I2 = 24.64%). The pooled TFR for infrared methods (PCI and LCOR) was 11.12% (95% CI: 8.45–14.52%; I2 = 78.28%), and was also significantly different to that of SS-OCT: 1.51% (95% CI: 0.94–2.41%; I2 = 24.64%; p < 0.001).

Conclusions

A meta-analysis of the TFR of different biometry methods highlighted that SS-OCT biometry resulted in significantly decreased TFR compared to PCI/LCOR devices.

Peer Review reports

Introduction

Precise measurement of ocular structure distances is essential for excellent refractive outcomes, both in cataract surgery and refractive lens exchange. More than twenty years ago the first optical biometer (IOLMaster®, Carl Zeiss Meditec, Jena, Germany) was introduced; since then several commercially-available instruments have been developed. These devices might employ partial coherence interferometry (PCI), low-coherence optical reflectometry (LCOR), and swept-source optical coherence tomography (SS-OCT) [1,2,3,4]. Currently, optical biometry is considered as the gold standard for preoperative biometry.

The first intraocular lens (IOL) calculations formulas were based on three variables that could be extracted from biometry data: axial length, corneal refractive power and the predicted postoperative anterior chamber depth. Newer-generation formulas employ additional parameters to estimate the postoperative effective lens position [5]. Even further modifications are still being made to vergence-based IOL formulas for improved lens power accuracy. The standard deviation of the new-generation IOL calculation formulas in patients undergoing cataract surgery reaches 0.4 D [6], which translates into more than 80% of eyes within 0.5 D target refraction [7]. A single study has even reported ≥ 88% of eyes within 0.5 D target refraction [8]. Not only does this lead to excellence in outcomes, but also increases patient expectations. This is particularly important due to the constantly growing popularity for spectacle-independence with non-toric and toric premium IOLs.

SS-OCT devices employ longer wavelengths (1055–1300 nm), so they are supposed to have better penetration through opaque media than PCI/LCOR devices [9, 10]. However, to date, a pooled analysis showing the technical failure rate (TFR) in these methods has not been published. The aim of this study was to compare the TFR in SS-OCT biometry and PCI/LCOR.

Methods

PubMed and Scopus were the main resources used to search the medical literature. An extensive look up was performed to identify relevant articles concerning the TFRs in optical biometry as of Feb 1, 2022. The following keywords were used in various combinations: optical biometry, partial coherence interferometry, low-coherence optical reflectometry, swept-source optical coherence tomography. Of the studies retrieved by this method, we reviewed all publications in English and abstracts of non-English publications [11]. The search did not aim to find studies designed to report TFR in healthy individuals, [12] but rather to report the rate of technical or capture error in currently published cataract surgery studies. A study was included in this meta-analysis if: (i) it was referring to patients undergoing phacoemulsification cataract surgery, (ii) presented results of a general population, but not only for a subgroup of patients with mature or white cataracts (iii) the study employed at least two or more optical methods for optical biometry in the same cohort of patients: one of them was SS-OCT and the other either PCI, LCOR or both, (iiii) the TFRs for each method used was reported with a 95% confidence interval (CI), or included data to calculate them. Studies were critically reviewed to create an overview and guidance for further search. No attempts to discover unpublished data were made. The search strategy in detail is presented in Supplement 1

TFR was defined as failure in obtaining ocular distance measurements. Such failure could be a result of the presence of dense cataracts, but also posterior subcapsular cataract, macular diseases, staphyloma, vitreous or corneal opacities, and poor fixation [13]. The reason for failure was not relevant as commonly it was not presented in the analyzed study and this study aimed to show real-life results. Devices classified as PCI biometers included the IOLMaster 500 (Carl Zeiss Meditec, Jena, Germany), Galilei G6 (Ziemer Ophthalmic Systems AG, Brügg, Switzerland) and the Pentacam AXL/AXL Wave (OCULUS Optikgeräte GmbH, Wetzlar, Germany), while LCOR biometers were the Lenstar LS900 (Haag-Streit AG, Köniz, Switzerland) and Aladdin (Topcon Corporation, Tokyo, Japan) [14, 15]. SS-OCT biometers which are currently on the market include the IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany), OA-2000 (Tomey Corporation, Nagoya, Japan), Argos (Alcon Laboratories, Inc., Fort Worth, TX, USA), ANTERION (Heidelberg Engineering, Heidelberg, Germany) and Eyestar 900 (Haag-Streit AG, Köniz, Switzerland).

Statistical analysis was performed using Medcalc v. 20.027 (Medcalc Software LTD, Ostend, Belgium). If CIs were not provided within the study, the Wilson method for calculating CIs was applied, as it provides more reliable results than normal approximation with symmetrical confidence intervals [16]. The Freeman–Tukey transformation was employed to calculate the weighted summary proportion. Corresponding heterogeneity (I2) estimates were generated, and the I2 estimates equal to 25%, 50%, and 75%, were considered as indicating the presence of low, moderate, and high heterogeneity, respectively. As studies did not come from a common population, random effects models were applied. The DerSimonian and Laird approach was employed to calculate the summary proportions under the random effects model. Pooled-analysis was used in order to avoid problems arising from simple pooling. Results with p levels under 0.05 were considered statistically significant.

Results

The search identified 1,860 eligible publications from the PubMed and Scopus databases; the strategy in detail is presented in Fig. 1 (study chart flow). Several studies have been excluded as they have compared results of optical biometry in healthy but not in cataract patients [17,18,19,20,21,22], did not compare outcomes of PCI/LCOR with SS-OCT devices [23,24,25,26,27,28,29] or both [30]. Finally, fourteen studies were included in the final analysis (Table 1), which presented results of 2,459 eyes of at least 1,853 patients (since in one study [31] only the number of eyes, and not patients was reported).

Fig. 1
figure 1

Flow diagram outlining the selection process of the systematic review for on swept-source and older-generation optical coherence methods

Table 1 Comparative studies showing the technical failure rate for three different optical biometry technologies in patients undergoing routine cataract surgery

The overall pooled TFR of all included studies was 5.47% (95% CI: 3.66–8.08%; overall I2 = 91.49%). The TFR was significantly different among the three methods (p < 0.001): 15.72% for PCI (95% CI: 10.73–22.46%; I2 = 99.62%), 6.88% for LCOR (95% CI: 3.26–13.92%; I2 = 86.44%), and 1.51% for SS-OCT (95% CI: 0.94–2.41%; I2 = 24.64%).

The TFR for infrared methods (LCOR and PCI) was 11.12% (8.45–14.52%; I2 = 78.28%), and was also significantly different to that of SS-OCT: 1.51% (95% CI: 0.94–2.41%; I2 = 24.64%; p < 0.001). A pooled analysis of the included studies is shown in Fig. 2. The funnel plots for PCI/LCOR (Fig. 3) and for SS-OCT (Fig. 4) devices highlighted that several studies regarding PCI/LCOR were situated outside the funnel which confirmed the evidently greater heterogeneity of these studies.

Fig. 2
figure 2

Forest plot showing the technical failure rate in partial coherence interferometry and low-coherence optical reflectometry vs. swept-source optical coherence tomography biometers. Abbreviation: LCOR - low-coherence optical reflectometry, PCI - partial coherence interferometry, SS-OCT - swept-source optical coherence tomography

Fig. 3
figure 3

Funnel plot for partial coherence interferometry and low-coherence optical reflectometry biometers

Fig. 4
figure 4

Funnel plot for swept-source optical coherence tomography biometers

Discussion

In this review and meta-analysis, the TFR was estimated and compared between different optical biometry methods. Results show that the TFR for infrared methods (LCOR and PCI) used in devices such as the IOLMaster 500, Lenstar LS900 or Aladdin were significantly higher than those using SS-OCT technology, such as the IOLMaster 700, Argos or OA-2000. Not only was the TFR lower among SS-OCT based devices (1.51% vs. 11.12%) but the heterogeneity in reported studies evidently was lower. The highest TFR was seen with devices based on PCI, followed by LCOR based devices, with the lowest rates seen with SS-OCT.

Achieving excellent refractive outcomes following surgery is becoming a common expectation for many patients; in order to achieve this, accurate biometry is essential. Optical biometry is considered superior to other methods, and a high TFR invariably leads to a higher percentage of patients requiring alternative methods to obtain biometry. This could have a negative influence on the refractive outcomes. The differences seen in this study between methods were clinically significant, as an increase from 1.5% to over 10% of patients with insufficient optical biometry can significantly decrease the expected refractive outcomes in any clinic. Considering the efforts routinely used to minimize any residual refraction, the difference seen here could be highly meaningful. The TFR of the biometer is usually unknown and the ocular distances obtained with different biometers cannot be considered interchangeable [18, 19]. Other aspects which are expected to be variable are corneal tomography or the results of IOL calculation made with different formulas; in such a case several measurements or calculations are often conducted and compared before a decision on IOL power is reached. IOL calculations using several biometers based on different technologies is not commonly performed, so the variability in TFR remains an uncontrolled factor that might influence the refractive results.

Phacoemulsification cataract surgery among patients with dense cataracts is associated with increased risk of intra- and postoperative complications e.g., risk of zonulopathy and posterior capsule rupture, greater corneal endothelial damage and prolonged operative time [14, 43,44,45,46]. Lower proportion of eyes with dense cataracts achieve excellent uncorrected and best-corrected visual acuity when compared to non-dense cataract [47, 48]. Thus, use of optical biometers with lower TFR might be especially important for patients with dense cataracts which are already prone to other undesirable outcomes. Furthermore, some of the current SS-OCT devices have a retinal visualization mode which can be utilized instead of ultrasound biometry; in this mode, the optical reflex from the retina is enhanced ten-fold [49]. A single study has shown that in cataracts grade IV or higher, according to the Emery-Little classification, the acquisition rates of SS-OCT might range from 63.6% (for IOLMaster 700) to 89.9% (for Argos) [29]. One might estimate that in these advanced cataract cases the TFR might be even higher in LCOR and particularly PCI biometers than in SS-OCT devices. Importantly, the study by Hirnschall et al. has shown that 91.3% of the eyes that were unsuccessfully scanned with PCI IOLMaster 500 were measurable with the SS-OCT IOLMaster 700 [12].

Our study has found a pooled 15.72% TFR (95% CI: 10.73–22.46%; I2 = 99.62%) for PCI biometers. This appears very high, keeping in mind that many clinicians are using PCI biometers without any problems on a daily basis; however, this might be dependent on the setting where the device is used and the grading of cataracts. The signal penetration even through a medium-dense cataract in PCI/LCOR devices is lower than with SS-OCT. Not only could this lead to technical failure, but result in a greater variability of the axial length measurements and intraocular lens power calculation predictability. Most of the analyzed studies have reported excellent repeatability and reproducibility in PCI, LCOR, and SS-OCT devices. However, in the study by Kurian et al. the within-subject coefficient of variation in axial length was significantly greater for the LCOR Lenstar LS900 than in the IOLMaster 700 (0.21% vs. 0.05%, respectively) [33]. Jung et al. has shown that the proportion of eyes with an absolute prediction error within 0.5 D was 80.0% for the PCI Galilei G6 and 85.0% for the SS-OCT IOLMaster 700 based on the SRK/T formula [35]. In the study by Yang et al. there was no difference in the percentage of eyes within ± 0.5 D postoperatively between the PCI IOLMaster 500, and SS-OCTs IOLMaster 700 and Argos [41]. The mean absolute prediction error was 0.41 ± 0.31 D, 0.42 ± 0.32 D, and 0.35 ± 0.30 D for IOLMaster 500, IOLMaster 700 and Argos, respectively [41]. Finally, the study by An et al. has shown a trend towards greater postoperative mean absolute error with the PCI IOLMaster 500 than the SS-OCT IOLMaster 700 (0.39 ± 0.30 D vs. 0.36 ± 0.27 D, respectively), but both of the methods outperformed A-scan ultrasound (0.47 ± 0.39 D) [39]. It might be concluded that in cases without a technical failure, the outcome of PCI/LCOR is not inferior to SS-OCT devices [24, 32, 33, 36, 40,41,42]. A recent review by Montés-Micó et al. reported excellent repeatability and reproducibility in SS-OCT biometers, but was not able to draw conclusions regarding the superiority of SS-OCT when compared to older biometers in terms of precision in IOL calculation [50].

The risk of bias assessment [Supplementary Table 1] represents that several of the studies did not explain in detail the reason for technical failure [13, 24, 31, 34, 36, 38,39,40]. There are some usability differences between the devices. The time required to acquire measurements might not only affect the patient experience, but could potentially influence the TFR. Although none of the analyzed studies has assessed the difference in acquisition time between the devices, Ruiz-Mesa et al. has shown that IOLMaster 700 measurements, whether they are performed with or without central topography, take less time to perform than older generation optical biometry and corneal tomography with a separate device [51]. It is beyond the scope of this manuscript, but with some devices it might be easier to obtain measurements, particularly if they are automatic rather than semi-automatic in use.

A significant limitation of the meta-analysis is that some studies included results for both eyes [13, 32, 34, 35, 40, 41], while others did not present the number of patients nor the laterality but just the number of eyes [31]. Using a combined measurement from both eyes is likely to be an underestimate of the true variance of a sample [52, 53]. It should also be acknowledged that no comparative studies have been published yet for some new SS-OCT biometers. For example, the ANTERION and the Eyestar 900 biometers have been released just recently (September 2018, and spring 2021, respectively).

In conclusion, in this meta-analysis of the TFR of different optical biometry methods, SS-OCT biometry provided significantly lower TFR compared to devices based on PCI and LCOR. These results could have significant implications on refractive outcomes and highlight the importance of variability in TFR between different devices.

Data Availability

The datasets generated and analysed during the current study available from the corresponding author on request.

References

  1. Savini G, Hoffer KJ, Shammas HJ, Aramberri J, Huang J, Barboni P. Accuracy of a new swept-source Optical Coherence Tomography Biometer for IOL Power calculation and comparison to IOLMaster. J Refract Surg. 2017;33:690–5.

    Article  PubMed  Google Scholar 

  2. Huang J, Savini G, Hoffer KJ, Chen H, Lu W, Hu Q, et al. Repeatability and interobserver reproducibility of a new optical biometer based on swept-source optical coherence tomography and comparison with IOLMaster. Br J Ophthalmol. 2017;101:493–8.

    Article  PubMed  Google Scholar 

  3. Muzyka-Woźniak M, Oleszko A. Comparison of anterior segment parameters and axial length measurements performed on a Scheimpflug device with biometry function and a reference optical biometer. Int Ophthalmol. 2019;39:1115–22.

    Article  PubMed  Google Scholar 

  4. Kanclerz P. Optical biometry in a commercially available anterior and posterior segment optical coherence tomography device. Clin Exp Optom. 2019;102:533–4.

    Article  PubMed  Google Scholar 

  5. Xia T, Martinez CE, Tsai LM. Update on intraocular Lens Formulas and Calculations. Asia Pac J Ophthalmol (Phila). 2020;9:186–93.

    Article  PubMed  Google Scholar 

  6. Melles RB, Holladay JT, Chang WJ. Accuracy of intraocular Lens calculation formulas. Ophthalmology. 2018;125:169–78.

    Article  PubMed  Google Scholar 

  7. Savini G, Taroni L, Hoffer KJ. Recent developments in intraocular lens power calculation methods-update 2020. Ann Transl Med. 2020;8:1553.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Savini G, Hoffer KJ, Balducci N, Barboni P, Schiano-Lomoriello D. Comparison of formula accuracy for intraocular lens power calculation based on measurements by a swept-source optical coherence tomography optical biometer. J Cataract Refract Surg. 2020;46:27–33.

    PubMed  Google Scholar 

  9. Povazay B, Hermann B, Unterhuber A, Hofer B, Sattmann H, Zeiler F, et al. Three-dimensional optical coherence tomography at 1050 nm versus 800 nm in retinal pathologies: enhanced performance and choroidal penetration in cataract patients. J Biomed Opt. 2007;12:041211.

    Article  PubMed  Google Scholar 

  10. Unterhuber A, Povazay B, Hermann B, Sattmann H, Chavez-Pirson A, Drexler W. In vivo retinal optical coherence tomography at 1040 nm - enhanced penetration into the choroid. Opt Express. 2005;13:3252–8.

    Article  PubMed  Google Scholar 

  11. Grzybowski A, Kanclerz P. Language Bias and Methodological issues in determining Reliable evidence for systematic reviews. JAMA Ophthalmol. 2019;137:118–9.

    Article  PubMed  Google Scholar 

  12. Hirnschall N, Varsits R, Doeller B, Findl O. Enhanced penetration for axial length measurement of eyes with dense cataracts using swept source Optical Coherence Tomography: a consecutive observational study. Ophthalmol Ther. 2018;7:119–24.

    Article  PubMed  PubMed Central  Google Scholar 

  13. McAlinden C, Wang Q, Gao R, Zhao W, Yu A, Li Y, et al. Axial length measurement failure Rates with Biometers using swept-source Optical Coherence Tomography compared to partial-coherence interferometry and optical low-coherence interferometry. Am J Ophthalmol. 2017;173:64–9.

    Article  PubMed  Google Scholar 

  14. Grzybowski A, Kanclerz P. Recent Developments in Cataract Surgery. Curr Concepts Ophthalmol. 2020;:55–97.

  15. Kanclerz P, Khoramnia R, Wang X. Current Developments in Corneal Topography and Tomography. Diagnostics (Basel). 2021;11.

  16. DasGupta A, Tony Cai T, Brown LD. Interval estimation for a binomial proportion. Stat Sci. 2001;16:101–33.

    Article  Google Scholar 

  17. Tu R, Yu J, Savini G, Ye J, Ning R, Xiong J, et al. Agreement between two Optical Biometers based on large coherence length SS-OCT and Scheimpflug Imaging/Partial coherence interferometry. J Refract Surg. 2020;36:459–65.

    Article  PubMed  Google Scholar 

  18. Kanclerz P, Hoffer KJ, Przewłócka K, Savini G. Comparison of an upgraded optical biometer with 2 validated optical biometers. J Cataract Refract Surg. 2021;47:859–64.

    Article  PubMed  Google Scholar 

  19. Kanclerz P, Hoffer KJ, Rozema JJ, Przewłócka K, Savini G. Repeatability and reproducibility of optical biometry implemented in a new optical coherence tomographer and comparison with a optical low-coherence reflectometer. J Cataract Refract Surg. 2019;45:1619–24.

    Article  PubMed  Google Scholar 

  20. Shajari M, Cremonese C, Petermann K, Singh P, Müller M, Kohnen T. Comparison of axial length, corneal curvature, and Anterior Chamber depth measurements of 2 recently introduced Devices to a known Biometer. Am J Ophthalmol. 2017;178:58–64.

    Article  PubMed  Google Scholar 

  21. Schiano-Lomoriello D, Hoffer KJ, Abicca I, Savini G. Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices. Sci Rep. 2021;11:983.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Kanclerz P, Hoffer KJ, Bazylczyk N, Wang X, Savini G. Optical Biometry and IOL calculation in a commercially available Optical Coherence Tomography device and comparison with Pentacam AXL. Am J Ophthalmol. 2023;246:236–41.

    Article  PubMed  Google Scholar 

  23. Hoffer KJ, Shammas HJ, Savini G, Huang J. Multicenter study of optical low-coherence interferometry and partial-coherence interferometry optical biometers with patients from the United States and China. J Cataract Refract Surg. 2016;42:62–7.

    Article  PubMed  Google Scholar 

  24. Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg. 2015;41:2224–32.

    Article  PubMed  Google Scholar 

  25. Panthier C, Rouger H, Gozlan Y, Moran S, Gatinel D. Comparative analysis of 2 biometers using swept-source OCT technology. J Cataract Refract Surg. 2022;48:26–31.

    Article  PubMed  Google Scholar 

  26. Dong J, Yao J, Chang S, Kanclerz P, Khoramnia R, Wang X. Comparison study of the two Biometers based on swept-source Optical Coherence Tomography Technology. Diagnostics. 2022;12:598.

    Article  PubMed  PubMed Central  Google Scholar 

  27. McAlinden C, Wang Q, Pesudovs K, Yang X, Bao F, Yu A, et al. Axial length measurement failure rates with the IOLMaster and Lenstar LS 900 in eyes with cataract. PLoS ONE. 2015;10:e0128929.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Sabatino F, Matarazzo F, Findl O, Maurino V. Comparative analysis of 2 swept-source optical coherence tomography biometers. J Cataract Refract Surg. 2019;45:1124–9.

    Article  PubMed  Google Scholar 

  29. Tamaoki A, Kojima T, Hasegawa A, Yamamoto M, Kaga T, Tanaka K, et al. Clinical evaluation of a new swept-source Optical Coherence Biometer that uses individual refractive indices to measure axial length in cataract patients. Ophthalmic Res. 2019;62:11–23.

    Article  PubMed  Google Scholar 

  30. Pardeshi AA, Song AE, Lazkani N, Xie X, Huang A, Xu BY. Intradevice repeatability and interdevice agreement of ocular biometric measurements: a comparison of two swept-source anterior segment OCT Devices. Transl Vis Sci Technol. 2020;9:14.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Cummings AB, Naughton S, Coen A-M, Brennan E, Kelly GE. Comparative analysis of swept-source optical coherence tomography and partial coherence Interferometry Biometers in the prediction of cataract surgery refractive outcomes. Clin Ophthalmol. 2020;14:4209–20.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Shammas HJ, Ortiz S, Shammas MC, Kim SH, Chong C. Biometry measurements using a new large-coherence-length swept-source optical coherence tomographer. J Cataract Refract Surg. 2016;42:50–61.

    Article  PubMed  Google Scholar 

  33. Kurian M, Negalur N, Das S, Puttaiah NK, Haria D. Biometry with a new swept-source optical coherence tomography biometer: repeatability and agreement with an optical low-coherence reflectometry device. J Cataract Refract Surg. 2016;42:577–81.

    Article  PubMed  Google Scholar 

  34. Akman A, Asena L, Güngör SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol. 2016;100:1201–5.

    Article  PubMed  Google Scholar 

  35. Jung S, Chin HS, Kim NR, Lee KW, Jung JW. Comparison of repeatability and agreement between swept-source Optical Biometry and Dual-Scheimpflug Topography. J Ophthalmol. 2017;2017:1516395.

    Article  PubMed  PubMed Central  Google Scholar 

  36. Arriola-Villalobos P, Almendral-Gómez J, Garzón N, Ruiz-Medrano J, Fernández-Pérez C, Martínez-de-la-Casa JM, et al. Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer. Eye. 2017;31:437–42.

    Article  CAS  PubMed  Google Scholar 

  37. Higashiyama T, Mori H, Nakajima F, Ohji M. Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry. PLoS ONE. 2018;13:e0196401.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Lee HK, Kim MK. Comparison of a new swept-source optical biometer with a partial coherence interferometry. BMC Ophthalmol. 2018;18:269.

    Article  PubMed  PubMed Central  Google Scholar 

  39. An Y, Kang E-K, Kim H, Kang M-J, Byun Y-S, Joo C-K. Accuracy of swept-source optical coherence tomography based biometry for intraocular lens power calculation: a retrospective cross-sectional study. BMC Ophthalmol. 2019;19:30.

    Article  PubMed  PubMed Central  Google Scholar 

  40. Huang J, Chen H, Li Y, Chen Z, Gao R, Yu J, et al. Comprehensive comparison of axial length measurement with three swept-source OCT-Based biometers and partial coherence interferometry. J Refract Surg. 2019;35:115–20.

    Article  PubMed  Google Scholar 

  41. Yang CM, Lim DH, Kim HJ, Chung T-Y. Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer. PLoS ONE. 2019;14:e0223114.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  42. El Chehab H, Agard E, Dot C. Comparison of two biometers: a swept-source optical coherence tomography and an optical low-coherence reflectometry biometer. Eur J Ophthalmol. 2019;29:547–54.

    Article  PubMed  Google Scholar 

  43. Chan E, Mahroo OAR, Spalton DJ. Complications of cataract surgery. Clin Exp Optom. 2010;93:379–89.

    Article  PubMed  Google Scholar 

  44. Miyata K, Nagamoto T, Maruoka S, Tanabe T, Nakahara M, Amano S. Efficacy and safety of the soft-shell technique in cases with a hard lens nucleus. J Cataract Refract Surg. 2002;28:1546–50.

    Article  PubMed  Google Scholar 

  45. Aaronson A, Viljanen A, Kanclerz P, Grzybowski A, Tuuminen R. Cataract complications study: an analysis of adverse effects among 14,520 eyes in relation to surgical experience. Ann Transl Med. 2020;8:1541.

    Article  PubMed  PubMed Central  Google Scholar 

  46. Grzybowski A, Kanclerz P, Huerva V, Ascaso FJ, Tuuminen R. Diabetes and Phacoemulsification Cataract Surgery: Difficulties, Risks and Potential Complications. J Clin Med Res. 2019;8.

  47. Grzybowski A, Kanclerz P. The role of steroids and NSAIDs in Prevention and Treatment of Postsurgical Cystoid Macular Edema. Curr Pharm Des. 2018;24:4896–902.

    Article  CAS  PubMed  Google Scholar 

  48. Anastasilakis K, Mourgela A, Symeonidis C, Dimitrakos SA, Ekonomidis P, Tsinopoulos I. Macular edema after uncomplicated cataract surgery: a role for phacoemulsification energy and vitreoretinal interface status? Eur J Ophthalmol. 2015;25:192–7.

    Article  PubMed  Google Scholar 

  49. Khoramnia R, Auffarth G, Łabuz G, Pettit G, Suryakumar R. Refractive Outcomes after Cataract Surgery. Diagnostics (Basel). 2022;12.

  50. Montés-Micó R, Pastor-Pascual F, Ruiz-Mesa R, Tañá-Rivero P. Ocular biometry with swept-source optical coherence tomography. J Cataract Refract Surg. 2021;47:802–14.

    Article  PubMed  Google Scholar 

  51. Ruiz-Mesa R, Ruiz-Santos M, Blanch-Ruiz J, Jiménez-Nieto A. Acquisition Time for swept-source optical biometry plus corneal power measurement during cataract evaluation. Clin Ophthalmol. 2022;16:661–8.

    Article  PubMed  PubMed Central  Google Scholar 

  52. Armstrong RA. Statistical guidelines for the analysis of data obtained from one or both eyes. Ophthalmic Physiol Opt. 2013;33:7–14.

    Article  PubMed  Google Scholar 

  53. McAlinden C, Khadka J, Pesudovs K. Precision (repeatability and reproducibility) studies and sample-size calculation. J Cataract Refract Surg. 2015;41:2598–604.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Nothing to acknowledge.

Funding

Open Access funding provided by University of Helsinki including Helsinki University Central Hospital.

Author information

Authors and Affiliations

Authors

Contributions

PK, IH, RT conceptualized the paper. PK drafted the initial version of the manuscript, IH and RT revised the paper. PK prepared the Figs. 1, 2, 3 and 4. All authors reviewed the manuscript.

Corresponding author

Correspondence to Piotr Kanclerz.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors report no conflicts of interest and have no proprietary interest in any of the materials mentioned in this article.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Table 1

PRISMA Checklist

Supplement 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kanclerz, P., Hecht, I. & Tuuminen, R. Technical failure rates for biometry between swept-source and older-generation optical coherence methods: a review and meta-analysis. BMC Ophthalmol 23, 182 (2023). https://doi.org/10.1186/s12886-023-02926-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-023-02926-0

Keywords