A custom-made vitreoretinal surgical simulator using a silicone mold

Abstract

Purpose

We constructed a custom-made vitreoretinal surgical simulator using a silicone mold and described its practicality.

Methods

We obtained spherical silicone molds, mannequins, and spray material from an internet-based vendor and combined them with expired surgical instruments to complete the simulator. Vitreoretinal experts confirmed the practicality of the simulator after simulated vitrectomy, and the results of the questionnaires were confirmed by nonvitreoretinal experts.

Results

Vitreoretinal experts observed that the simulated eyeball and the actual eyeball were similar in size and rigidity and that the intraocular practice swing seemed to be useful for the prevention of complications. The semitransparency and open-sky structure of the silicone material ensured visibility. The simulated membrane, which was spray glue, provided an excellent peeling sensation. In the results of the nonvitreoretinal experts’ questionnaires, the average scores of all items were generally high, which supported the claims of the simulator’s usefulness.

Conclusion

This report describes the simplicity and cost-effectiveness of our custom-made simulator and its contribution in creating an ideal training environment that does not necessitate travel to special facilities that offer a large number of pig eyes and vitreous surgical machines. The simple shape seems to allow many possibilities, and further verification at multiple facilities is necessary.

Peer Review reports

Introduction

To improve ophthalmic surgeons’ technique, it would be desirable to provide surgical training that is remarkably simple and affordable and can be casually performed every day. Wet lab training, involving the use of pig eyes, is a common method of practicing cataract surgery [1]. Vitrectomy using pig eyes is also possible [2], but there is no consensus on vitreoretinal surgery training in Japan [3]. It is speculated that there are many countries with similar circumstances. Furthermore, the COVID-19 pandemic has restricted the use of numerous wet lab facilities worldwide [4]. Therefore, such a complex situation may cause complications to occur because of unskilled surgeons.

Practical methods for vitreoretinal surgery, other than wet lab procedures, have been previously reported [3, 5,6,7,8,9,10,11,12]. Virtual reality (VR) machines that support the visualization of various training surgical techniques are attractive simulators [5,6,7,8]. Other surgical simulators that enable the use of actual surgical instruments are also useful [3, 9,10,11,12]. However, young surgeons have requested that more easy-to-use eye model simulations be created, and we have therefore explored novel surgical simulators.

The purpose of this report is to describe the practicality of a simple custom-made vitreoretinal surgical simulator that is cost-effective and easily obtained.

Materials and methods

One of the most important structures of this simulator is its spherical silicone mold, which makes it an ocular substitute. It is generally used as a tool to make accessories by pouring resin into the mold. The material we used had an 8 mm hole on the top and a semitransparent spherical form, with a 25 mm inner diameter and a 1 mm thickness (Fig. 1). The silicone mold is a flexible material that can be easily cut with scissors; we enlarged the hole by several mm and brought it close to the corneal diameter. Furthermore, after pinning the spherical silicone mold onto the optic nerve of the Styrofoam mannequin head, whose orbit was already dug, a very precise human eye model was constructed (Fig. 2).

Fig. 1
figure 1

Spherical silicone mold. We used a spherical silicone mold that had an 8-mm hole at the top and a semitransparent spherical form with a 25 mm inner diameter and a 1 mm thickness

Fig. 2
figure 2

Overviews of our simulator. a Photo taken while using the simulator. b Side view. c Front view

The intraocular membrane was mimicked by spray material (Scotch Spray Glue 55, 3 M Japan Limited, Shinagawa-ku, Tokyo, Japan). The spherical silicone mold was reversed and lightly sprayed on the bottom. Approximately 10–20 min for drying enabled us to use it as our very thin material.

We added sterilized or expired disposable vitreous surgical instruments that were used in the facility to this substitute eyeball. In our hospital, we used vitreous forceps (25G Grieshaber Advanced DSP Tips ILM Forceps, Alcon Laboratories Inc., Fort Worth, TX, USA), cutter and light guide probes (25G Constellation Vision System, Alcon Laboratories Inc., Fort Worth, TX, USA), and other devices.

For observation, we used a surgical microscope (OPMI Lumera 700, Carl Zeiss Meditec, Oberkochen, Germany) and a wide viewing system (Resight, Carl Zeiss Meditec, Oberkochen, Germany).

The practicality was investigated as follows: (1) vitreoretinal experts reviewed the usability of this simulator in the same way as for regular surgery procedures, and (2) nonvitreoretinal experts who used this simulator for practice evaluated the simulator by completing a questionnaire regarding the surgical training (Table 1). Parts I and II asked questions about the preparation, and Parts III-XI asked questions about the actual procedures. With respect to the questions in Part XII, we used these questions to investigate whether they were suitable for continuous practice. Primary surgeons who performed a minimum of 30 vitrectomies for rhegmatogenous retinal detachment and 30 scleral bucklings and who subsequently had continuous surgical practice as vitreoretinal consultants were classified as vitreoretinal experts [13].

Table 1 Questionnaire for nonvitreoretinal experts

Results

Four surgeons were classified as vitreoretinal experts, and five surgeons were classified as nonvitreoretinal experts in our hospital. Reviews of this surgical simulator, using silicone molds, were summarized as follows for each procedure (Online Resource 1 has instructions provided by H.K., a vitreoretinal expert, on how to use this simulator).

Reviews by vitreoretinal experts

The experts made 3 ports 3–4 mm from the simulated corneal limbs (Fig. 3a). Because they considered the rigidity of this material to closely resemble that of a real eyeball, this practice may mimic the sensation of performing sclerotomies, such as the angled incision technique.

Fig. 3
figure 3

Simulated vitreoretinal surgery using our simulator. a Sclerotomies. b Vitrectomy. c Vitreous base shaving under indentation. d Membrane peeling. e Wound closure

Core vitrectomy, posterior vitreous detachment induction and peripheral vitrectomy were performed while taking care not to unknowingly move or distort the simulated eyeball (Fig. 3b). Its open-sky structure made the experts feel that the visibility was excellent. They felt that they had a better surgical field, which was achieved with the help of the wide viewing system. However, with only a surgical microscope, even with the naked eye, they could obtain enough visibility. There were some differences between this simulator and an actual eyeball due to the lack of refractivity of the anterior segment, but it was sufficient and useful for practice. This stroke training seems to facilitate the actual approach from all directions into the vitreous cavity while maintaining a good surgical field, which is important for the prevention of intraoperative complications.

The experts moved a cutter probe and indented each quadrant of the spherical silicone mold using a chandelier lighting system (25G Vivid Chandelier, Synergetics, Inc., St, Charles, Missouri, USA) (Fig. 3c). The material, which features semitransparency and moderate rigidity, was expected to help inexperienced surgeons gain a sense of the coordinated movement of a cutter probe and a scleral depressor.

The membrane made of Scotch Spray Glue 55 was peeled off by vitreous forceps (Fig. 3d). The experts felt that its touch resembled that of the sticky epiretinal membrane (ERM). This feeling induced excitement and allowed trainees who had not experienced vitreous surgery to “peel” the membrane.

The experts removed the cannulas and closed the wounds with surplus surgical sutures (Fig. 3e). This practice contributed to the acquisition of skills to create smooth sutures, although the frictional resistance was slightly increased compared with that of actual eyeballs.

Although Online Resource 1 did not demonstrate this, if a lens (HHV DISPO, Hoya, Shinjuku-ku, Tokyo, Japan) was fitted into the simulated cornea to create a closed space and a gel-like substance was poured, then a cutter probe and infusion system was driven. This imitated vitreous body resection, using a foot pedal, allowed the experts to achieve a more realistic sensation, but the disposable pack was impractical due to its cost and effort.

Evaluations by nonvitreoretinal experts

The results of the questionnaire are shown in Table 2. The mean scores of all the items were 3 or higher, which was a high value. The evaluations of Parts III-XI, regarding the actual procedure, were widely varied, but the evaluations of Parts I and II, regarding preparation, and part XII, regarding continuity, were less varied.

Table 2 Results of the questionnaire completed by nonvitreoretinal experts

Discussion

On occasion, inexperienced trainees have to perform surgery on patients before their skills are well developed [14]. Our simple simulator, using a spherical silicone mold, may be a solution for such a problem in vitreoretinal surgery. This reusable silicone material has moderate flexibility and similarities to the ocular axial length [15] and the thickness of the sclera [16]. Moreover, its semitransparency and open-sky structure are suitable for trainees to observe the model procedures of senior doctors as well as their own procedures. The results of the questionnaire were generally good, seemed to support the abovementioned facts and proved that the simulator corresponded to various vitreous surgery procedures.

Referring to past reports [9, 10], we reproduced extremely thin membranes by spray material, Scotch Spray Glue 55. Moderate spraying gave vitreoretinal experts a sticky ERM-like sensation, which seemed to be easily reproduced and of sufficient quality and helped nonvitreoretinal experts to understand the procedure. We tried numerous kinds of tape and paint materials, but the silicone material prevented them from adhering well. Other unknown ingredients with the potential to produce an excellent membrane have not been sufficiently examined; thus, further studies on membrane materials are necessary.

Traditionally, surgical practice with pig eyes has been common [1, 2], but there are problems in terms of preservation, hygiene, and reusability. Recently, the progress of vitreoretinal surgical training appliances using VR, for example, EyeSi as a representative, has been remarkable [5,6,7,8]. However, EyeSi is extremely expensive and not generally used yet [17]. Dry lab surgical simulators, based on previous reports [3, 9,10,11,12], have been fascinating for understanding direct feedback. Bioniko (https://www.bioniko.com/, accessed as of June 8, 2023), Phillips Studio (https://phillipsstudio.co.uk/, accessed as of June 8, 2023), and other devices are famous eyeball models constructed from marketed products, some of which were used in several past reports [9, 10]. These models were high quality but relatively expensive [11], while spherical silicone molds were likely to be available on various online markets for approximately $2–10 at the time of this experiment (Table 3). Our results showed that nonvitreoretinal experts mainly value the cost-effectiveness, and together, these facts suggest that anyone can casually purchase silicone molds in many countries. Rice et al. also reported a similar low-cost simulator with an open-sky structure [11]. However, this eye model was crafted from a table tennis ball and had a 40 mm diameter, which seemed to differ in terms of size and rigidity when compared to an actual eyeball.

Table 3 Price list of spherical silicone molds in typical online shops

This report has some obvious limitations. The simple shape of the model, which does not take much time and effort to construct, may introduce further creative uses, but its form lacks the cornea, lens, and vitreous body. Additionally, the evaluation of the viscosity of the membrane or the rigidity of the pseudo-eyeball wall is not scientific but subjective because we do not have a device to measure these aspects. Moreover, due to the small number of evaluators and lack of comparisons, further investigation in many other facilities is needed. However, this report reveals the simplicity and cost-effectiveness of our simulator, and the results of this questionnaire substantiated its practicality. Our custom-made simulator seemed to have the potential to create an ideal training environment without necessitating travel to special facilities that offer plenty of pig eyes and vitreous surgical machines. We hope that this report will help young vitreous surgeons and patients worldwide.

Data Availability

The datasets generated and/or analysed during the current study are not publicly available but are available from the corresponding author on reasonable request.

References

  1. Sugiura T, Kurosaka D, Uezuki Y, Eguchi S, Obata H, Takahashi T. Creating cataract in a pig eye. J Cataract Refract Surg. 1999;25:615–21.

    Article  CAS  PubMed  Google Scholar 

  2. Uhlig CE, Gerding H. Illuminated artificial orbit for the training of vitreoretinal surgery in vitro. Eye (Lond). 2004;18:183–7.

    Article  CAS  PubMed  Google Scholar 

  3. Hirata A, Iwakiri R, Okinami S. A simulated eye for vitreous surgery using japanese quail eggs. Graefes Arch Clin Exp Ophthalmol. 2013;251:1621–4.

    Article  PubMed  Google Scholar 

  4. Ferrara M, Romano V, Steel DH, Gupta R, Iovino C, van Dijk EHC, et al. Reshaping ophthalmology training after COVID-19 pandemic. Eye (Lond). 2020;34:2089–97.

    Article  PubMed  Google Scholar 

  5. Rossi JV, Verma D, Fujii GY, Lakhanpal RR, Wu SL, Humayun MS, et al. Virtual vitreoretinal surgical simulator as a training tool. Retina. 2004;24:231–6.

    Article  PubMed  Google Scholar 

  6. Grodin MH, Johnson TM, Acree JL, Glaser BM. Ophthalmic surgical training: a curriculum to enhance surgical simulation. Retina. 2008;28:1509–14.

    Article  PubMed  Google Scholar 

  7. Vergmann AS, Vestergaard AH, Grauslund J. Virtual vitreoretinal surgery: validation of a training programme. Acta Ophthalmol. 2017;95:60–5.

    Article  PubMed  Google Scholar 

  8. Cissé C, Angioi K, Luc A, Berrod JP, Conart JB. EYESI surgical simulator: validity evidence of the vitreoretinal modules. Acta Ophthalmol. 2019;97:e277–82.

    Article  PubMed  Google Scholar 

  9. Yeh S, Chan-Kai BT, Lauer AK. Basic training module for vitreoretinal surgery and the Casey Eye Institute Vitrectomy Indices Tool for Skills Assessment. Clin Ophthalmol. 2011;5:1249–56.

    PubMed  PubMed Central  Google Scholar 

  10. Iyer MN, Han DP. An eye model for practicing vitreoretinal membrane peeling. Arch Ophthalmol. 2006;124:108–10.

    Article  PubMed  Google Scholar 

  11. Rice JC, Steffen J, du Toit L. Simulation training in vitreoretinal surgery: a low-cost, medium-fidelity model. Retina. 2017;37:409–12.

    Article  PubMed  Google Scholar 

  12. Omata S, Someya Y, Adachi S, Masuda T, Hayakawa T, Harada K, et al. A surgical simulator for peeling the inner limiting membrane during wet conditions. PLoS ONE. 2018;13:e0196131.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Koch KR, Hermann MM, Kirchhof B, Fauser S. Success rates of retinal detachment surgery: routine versus emergency setting. Graefes Arch Clin Exp Ophthalmol. 2012;250:1731–6.

    Article  PubMed  Google Scholar 

  14. McCannel CA. Simulation Surgical teaching in Ophthalmology. Ophthalmology. 2015;122:2371–2.

    Article  PubMed  Google Scholar 

  15. Wiseman SJ, Tatham AJ, Meijboom R, Terrera GM, Hamid C, Doubal FN. Measuring axial length of the eye from magnetic resonance brain imaging. BMC Ophthalmol. 2022;22:54. https://doi.org/10.1186/s12886-022-02289-y.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Olsen TW, Aaberg SY, Geroski DH, Edelhauser HF. Human sclera: thickness and surface area. Am J Ophthalmol. 1998;125:237–41.

    Article  CAS  PubMed  Google Scholar 

  17. Young BK, Greenberg PB. Is virtual reality training for resident cataract surgeons cost effective? Graefes Arch Clin Exp Ophthalmol. 2013;251:2295–6.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors thank Kazuya Yamashita, Yoshiko Ofuji, Ryuki Fukumoto, Rio Sato, and Natsuko Shimizu for the critical review of the simulator using the questionnaires.

The authors also thank American Journal Experts for professional medical English editing.

Funding

The authors have no relevant financial or nonfinancial interests to disclose.

Author information

Authors and Affiliations

Authors

Contributions

Takashi Nagamoto conceptualized and drafted the manuscript as a major contributor. Hirohisa Kubono was involved in reviewing this simulator and revised the paper for intellectual content. Mari Kawamura and Kotaro Suzuki supervised this report and were involved in reviewing this simulator.

Corresponding author

Correspondence to Takashi Nagamoto.

Ethics declarations

Ethics approval and consent to participate

This study protocol, conducted according to the principles of the Declaration of Helsinki, is under review by the Institutional Review Board of Keiyu Hospital. Written informed consent was obtained from all participants.

Consent for publication

All participants read and approved this manuscript for publication.

Competing interests

The authors have no conflicts of interest to declare that are relevant to the content of this article.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nagamoto, T., Kubono, H., Kawamura, M. et al. A custom-made vitreoretinal surgical simulator using a silicone mold. BMC Ophthalmol 23, 311 (2023). https://doi.org/10.1186/s12886-023-03070-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-023-03070-5

Keywords