Accuracy of intraocular lens calculation formulas based on swept-source OCT biometer in cataract patients with phakic intraocular lens

Abstract

Purpose

To research the accuracy of intraocular lens (IOL) calculation formulas and investigate the effect of anterior chamber depth (ACD) and lens thickness (LT) measured by swept-source optical coherence tomography biometer (IOLMaster 700) in patients with posterior chamber phakic IOL (PC-pIOL).

Methods

Retrospective case series. The IOLMaster 700 biometer was used to measure axial length (AL) and anterior segment parameters. The traditional formulas (SRK/T, Holladay 1 and Haigis) with or without Wang-Koch (WK) AL adjustment, and new-generation formulas (Barret Universal II [BUII], Emmetropia Verifying Optical [EVO] v2.0, Kane, Pearl-DGS) were utilized in IOL power calculation.

Results

This study enrolled 24 eyes of 24 patients undergoing combined PC-pIOL removal and cataract surgery at Xiamen Eye Center of Xiamen University, Xiamen, Fujian, China. The median absolute prediction error in ascending order was EVO 2.0 (0.33), Kane (0.35), SRK/T-WKmodified (0.42), Holladay 1-WKmodified (0.44), Haigis-WKC1 (0.46), Pearl-DGS (0.47), BUII (0.58), Haigis (0.75), SRK/T (0.79), and Holladay 1 (1.32). The root-mean-square absolute error in ascending order was Haigis-WKC1 (0.591), Holladay 1-WKmodified (0.622), SRK/T-WKmodified (0.623), EVO (0.673), Kane (0.678), Pearl-DGS (0.753), BUII (0.863), Haigis (1.061), SRK/T (1.188), and Holladay 1 (1.513). A detailed analysis of ACD and LT measurement error revealed negligible impact on refractive outcomes in BUII and EVO 2.0 when these parameters were incorporated or omitted in the formula calculation.

Conclusion

The Kane, EVO 2.0, and traditional formulas with WK AL adjustment displayed high prediction accuracy. Furthermore, the ACD and LT measurement error does not exert a significant influence on the accuracy of IOL power calculation formulas in highly myopic eyes implanted with PC-pIOL.

Peer Review reports

Introduction

Owing to the commendable efficacy, predictability, reversibility and preservation of the native corneal biostructure, posterior chamber phakic intraocular lens (PC-pIOL) have emerged as a preferred therapeutic modality for individuals presenting with moderate to high degrees of myopia [1,2,3]. Despite the significant achievements in PC-pIOL implantation for the correction of refractive errors in recent years, there are some complications that warrant attention, including but not limited to lens opacification, endothelial cell loss, and angle-closure glaucoma [4, 5]. Cataract development, a well-recognized complication after PC-pIOL implantation [6], has been reported with an incidence rate ranging from 1.8–9.8% [7, 8]. Notably, Guber et al. reported a noteworthy 18.3% of patients with intraocular collamer lens (ICL) implantation necessitating cataract surgery a decade post-surgery [9]. Given the increasing global utilization of PC-pIOLs, it is prudent to anticipate a growing cohort of individuals necessitating PC-pIOL removal concomitant with cataract extraction and pseudophakic IOL implantation in the foreseeable future [10].

Nowadays, the precision of IOL power calculation formulas has witnessed significant enhancement with advancements in technology and data science, making the cataract surgery much more precise. Beyond common parameters such as axial length (AL) and corneal keratometry (Km), numerous additional variables, including anterior chamber depth (ACD), lens thickness (LT), central corneal thickness (CCT), and white-to-white (WTW), particularly preoperative ACD, contributes to improve the accuracy of IOL power calculation formulas [11,12,13,14]. Existing evidence posits that the presence of PC-pIOL may influence the measurement accuracy of ACD and LT in swept-source optical coherence tomography (SS-OCT) [15, 16]. The inaccurate measurement of ACD and LT would potentially affect the calculation outcome by formulas incorporating ACD or LT into calculation. Furthermore, Ouchi discerned that ACD measurement inaccuracy exerted a discernible influence on the accuracy of Haigis and Barrett Universal II (BUII) TK formulas, resulting in an approximate decrease of 0.3 D in IOL power [16]. The ACD and LT measurement errors attributed to PC-pIOL and its impact on IOL power calculation need further exploration.

Notably, the current research lacks investigations into whether the precision of IOL calculation formulas is susceptible to alterations induced by ACD and LT measurement errors in cataract patients with PC-pIOL, particularly in formulas incorporating ACD or LT into IOL power calculation. Therefore, we examined the prediction accuracy of several traditional formulas (SRK/T, Holladay 1 and Haigis) with or without Wang-Koch (WK) AL adjustment, and new-generation formulas (BUII, Emmetropia Verifying Optical [EVO] v2.0, Kane, Pearl-DGS) in patients implanted with pIOL undergoing cataract surgery, and explore the potential influence of ACD and LT measurement errors on prediction outcomes.

Materials and methods

Study Design

This was a retrospective case series study and performed in accordance with tenets of the Declaration of Helsinki. Ethical clearance was obtained from the Ethics Committee of Xiamen Eye Center of Xiamen University. We retrospectively reviewed the medical records of the patients underwent pIOL removal and phacoemulsification with pseudophakic IOL implantation at the Xiamen Eye Center of Xiamen University, Xiamen, Fujian, China, from March 2018 to October 2023. The general exclusion criteria were as follows: corneal endothelial decompensation (ECD ≤ 1000 cells/mm2), glaucoma, lens or IOL dislocation, posterior capsular rupture during the surgery, ocular fundus diseases, and postoperative best corrected distance VA less than 20/40. Figure 1 shows the flowchart of study enrollment.

Fig. 1
figure 1

Flowchart of patient enrollment

According to the anterior surface of the crystalline lens was identified in IOLMaster 700, we set up two subgroups. The eyes with misidentified segmentation line on the anterior surface of pIOL as the anterior surface of crystalline lens were divided into misidentification group, and correctly identified the anterior surface of crystalline lens were divided into correct identification group (Fig. 2).

Fig. 2
figure 2

The segmentation line correctly identified the anterior surface of crystalline lens (A) and misidentified the anterior surface of pIOL as the anterior surface of crystalline lens (B)

Data Collection and IOL calculation formula

The data collection included demographic data, biometric characteristics, endothelium cell density (ECD), as well as preoperative and postoperative corrected distance visual acuity (CDVA). Biometry was measured by IOLMaster 700 (Carl Zeiss Meditec AG, Jena, Germany), included AL, Km, ACD, LT, CCT and WTW. Instances wherein the anterior surface of the PC-pIOL led to erroneous measurements of ACD and LT were documented. Refractive spherical equivalent (SE) was collected at 1 month after surgery or later.

In this study, several traditional formulas with or without WK-AL adjustment were assessed, Haigis formula used the 1-center regression version (Haigis-WKC1) [18], SRK/T and Holladay 1 formula used the modified version published in 2018 (SRK/T-WKmodified and Holladay 1-WKmodified) [19]. The new-generation formulas included Kane (available at https://www.iolformula.com; accessed on November 2023), Pearl-DGS (available at http://www.iolsolver.com; accessed on November 2023), BUII (available at http://calc.apacrs.org/barrett_universal2105/; accessed on November 2023), and EVO 2.0 (available at www.evoiolcalculator.com; accessed on November 2023). We also assessed the outcomes of BUII and EVO 2.0 without ACD and LT into formula calculation (BUIInoACD+LT and EVO 2.0noACD + LT). Furthermore, the Pearl-DGS formula provided an option to adjust AL after ICL implantation, and this calculation pattern was also assessed (Pearl-DGSICL). The constants used in this study were refer to https://iolcon.org. The Sensar AR40E (Johnson & Johnson Vision), MI60 Akreos (Bausch & Lomb), and CT ASPHINA 409MP (Carl Zeiss) were used in this study, the constants were shown in Supplementary Table 1.

Only AL and Km parameters were considered for BUII and EVO 2.0 formula calculation when ACD and LT was excluded. In contrast, when ACD was integrated into IOL power calculation, variables such as LT, CCT, WTW, or other relevant factors were entered into the formula as requested. The formulas incorporating ACD and/or LT into calculation included Haigis, Haigis-WKC1, Kane, BUII, EVO 2.0, Pearl-DGS, and Pearl-DGSICL, the accuracy of formulas calculation outcomes in both the misidentification and correct identification groups was evaluated. Additionally, within the misidentification group, a comparative analysis was conducted, specifically focusing on the prediction outcomes of BUII and EVO 2.0, with and without ACD and LT incorporation into calculation (BUIInoACD+LT and EVO 2.0noACD + LT).

Statistical analysis

To evaluate the accuracy of the formula, we calculated the refractive prediction error (PE) as the difference between the measured and predicted postoperative SE (actual refraction – predicted refraction). The mean prediction error (ME) is the average of all the PE for each formula. Statistical parameter, including ME, mean absolute prediction error (MAE), median absolute prediction error (MedAE), root-mean-square absolute prediction error (RMSAE), and percentage of eyes within ± 0.25 D, ± 0.5 D, ± 0.75 D and ± 1.0 D of the PE.

Statistical analysis was performed using SPSS software version 26.0 (IBM Corporation) and R project 4.3.0. The normal distribution of data was evaluated using the Shapiro-Wilk test. The one-sample t test was used to assess whether the ME was significantly different from zero. The Friedman test was used to compare the absolute prediction error of formulas. The bootstrap-t method was used to compare the RMSAE values between IOL calculation formulas and the Holm’s adjusted p-value was used for multiple comparisons, according to the statistical analysis suggested by Holladay et al. [32]. and Stopyra et al. [33]. The Cochran’s Q test was used to compare the proportion of eyes within ± 0.25, ± 0.50, ± 0.75 and ± 1.00 D of PE between formulas. The post-hoc Bonferroni correction was used for multiple comparisons. To compare the biometry between misidentification and correct identification groups, the normally distributed data used independent-sample t test and non-normally distributed data used Mann-Whitney U test. A P value of less than 0.05 was considered statistically significant.

Results

This study enrolled 24 eyes of 24 patients with PC-pIOL implantation undergoing cataract surgery, the mean age of the patients was 37.75 ± 5.89 years, 15 females and 9 males. In total, the mean AL was 31.20 ± 2.06 mm, Km was 44.08 ± 1.78 D, ACD was 3.05 ± 0.37 mm, LT was 4.38 ± 0.35 mm, WTW was 11.80 ± 0.45 mm, and CCT was 534.79 ± 30.15 μm. The average IOL power was 3.98 ± 3.88 D. The preoperative and postoperative CDVA was 0.64 ± 0.43 and 0.14 ± 0.23 logMAR, respectively (Table 1).

Table 1 Characteristic biometric parameters

Accuracy of IOL Power Calculation Formulas

The prediction outcomes of formulas are shown in Table 2. Among the formulas, the ME of Kane, SRK/T-WKmodified, Holladay 1-WKmodified, and Haigis-WKC1 displayed no significant difference from zero (P = 0.256, P = 0.413, P = 0.995, and P = 0.530, respectively). The myopic shift was found in Kane and SRK/T-WKmodified, and the hyperopic shift was found in SRK/T, Holladay 1, Haigis, Haigis-WKC1, BUII, BUIInoACD+LT, EVO 2.0, EVO 2.0noACD + LT, Pearl-DGS, and Pearl-DGSICL.

Table 2 Accuracy of IOL Power Calculation Formulas (N=24)

Comparing the MedAE, the best results were obtained by EVO 2.0 (0.33), Kane (0.35), SRK/T-WKmodified (0.42), Holladay 1-WKmodified (0.44), and Haigis-WKC1 (0.46), and the worst by Haigis (0.75), SRK/T (0.79), and Holladay 1 (1.32). The MedAE among formulas had significant difference (P < 0.001). In terms of the RMSAE, the Haigis-WKC1 obtained the lowest RMSAE (0.591), followed by Holladay 1-WKmodified (0.622), SRK/T-WKmodified (0.623), EVO 2.0noACD + LT (0.670), EVO 2.0 (0.673), Kane (0.678), and the Haigis (1.061), SRK/T (1.188), Holladay 1 (1.513) yielded the highest RMSAE. The detailed comparison outcomes of the RMSAE were shown in Supplementary Table 2.

Comparing the proportion of eyes within ± 0.25, ± 0.50, ± 0.75 and ± 1.00 D of PE between formulas (Fig. 3), no significant difference was found in PE within ± 0.25 D between formulas (P = 0.573), but PE within ± 0.5 D, ± 0.75 D and ± 1.0 D were significantly different (P = 0.005, P < 0.001, and P < 0.001, respectively). Most formulas achieved a percentage of > 50% within the ± 0.5 D category. Kane demonstrated the highest percentage (62.5%), followed by EVO 2.0, EVO 2.0noACD + LT, and Haigis-WKC1 (both 58.3%). SRK/T, Holladay 1, and Haigis had low percentage of PE within ± 0.5 D, with a noticeable increase in percentage facilitated by WK-AL adjustment (54.2–58.3%).

Fig. 3
figure 3

Proportion of eyes with absolute PE within ±0.25, ±0.50, ±0.75 and ±1.00 D

Comparisons of misidentification and correct identification groups

In this study, a total of 14 eyes were misidentified by the IOLMaster 700, resulting in a misidentification rate of approximately 58.3%. Upon comparing the biometry between the misidentification and correct identification groups, no statistically significant differences were found in AL, Km, LT, WTW, ECD, and IOL power. For ACD, the misidentification and correct identification groups were 2.81 ± 0.23 and 3.38 ± 0.25, respectively (P < 0.001). No significant difference was found in preoperative and postoperative CDVA between the two groups (Table 1).

The calculation outcomes of formulas incorporating ACD into calculation in both the misidentification and correct identification groups were presented in Table 3. In the misidentification and correct identification groups, the MedAE in ascending order was Kane (0.31; 0.36), EVO 2.0 (0.33; 0.48), Haigis-WKC1 (0.35; 0.55), Pearl-DGS (0.43; 0.73), Pearl-DGSICL (0.43; 0.75), BUII (0.47, 0.79), Haigis (0.75, 1.06), respectively. In terms of the RMSAE, the misidentification group in ascending order was Haigis-WKC1 (0.458), EVO 2.0 (0.619), Pearl-DGS (0.642), Pearl-DGSICL (0.647), Kane (0.746), BUII (0.817), Haigis (0.883), and the correct identification group was Kane (0.569), Haigis-WKC1 (0.738), EVO 2.0 (0.743), Pearl-DGS (0.887), Pearl-DGSICL (0.898), BUII (0.925), Haigis (1.268).

Table 3 Prediction outcomes of formulas incorporated ACD into calculation in misidentification and correct identification groups

Within the misidentification group, BUII and BUIInoACD+LT exhibited ME values of 0.27 ± 0.80 and 0.21 ± 0.87 D, MAE was 0.63 ± 0.54 and 0.69 ± 0.53 D, MedAE was 0.47 and 0.53 D, and RMSAE was 0.817 and 0.862 D, respectively. For EVO 2.0 and EVO 2.0noACD + LT, ME was 0.16 ± 0.62 and 0.15 ± 0.60 D, MAE was 0.49 ± 0.39 and 0.47 ± 0.38 D, MedAE was 0.33 and 0.36 D, and RMSAE was 0.619 and 0.598 D, respectively.

Discussion

The importance of preoperative ACD as a prediction factor for IOL power calculation is only second to AL [12]. Previous study had reported that inaccurate estimations of postoperative ACD contribute substantially, accounting for 42% of the prediction error in IOL power calculation [20]. The IOLMaster 700, based on SS-OCT, stands as a widely utilized biometry measurement device in cataract surgery [21]. Nevertheless, previous studies found that the presence of PC-pIOL may lead to the misidentification of the anterior surface of PC-pIOL as the anterior surface of the lens by the IOLMaster 700 [15, 16, 22]. Zhang J et al. reported a misjudgment rate of approximately 62.5% for the IOLMaster 700 in measuring ACD and LT in cataract patients with PC-pIOL [22], similarly, it was 58.3% in our study. Therefore, it is of great significance to investigate the impact of anterior segment measurement error caused by PC-pIOL on prediction accuracy in this specific cohort.

We suppose that the IOLMaster 700 is based on the segmented measurement principle, and existence of PC-pIOL in the light pathway may alter anterior segment measurement, which contribute to the instrument mistakenly recognizing the anterior surface of the PC-pIOL as the anterior surface of the lens. Intriguingly, we observed that the measurement error of ACD and LT did not exert an influence on the prediction accuracy of formulas incorporating ACD and/or LT into calculation in this study. A prior investigation demonstrated that a 1 D discrepancy in IOL prediction error translates to approximately 0.7 D of refractive error at the spectacle plane [17]. Olsen reported that 1 mm change in ACD results in a 0.32 D refractive shift after the cataract surgery [20]. In our study, the difference in ACD between the misidentification group and correct identification group was 0.57 mm. Despite the influence of PC-pIOL on ACD and LT measurement, this minor difference would not cause significant refractive error. Meier PG et al. also reported that the presence of pIOL did not significantly affect IOL power calculation [23], aligning with our study.

Comparing with normal eyes, the highly myopic eyes have deeper ACD, Miao et al. reported that ACD of elongated eyes was approximately 2.96 ± 0.50 mm [24]. The cases in this study were almost extremely elongated eyes, the average AL was 31.20 ± 2.06 mm and ACD was 3.05 ± 0.37 mm. Norrby S demonstrated that the change of ACD have less impact on the PE of formulas calculation in highly myopic eyes [25]. Similarly, Vega et al. suggested that the influence of ACD is minimal in the BUII formula calculation for eyes with AL longer than 22 mm [26]. Consequently, it is plausible to integrate error ACD and LT directly into formula calculations for highly myopic eyes without manual correction, while exercising caution in shorter eyes where the impact of error ACD and LT warrants consideration.

In the comparisons of IOL calculation formulas, Kane, EVO 2.0, Haigis-WKC1, Holladay 1-WKmodified, and SRK/T-WKmodified demonstrated the good calculation accuracy in both MedAE and RMSAE. The analysis of MAE, MedAE and PE within ± 0.50 D among formulas revealed that Kane (0.49 ± 0.48 D, 0.35 D, and 62.5%) and EVO 2.0 (0.53 ± 0.43, 0.33 D, and 58.3%) performed optimally among the new-generation formulas. As previous studies have demonstrated, these two formulas showed high prediction accuracy in highly myopic eyes, Kane (0.318 ± 0.227 D, 0.271 D, and 78.9%), and EVO 2.0 (0.314 ± 0.216 D, 0.288 D, and 82.3%) [27]. The Kane formula uses a combination of theoretical optics, thin lens formulas, and ‘‘big data’’ techniques, it is based on optics and includes both regression and AI elements to improve the prediction outcomes, especially have high prediction accuracy in extreme eyeball AL [34, 35]. The EVO is a thick-lens formula based on the theory of emmetropization that generates an ‘‘emmetropia factor’’ for each eye, EVO version 2.0 has improved prediction for long and short AL, or steep and flat Km [35]. Within traditional formulas, the WK AL adjustment significantly improved refractive prediction accuracy. We hypothesize that the formulas which are suitable for elongated eyes may also be applicable to highly myopic patients implanted with PC-pIOL, yielding good prediction accuracy.

Notably, we compared the outcomes of EVO 2.0 and BUII with and without incorporating error ACD and LT into IOL power calculation in the misidentification group, and the results indicated that error ACD and LT did not significantly influence PE in these two formulas. However, we found the EVO 2.0noACD + LT performed slightly better, aligning with the findings of Savini G et al. that EVO 2.0 operates more accurately without factoring in ACD [28]. This interesting observation leads us to recommend the utilization of EVO 2.0 and BUII formulas without ACD and LT incorporation in this particular group, thereby not only circumventing the potential impact of error ACD and LT but also potentially enhancing prediction accuracy.

In terms of Pearl-DGS formula, we found that the optimization calculation for patients implanted with ICL had no significant effect on prediction accuracy. Prior studies have reported that the presence of pIOL and their materials can influence ultrasound AL measurements [29, 30], but have minimal impact on optical measurements [16]. The optimization calculation in Pearl-DGS involves a slight correction of AL, guided by prior studies indicating that the presence of phakic IOLs can lead to AL measurement errors [15, 31]. We posit that the adjusted AL is approximately 0.01 to 0.02 mm, a negligible change that does not significantly affect IOL power calculation in highly myopic eyes. Our results indicate that the calculation outcomes of Pearl-DGS are almost identical to Pearl-DGSICL.

There are some limitations in this study. Firstly, the sample size was constrained, warranting a larger cohort in future investigations. Secondly, the eyes enrolled in this study were elongated AL, thus limiting the generalization of findings to normal and shorter AL eyes. Thirdly, the study exclusively employed the IOLMaster 700 due to equipment constraints, necessitating further exploration of other optical biometry measurement devices’ performance. Fourthly, although optimizing the lens constants can improve the prediction accuracy of IOL power calculation formulas, it is difficult to optimize the constants in this small size, retrospective, and single-center study, thus we used the optimized constants recommended in https://iolcon.org.

In conclusion, the measurement error of ACD and LT in patients with PC-pIOL does not exert a significant influence on the prediction accuracy of IOL power calculation formulas. Among the new-generation formulas, the Kane and EVO 2.0 performed best, and the traditional formulas with WK-AL adjustment demonstrate high prediction accuracy in this specific cohort.

Data availability

The data presented in this study are included in the article. The data are not publicly available due to restrictions that apply to the availability of the data (e.g., privacy or ethical). Datasets from this study may be available upon request from the corresponding author and provided upon approval from the sponsor and in accordance with data privacy and ethical provisions.

Abbreviations

ACD:

Anterior chamber depth

AL:

Axial length

CCT:

Central corneal thickness

CDVA:

Corrected distance visual acuity

D:

Diopter

ECD:

Endothelium cell density

ICL:

Intraocular collamer lens

IOL:

Intraocular lens

Km:

Corneal keratometry

LT:

Lens thickness

MAE:

Mean absolute prediction error

ME:

Mean prediction error

MedAE:

Median absolute prediction error

PC-pIOL:

Posterior chamber phakic intraocular lens

PE:

Prediction error

RMSAE:

Root-mean-square absolute error

SE:

Spherical equivalent

SS-OCT:

Swept-source optical coherence tomography

WTW:

White-to-white

WK:

Wang-Koch

References

  1. Montés-Micó R, Ruiz-Mesa R, Rodríguez-Prats JL, Tañá-Rivero P. Posterior-chamber phakic implantable collamer lenses with a central port: a review. Acta Ophthalmol. 2021;99(3):e288–301.

    Article  PubMed  Google Scholar 

  2. Moshirfar M, Webster CR, Ronquillo YC. Phakic intraocular lenses: an update and review for the treatment of myopia and myopic astigmatism in the United States. Curr Opin Ophthalmol. 2022;33(5):453–63.

    Article  PubMed  Google Scholar 

  3. Nakamura T, Isogai N, Kojima T, Yoshida Y, Sugiyama Y. Posterior Chamber Phakic Intraocular Lens Implantation for the correction of myopia and myopic astigmatism: a retrospective 10-Year follow-up study. Am J Ophthalmol. 2019;206:1–10.

    Article  PubMed  Google Scholar 

  4. Kohnen T, Kook D, Morral M, Güell JL. Phakic intraocular lenses: part 2: results and complications. J Cataract Refract Surg. 2010;36(12):2168–94.

    Article  PubMed  Google Scholar 

  5. Sucu ME, Cakmak S, Yildirim Y, Yildiz BK, Yalçınkaya G, Beşek NK, Yasar T. Explantation of phakic intraocular lenses: causes and outcomes. Int Ophthalmol. 2021;41(1):265–71.

    Article  CAS  PubMed  Google Scholar 

  6. Moshirfar M, Mifflin M, Wong G, Chang JC. Cataract surgery following phakic intraocular lens implantation. Curr Opin Ophthalmol. 2010;21(1):39–44.

    Article  PubMed  Google Scholar 

  7. Chen LJ, Chang YJ, Kuo JC, Rajagopal R, Azar DT. Metaanalysis of cataract development after phakic intraocular lens surgery. J Cataract Refract Surg. 2008;34(7):1181–200.

    Article  PubMed  Google Scholar 

  8. Kamiya K, Shimizu K, Igarashi A, Hikita F, Komatsu M. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol. 2009;127(7):845–50.

    Article  PubMed  Google Scholar 

  9. Guber I, Mouvet V, Bergin C, Perritaz S, Othenin-Girard P, Majo F. Clinical outcomes and cataract formation rates in eyes 10 years after posterior phakic Lens Implantation for Myopia. JAMA Ophthalmol. 2016;134(5):487–94.

    Article  PubMed  Google Scholar 

  10. Jonker SMR, Berendschot T, Saelens IEY, Bauer NJC, Nuijts R. Phakic intraocular lenses: an overview. Indian J Ophthalmol. 2020;68(12):2779–96.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Kesim C, Yıldız-Taş A, Karslıoğlu MZ, Hasanreisoğlu M, Müftüoğlu O, Şahin A. The Effect of Anterior Segment depth on the Accuracy of 7 different intraocular Lens calculation formulas. Turk J Ophthalmol. 2022;52(4):228–36.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Olsen T. Prediction of the effective postoperative (intraocular lens) anterior chamber depth. J Cataract Refract Surg. 2006;32(3):419–24.

    Article  PubMed  Google Scholar 

  13. Jeong J, Song H, Lee JK, Chuck RS, Kwon JW. The effect of ocular biometric factors on the accuracy of various IOL power calculation formulas. BMC Ophthalmol. 2017;17(1):62.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Ning X, Yang Y, Yan H, Zhang J. Anterior chamber depth - a predictor of refractive outcomes after age-related cataract surgery. BMC Ophthalmol. 2019;19(1):134.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Chen X, Zhang D, Liu Z, Liu Y, Cai H, Wu Q, Zhang Y. Effect of Implantable Collamer Lens on Anterior Segment Measurement and Intraocular Lens Power Calculation Based on IOLMaster 700 and Sirius. J Ophthalmol. 2021;2021:8988479.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Ouchi M. Evaluation of impact of posterior phakic IOL implantation on biometry and effectiveness of concomitant use of anterior segment OCT on IOL power calculation for cataract surgery. J Cataract Refract Surg. 2022;48(6):657–62.

    Article  PubMed  Google Scholar 

  17. Wang L, Tang M, Huang D, Weikert MP, Koch DD. Comparison of newer intraocular Lens Power calculation methods for eyes after corneal refractive surgery. Ophthalmology. 2015;122(12):2443–9.

    Article  PubMed  Google Scholar 

  18. Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD. Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg. 2011;37(11):2018–27.

    Article  PubMed  Google Scholar 

  19. Wang L, Koch DD. Modified axial length adjustment formulas in long eyes. J Cataract Refract Surg. 2018;44(11):1396–7.

    Article  PubMed  Google Scholar 

  20. Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol Scand. 2007;85(5):472–85.

    Article  PubMed  Google Scholar 

  21. Bullimore MA, Slade S, Yoo P, Otani T. An evaluation of the IOLMaster 700. Eye Contact Lens. 2019;45(2):117–23.

    Article  PubMed  Google Scholar 

  22. Zhang J, Xia Z, Han X, Liu Z, Lin H, Qiu X, Zhang M, Ruan X, Chen X, Jin G, Gu X, Tan X, Luo L, Liu Y. Accuracy of Intraocular Lens Calculation Formulas in patients undergoing combined phakic intraocular Lens removal and cataract surgery. Am J Ophthalmol. 2022;234:241–9.

    Article  PubMed  Google Scholar 

  23. Meier PG, Majo F, Othenin-Girard P, Bergin C, Guber I. Refractive outcomes and complications after combined copolymer phakic intraocular lens explantation and phacoemulsification with intraocular lens implantation. J Cataract Refract Surg. 2017;43(6):748–53.

    Article  PubMed  Google Scholar 

  24. Miao A, Tang Y, Zhu X, Qian D, Zheng T, Lu Y. Associations between anterior segment biometry and high axial myopia in 3438 cataractous eyes in the Chinese population. BMC Ophthalmol. 2022;22(1):71.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008;34(3):368–76.

    Article  PubMed  Google Scholar 

  26. Vega Y, Gershoni A, Achiron A, Tuuminen R, Weinberger Y, Livny E, Nahum Y, Bahar I, Elbaz U. High agreement between Barrett Universal II calculations with and without utilization of Optional Biometry parameters. J Clin Med 2021;10(3).

  27. Lin L, Xu M, Mo E, Huang S, Qi X, Gu S, Sun W, Su Q, Li J, Zhao YE. Accuracy of newer Generation IOL Power Calculation Formulas in eyes with high Axial Myopia. J Refract Surg. 2021;37(11):754–8.

    Article  PubMed  Google Scholar 

  28. Savini G, Di Maita M, Hoffer KJ, Næser K, Schiano-Lomoriello D, Vagge A, Di Cello L, Traverso CE. Comparison of 13 formulas for IOL power calculation with measurements from partial coherence interferometry. Br J Ophthalmol. 2021;105(4):484–9.

    Article  PubMed  Google Scholar 

  29. Morales AJ, Zadok D, Tardio E, Anzoulatous G Jr., Litwak S, Mora R, Martinez E, Chayet AS. Outcome of simultaneous phakic implantable contact lens removal with cataract extraction and pseudophakic intraocular lens implantation. J Cataract Refract Surg. 2006;32(4):595–8.

    Article  PubMed  Google Scholar 

  30. Hoffer KJ. Ultrasound axial length measurement in biphakic eyes. J Cataract Refract Surg. 2003;29(5):961–5.

    Article  PubMed  Google Scholar 

  31. Pitault G, Leboeuf C, Leroux les Jardins S, Auclin F, Chong-Sit D, Baudouin C. [Optical biometry of eyes corrected by phakic intraocular lenses]. J Fr Ophtalmol. 2005;28(10):1052–7.

    Article  CAS  PubMed  Google Scholar 

  32. Holladay JT, Wilcox RR, Koch DD, Wang L. Review and recommendations for univariate statistical analysis of spherical equivalent prediction error for IOL power calculations. J Cataract Refract Surg. 2021;47(1):65–77.

    Article  PubMed  Google Scholar 

  33. Stopyra W, Voytsekhivskyy O, Grzybowski A. Accuracy of 20 Intraocular Lens Power Calculation Formulas in medium-long eyes. Ophthalmol Ther. 2024;13(7):1893–907.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Stopyra W, Cooke DL, Grzybowski A. A review of intraocular Lens Power calculation formulas based on Artificial Intelligence. J Clin Med 2024;13(2).

  35. Stopyra W, Langenbucher A, Grzybowski A. Intraocular Lens Power Calculation Formulas-A Systematic Review. Ophthalmol Ther. 2023;12(6):2881–902.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

None.

Funding

The study was supported by Natural Science Foundation of Xiamen (3502Z20227127).

Author information

Authors and Affiliations

Authors

Contributions

Zongsheng Zeng: Conceptualization, Methodology, Validation, Investigation, Data Curation, Writing-Review & Editing, and Visualization. Meiyi Zhu: Formal Analysis, Data Curation, Writing-Original Draft, Writing-Review & Editing, and Visualization. Guangbin Zhang: Conceptualization, Methodology, Resources, Supervision, and Project administration. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Guangbin Zhang.

Ethics declarations

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Ethics approval and informed consent

The studies involving human participants were reviewed and approved by the Human Ethics Committee of Xiamen University affiliated with the Xiamen Eye Center. All participants provided written informed consent to take part in this study.

Financial support

None.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zeng, Z., Zhu, M. & Zhang, G. Accuracy of intraocular lens calculation formulas based on swept-source OCT biometer in cataract patients with phakic intraocular lens. BMC Ophthalmol 24, 326 (2024). https://doi.org/10.1186/s12886-024-03605-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-024-03605-4

Keywords