Easy and effective test to evaluate tear-film stability for self-diagnosis of dry eye syndrome: blinking tolerance time (BTT)

Abstract

Background

The tear film breakup time (tBUT) is a clinical evaluation of evaporative dry eye disease assessed by instilling topical fluorescein into the eyes. In the present study, we introduce a new diagnostic test, blinking tolerance time (BTT), for self-evaluation of tear-film stability. We compared the results with the tBUT and validated the BTT test for self-assessment of tear film instability.

Methods

This was a prospective controlled study involving 212 eyes of 106 participants 20–79 years of age. A total of 114 eyes of 57 dry eye patients and 98 eyes of 49 healthy subjects were included in the study. All patients and subjects were administered the following tests to diagnose dry eye disease: Ocular Surface Disease Index, BTT, tBUT, slit-lamp examination, corneal stain score, and Schirmer I test (without anesthesia). Patients and subjects were instructed not to blink for as long possible after reset blinking. The time interval between the reset blink and the next blink was measured. The mean of 3 tBUT values in both the right and left eyes was defined as tBUTBE. Correlations between the BTT and tBUTBE were also evaluated. To evaluate the diagnostic efficacy of the BTT and tBUT tests, receiver operating characteristic (ROC) curves were generated to obtain a cutoff score, and the sensitivities of the tests against the specificity at all possible thresholds were plotted.

Results

Spearman’s correlation analysis revealed a significant weakly positive correlation between BTT and tBUTBE (r = 0.447; p = 0.000). The intraclass correlation coefficient (ICC) of the tBUT was 0.679 (95% confidence interval [CI]: 0.575–0.765) and the ICC of the BTT was 0.904 (95% CI: 0.867–0.932). The area under the ROC curve did not significantly differ between the tBUTBE (0.678) and BTT (0.628, p = 0.641). When the cutoff value of the BTT test was set to 8.1 s, the sensitivity was 63.3% and the specificity was 56.1%.

Conclusion

The BTT test is a simple, inexpensive, and effective method for self-diagnosing dry eye that can also be used in the clinical setting.

Peer Review reports

Background

Tear film instability plays a crucial role in the pathophysiology of dry eye disease. Thinning of the aqueous layer of the tear film due to reduced tear secretion from the lacrimal glands or thinning of the lipid layer of the tear film due to reduced meibum (oil) secretion from the meibomian glands leads to rapid evaporation and tear film breakup [1]. The tear film breakup time (tBUT) is traditionally measured to assess tear-film stability in the clinic. In the tBUT test, fluorescein dye is applied to the conjunctival sac and the time to tear breakup is evaluated while the patient eye is open under the cobalt blue light of a slit-lamp [2]. Although this method is effective, there are some disadvantages. First, the use fluorescein dye does not allow for observation of the physiological state of the ocular surface; moreover, the breakup time is dependent on the amount of fluorescein dye used, so that the tBUT value may differ accordingly. Second, it is sometimes difficult to determine when the tear film begins to breakup, so the tBUT test has low reproducibility [3]. Finally, although the probability is very low, the use of fluorescein dye poses a risk of infection.

There have been some attempts to overcome these limitations of the tBUT test. Recently, a noninvasive tBUT test was developed to measure the tear breakup time without fluorescein dye using the Keratograph 5 M instrument (K5M; Oculus Optikgerate, Wetzlar, Germany) [4] or DR-1α (Kowa, Nagoya, Japan). While this method allows for tBUT to be performed noninvasively, these instruments may not be readily available.

A staring contest is a well-known game in which 2 people face each other and the person who can delay blinking for the longest time wins the game. We considered that this game could be an inexpensive and simple method to assess tear-film stability. The aim of the present study was to introduce a new test, the blinking tolerance time (BTT) test, that could be used for self-assessment of tear-film stability. To evaluate the clinical usefulness of the BTT test, the results were compared with tBUT measurements obtained with the traditional method using fluorescein dye.

Methods

Selection of participants

This study followed the tenets of the Declaration of Helsinki, and the Institutional Review Board of Chuncheon Sacred Hospital and Incheon St. Mary’s Hospital approved the protocol. Patients were recruited at the Cornea Service of Chuncheon Sacred Hospital. The study included healthy subjects (n = 49) and dry eye patients (n = 57) matched for age and sex. For the dry eye group, the inclusion criteria were age ≥ 20 years and at least mild dry eye symptoms (an Ocular Surface Disease Index [OSDI] score ≥ 13) and low tear film break-up time (TBUT < 5 s), or low Schirmer I score (< 10 mm per 5 min without anesthesia), or corneal punctate fluorescein staining (Oxford staining score of > 1) in at least 1 eye. Exclusion criteria were a history of ocular injury, infection, non-dry eye ocular inflammation, trauma, surgery within the prior 6 months, or the presence of uncontrolled systemic disease. For healthy subjects, inclusion criteria were no dry eye syndrome (no typical dry eye syndrome symptoms, tBUT > 5 s, Schirmer I score > 10 mm, Oxford staining score < 1), and the exclusion criteria were the same.

Evaluation of dry eye

The examinations were performed as follows: OSDI, BTT, slit-lamp examination, tBUT, corneal stain, and Schirmer I (without anesthesia) test. Using the OSDI score, subjective symptoms of dry eye were graded numerically from 0 to 4, and the sum of these scores was used in the analyses [5]. The OSDI, developed by the Outcomes Research Group at Allergan (Irvine, CA, USA), is a 12-item questionnaire yielding information on vision-related functioning [5]. Each item is scored on a 5-point scale, resulting in a total OSDI score ranging from 0 (no symptoms) to 100 (maximal symptoms). Slit-lamp examination was performed to identify exclusion criteria. For the tBUT test, a sterile fluorescein strip (Haag, Heidelberg, Germany) moistened with balanced salt solution was applied to the inferior fornix of both eyes. After approximately 2 min, the subjects were asked to blink several times to ensure mixing of the dye and tear fluid, and the time interval between the last blink and the appearance of the first dry spot of tear film was recorded. To evaluate intra-examiner repeatability, 3 consecutive measurements were obtained by the same clinician. The tBUT was recorded to the nearest 0.1 s using a digital stopwatch. The mean of three tBUT measurements was determined for each eye. The average of the mean value for both eyes was defined as tBUTBE. The lowest mean tBUT value between the 2 eyes was defined as tBUTs. Corneal fluorescein staining was also performed under cobalt blue slit-lamp illumination. The Oxford grading scheme [6] was applied to evaluate the ocular surface damage. Approximately 15 min after the tBUT test, the Schirmer test was performed without topical anesthesia.

Blinking tolerance time test

The BTT test was performed in eyes of all participants by the same clinician. The BTT test was conducted in a silent and windless examination room. The temperature and humidity of the room were carefully controlled (23 °C ~ 26 °C and 40% ~ 60%, respectively). Air conditioners and electric fans were not in use during the examination. For the BTT tests, subjects and patients sat in the examination chair with their eyes closed. Immediately after opening both eyes, the investigator instructed each participant to fix his or her eyes to a visual target 3 m away and keep their eyes open until they had to blink due to ocular pain. The investigator measured the time during which the eyes were open without blinking to within 0.1 s using a stopwatch, i.e., the time interval between the last blink and the unavoidable next blink. The unavoidable next blink resulted from the participant experiencing irritation or pain. The participants were instructed not to over-open or unnaturally open their eyes, and the BTT test was performed with the eyes open as naturally as possible. To measure the intra-examiner repeatability, the BTT test results were calculated based on 3 consecutive measurements obtained by the same clinician. Between each measurement, the participants were given a break for at least 30 s to ensure stable and reproducible data. If reflex tearing occurred during a BTT test, the results were excluded from the analysis.

Order of the diagnostic testing procedures

The order of the testing procedures during the ophthalmic examinations was as follows:

  1. 1.

    Subjective interview regarding symptoms of dry eye (i.e., OSDI) and recording of patient medical history.

  2. 2.

    BTT test, repeated 3 times.

  3. 3.

    Slit-lamp evaluation of the cornea, conjunctiva, eyelids, and Meibomian glands.

  4. 4.

    Schirmer I test (without anesthesia).

  5. 5.

    tBUT test using fluorescein dye, repeated 3 times.

  6. 6.

    Fluorescein staining of the cornea.

Statistical analysis

All statistical analyses were performed using SAS software (ver. 9.0; SAS Institute, Cary, NC, USA). A value of P < 0.05 was considered statistically significant, and 95% confidence intervals (CIs) were calculated. The mean ± standard deviation (SD) of age, OSDI, Schirmer values (mm), tBUTBE, and tBUTS values were calculated for all participants. For all participants, we also calculated the mean ± SD of the difference between the BTT and tBUTBE, and of the difference between the BTT and tBUTS. The distribution of the BTT values was measured, and the Kolmogorov-Smirnov test was performed to confirm that the data were normally distributed. The Wilcoxon signed-rank test was performed to determine whether there was a statistically significant difference between the mean values of the BTT and tBUTBE measured in all participants and between the mean values of the BTT and tBUTS. Spearman’s correlation test was performed to confirm the correlation between the BTT and tBUTBE, and Spearman’s rho values were calculated. In addition, Spearman’s correlation was used to determine whether there was a correlation between the BTT and OSDI values.

The mean values for age, sex, OSDI, Schirmer values, corneal staining score (according to the Oxford grading scheme), tBUTBE, tBUTS, and BTT were calculated for the control group and dry eye patient groups. Also, the difference between the BTT and tBUTBE, and between the BTT and tBUTS, were calculated for both groups, and the mean value of these differences was calculated. The Mann-Whitney U test was then used to confirm whether the mean values were significantly different between the 2 groups. In addition, the Wilcoxon signed-rank test was used to determine whether there was a statistically significant difference between the BTT and tBUTBE in each group.

For repeated measurements of the tBUT and BTT, intraclass correlation coefficients (ICCs) were calculated to determine the consistency between the measurements (i.e., the repeatability of the measurements within the same participant). The ICCs of the BTT and tBUT were used to determine the intra-examiner repeatability (ICC ≥ 0.75 indicated good reliability). To evaluate and maximize the diagnostic efficacy of the BTT and tBUT tests, receiver operating characteristic (ROC) curves were generated to obtain a cutoff score, and the sensitivities of the tests against the specificity at all possible thresholds were plotted. The area under the ROC curve (AUC) was used to determine how the tests performed as diagnostic instruments. An AUC value of 0.5 indicated a completely ineffective diagnostic test, and a value of 1.0 indicated a maximally effective diagnostic test [5, 7].

Results

Table 1 shows the age, sex, OSDI, Schirmer test, corneal staining, tBUTBE, BUTS, BTT, BTT-tBUTBE, and BTT-tBUTS data for the normal and dry eye patient groups.

Table 1 Basic characteristics of enrolled subjects

The mean values of tBUTBE and BTT for all 106 participants were 12.2 ± 9.2 s and 13.5 ± 17.1 s, respectively. There was no statistically significant difference between the 2 mean values (p = 0.215; Wilcoxon signed-rank test). The mean value of TBUTS was 8.8 ± 5.3 s, which was significantly shorter than that of BTT (p = 0.003; Wilcoxon signed-rank test). The mean value of the difference between the tBUTS and BTT in each participant was 4.8 ± 16.8 s.

Figure 1 shows that the distribution of BTT values was skewed to the left, and the BTT values were not normally distributed according to the results of the Kolmogorov-Smirnov normality test (p = 0.000). Figure 2 shows the correlation between the BTT and tBUTBE values. Spearman’s correlation analysis showed a statistically significant, weakly positive correlation between the BTT and tBUTBE values (r = 0.447; p = 0.000). There was no statistically significant correlation between the BTT and OSDI values among all participants (r = − 0.160; p = 0.105; Spearman’s correlation test). The ICC of the tBUT values was 0.679 (95% CI: 0.575–0.765) and the ICC of the BTT values was 0.904 (95% CI: 0.867–0.932).

Fig. 1
figure 1

The blinking tolerance time (BTT) values exhibited a left-skewed distribution. The Kolmogorov-Smirnov test indicated that the BTT test values were not normally distributed (p = 0.000)

Fig. 2
figure 2

The BTT and tear breakup time (tBUT) without anesthesia showed a statistically significant positive correlation (r = 0.447; p = 0.000; Spearman’s correlation analysis)

The age and sex distributions did not differ significantly between the normal and dry eye groups (p = 0.200 and 0.159, respectively; Mann-Whitney U test and Fisher’s exact test) (Table 1). The OSDI values and corneal stain were higher in the dry eye group compared with the normal group. The Schirmer, tBUTBE, and tBUTs values were lower in the dry eye group. The BTT was significantly longer in the normal group than in the dry eye group (18.1 ± 22.8 s vs 9.6 ± 8.1 s, p = 0.024, Mann-Whitney U test).

In the normal subject group, the BTT and tBUTBE did not differ significantly (p = 0.483, Wilcoxon signed-rank test). In the dry eye group, the BTT and tBUTBE values also did not differ significantly (p = 0.324, Wilcoxon signed-rank test). Further, BTT-tBUTBE did not differ significantly between the normal subjects and dry eye group (p = 0.534, Wilcoxon signed-rank test). The BTT-tBUTS did not differ significantly between the normal subjects and dry eye group (p = 0.626, Wilcoxon signed-rank test).

The ROC curves of the tBUTBE and BTT values are shown in Fig. 3. The AUC of the tBUTBE was 0.678 and that of the BTT was 0.628, with no statistically significant difference between these values (p = 0.641; method of DeLong et al. [8]). When the cutoff value of BTT was set to 8.1 s, the sensitivity of the BTT test was 63.3% and the specificity was 56.1%.

Fig. 3
figure 3

Receiver operating characteristic (ROC) curve of tear breakup time without anesthesia (TBUT[BE]) and the BTT. The area under the ROC curve (AUC) of the tBUTBE was 0.678 and the AUC of the BTT was 0.628. The difference between these 2 AUC values did not reach statistical significance (p = 0.641)

Discussion

We report a new diagnostic test, called the BTT test, for evaluating tear film instability. This test evaluates the stability of the tear film indirectly by measuring the time interval between eye blinks when the participant has been asked to refrain from blinking for as long as possible before feeling a foreign body sensation or irritation. Tear film instability is a key mechanism of dry eye disease [9], and new, effective diagnostic approaches that are inexpensive and noninvasive would be beneficial.

The present study revealed no statistically significant difference between the tBUTBE and BTT among all participants, and a significant positive correlation between the 2 measurements. Further, there was no significant difference in the AUC between the tBUTBE and BTT values. The BTT values for the dry eye group were significantly shorter than those for the normal control group. Thus, the BTT test can be used for self-diagnosis of dry eye disease.

Sophisticated scientific diagnostic instruments, such as the TearLab system (TearLab, San Diego, CA, USA) and LipiView (TearScience, Morrisville, NC, USA) interferometry systems, which incorporate remarkable technologic innovations, have been developed for clinical use. Dry eye disease is a prevalent disorder that represents a major public health burden due to its negative impact on patient vision and quality of life. The BTT test is a simple method for self-diagnosis of this disease with high reliability. This inexpensive new diagnostic tool can also be used widely by clinicians to diagnose dry eye disease.

Several studies have been conducted to determine the association between eye blinking and dry eye disease. Pult et al. [10] performed a computerized analysis of video recordings of spontaneous blinking in normal and dry eye patients, and reported that the inter-blinking time (IBT) differed significantly between the 2 groups (4.0 ± 2.0 s for the normal group and 1.5 ± 0.9 s for the dry eye group). The IBT in their study corresponds to the BTT used in our study. The BTT differs from the IBT, however, in that the participant is asked not to blink after the reset blink until he or she feels ocular pain. In addition, our study focused on analyzing the correlation between the BTT and tBUT tests without using complex equipment, such as video capture and analysis equipment, in contrast to the study by Pult et al. [10]. We first confirmed that there was no significant difference between the BTT and tBUTBE values; second, we found a significant positive correlation between the 2 tests; third, the ICC of the BTT test was superior to that of the tBUTBE test; and finally, there was no significant difference between the AUCs of the 2 tests. These findings confirmed that the BTT test is appropriate for dry eye diagnosis, having comparable utility to the tBUT test. The diversity of the blink rate is related to the severity of dry eye disease. Previous studies reported that the blink rate increased in the aqueous-deficient type of dry eye disease; furthermore, the increased blink rate was positively correlated with corneal staining and subjective dry eye symptoms, and negatively correlated with the tBUT [11, 12]. Several reports have described the mechanism underlying blinking in patients with dry eye disease.

Ocular surface temperature decreases over time after blinking [13]. The tear film begins to destabilize after blinking due to evaporation, which cools the ocular surface [14]. The rate of evaporation is increased in patients with dry eye due to the instability of the lipid layer [15]. The ocular surface temperature may therefore be decreased in patients with dry eye due to tear film instability immediately after blinking compared with normal controls [16]. Stimulation of cold thermoreceptors in the cornea between blinks can result in basal tear secretion, and such stimulation in normal controls is related to ocular comfort and wetness [17]. Reduced sensitivity of these thermoreceptors in patients with dry eye, who already have a cooler-than-normal corneal temperature of 34 °C, can decrease the physiologic tearing stimulation, thus leading to increased subjective dry eye symptoms such as pain and a stinging sensation [12].

The concept of the BTT is different from the blinking interval in a relaxed state. The BTT test measures the time until the participant blinks after sensing a foreign body sensation or irritation. The BTT measurement is therefore consistent with the concept of tear film stability, which is not related to the pain threshold or nociceptor activation. It is possible that the BTT is related to the pain threshold and nociceptor activation, and that this test is thus less objective than the tBUT test, being relatively strongly associated with the symptoms afflicting patients with dry eye. Therefore, we analyzed the correlation between the BTT and OSDI results, but detected no significant association.

In the present study, we aimed to demonstrate the efficacy of the BTT test, a new method for diagnosing dry eye, compared with the conventional tBUT test. The BTT test can serve as a useful dry eye diagnostic tool as an adjunct to the tBUT test: first, the mean values of the BTT and tBUT in the normal control group did not differ significantly from each other; second, there was no statistically significant difference in the AUC between the BTT and tBUT; and third, both the tBUT and the BTT were significantly shorter in patients with dry eye than in the normal controls. The BTT and tBUT measurements, however, showed a weakly positive correlation, for the following reasons: first, both the BTT and tBUT tests may not be highly reproducible; second, the pain threshold may differ among participants; and third, an understanding of and willingness to undergo the BTT test (which involves blinking naturally after feeling a foreign body sensation or irritation) may also differ among individuals.

The BTT test has several advantages compared with the tBUT test. First, it can be used as a self-test of dry eye disease. The BTT test is very easy and simple to perform, such that anyone can use it to evaluate dry eye disease without assistance from a clinician. In this study, we found no significant difference between the BTT and tBUT results among all participants, and the BTT was shorter in patients with dry eye than in normal subjects. Because the AUC of the BTT was 0.628, it was insufficient to diagnose dry eye disease alone, but could be used as a self-test for screening purposes. Second, the BTT test was superior to the tBUT test with respect to the reproducibility of the results. In our study, the ICC of the BTT test was 0.904 and that of the tBUT test was 0.679; the BTT test was therefore more reproducible than the tBUT test. Third, the BTT test better reflects the physiologic status of the tear film, because no fluorescein dye is needed for the testing procedure.

The BTT test has some weaknesses. First, the BTT is affected not only by tear film stability, but also by each individual pain threshold; a person with a high pain threshold will have a longer BTT and vice versa. As such, we expected that the OSDI, which reflects the subjective symptoms of patients with dry eye, would be significantly correlated with the BTT, but instead, the results of BTT test showed a significant positive correlation with the results tBUT test. Second, the BTT test examines both eyes simultaneously, so that the condition of each eye cannot be measured separately, unlike in the tBUT or Schirmer tests (both of which test 1 eye at a time). Of course, we can measure the BTT in 1 eye at a time, with the other eye closed, but this method would not reflect the physiologic status of the tear film.

Conclusions

The tBUTBE and BTT values among all participants did not differ significantly, and a significant positive correlation was detected between the 2 measurements. In addition, the ICC of the BTT test was superior to that of the tBUT test, and the AUC of the BTT test was not significantly different from that of the TBUT test. Most importantly, the BTT values were significantly shorter in dry eye patients than in normal subjects. These results suggest that the BTT test can be used as an adjunct to existing tests for evaluating dry eye disease as well as for self-diagnosis of the disorder.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

OSDI:

Ocular Surface Disease Index

BTT:

Blinking tolerance time

tBUT:

Tar break-up time

ICC:

Intraclass correlation coefficient

CIs:

Confidence intervals

SD:

Standard deviation

ROC:

Receiver operating characteristic

AUC:

Area under the curve

IBT:

Inter-blinking time

References

  1. Bron AJ, de Paiva CS, Chauhan SK, Bonini S, Gabison EE, Jain S, et al. TFOS DEWS II pathophysiology report. Ocul Surf. 2017;15(3):438–510. https://doi.org/10.1016/j.jtos.2017.05.011 Epub 2017/07/25. PubMed PMID: 28736340.

    Article  Google Scholar 

  2. Norn MS. Desiccation of the precorneal film. I. Corneal wetting-time. Acta Ophthalmol. 1969;47(4):865–80 PubMed PMID: 4187469.

    Article  CAS  Google Scholar 

  3. Nichols KK, Mitchell GL, Zadnik K. The repeatability of clinical measurements of dry eye. Cornea. 2004;23(3):272–85 PubMed PMID: 15084861.

    Article  Google Scholar 

  4. Markoulli M, Duong TB, Lin M, Papas E. Imaging the tear film: a comparison between the subjective Keeler Tearscope-plus and the objective oculus(R) Keratograph 5M and LipiView(R) interferometer. Curr Eye Res. 2018;43(2):155–62. https://doi.org/10.1080/02713683.2017.1393092 Epub 2017/11/15. PubMed PMID: 29135324.

    Article  Google Scholar 

  5. Schiffman RM, Christianson MD, Jacobsen G, Hirsch JD, Reis BL. Reliability and validity of the ocular surface disease index. Arch Ophthalmol. 2000;118(5):615–21 PubMed PMID: 10815152.

    Article  CAS  Google Scholar 

  6. Bron AJ, Evans VE, Smith JA. Grading of corneal and conjunctival staining in the context of other dry eye tests. Cornea. 2003;22(7):640–50 PubMed PMID: 14508260.

    Article  Google Scholar 

  7. Simpson TL, Situ P, Jones LW, Fonn D. Dry eye symptoms assessed by four questionnaires. Optom Vis Sci. 2008;85(8):692–9. https://doi.org/10.1097/OPX.0b013e318181ae36 PubMed PMID: 18677225.

    Article  Google Scholar 

  8. DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics. 1988;44(3):837–45 Epub 1988/09/01. PubMed PMID: 3203132.

    Article  CAS  Google Scholar 

  9. Herrero-Vanrell R, Peral A. International dry eye workshop (DEWS). Update of the disease. Arch Soc Esp Oftalmol. 2007;82(12):733–4 Epub 2007/11/28. PubMed PMID: 18040913.

    Article  CAS  Google Scholar 

  10. Pult H, Riede-Pult BH, Murphy PJ. The relation between blinking and conjunctival folds and dry eye symptoms. Optom Vis Sci. 2013;90(10):1034–9. https://doi.org/10.1097/OPX.0000000000000029 Epub 2013/09/27. PubMed PMID: 24067407.

    Article  Google Scholar 

  11. Gumus K, Crockett CH, Pflugfelder SC. Anterior segment optical coherence tomography: a diagnostic instrument for conjunctivochalasis. Am J Ophthalmol. 2010;150(6):798–806. https://doi.org/10.1016/j.ajo.2010.06.014 PubMed PMID: 20869039; PubMed Central PMCID: PMC3577926.

    Article  Google Scholar 

  12. Rahman EZ, Lam PK, Chu CK, Moore Q, Pflugfelder SC. Corneal sensitivity in tear dysfunction and its correlation with clinical parameters and blink rate. Am J Ophthalmol 2015;160(5):858–866 e5. doi: https://doi.org/10.1016/j.ajo.2015.08.005. PubMed PMID: 26255576; PubMed Central PMCID: PMC4661092.

  13. Efron N, Young G, Brennan NA. Ocular surface temperature. Curr Eye Res. 1989;8(9):901–6 PubMed PMID: 2791634.

    CAS  Google Scholar 

  14. Craig JP, Singh I, Tomlinson A, Morgan PB, Efron N. The role of tear physiology in ocular surface temperature. Eye. 2000;14(Pt 4):635–41. https://doi.org/10.1038/eye.2000.156 PubMed PMID: 11040913.

    Article  Google Scholar 

  15. Mathers WD, Binarao G, Petroll M. Ocular water evaporation and the dry eye. A new measuring device. Cornea. 1993;12(4):335–40 PubMed PMID: 8339563.

    Article  CAS  Google Scholar 

  16. Purslow C, Wolffsohn J. The relation between physical properties of the anterior eye and ocular surface temperature. Optom Vis Sci. 2007;84(3):197–201. https://doi.org/10.1097/OPX.0b013e3180339f6e PubMed PMID: 17435533.

    Article  Google Scholar 

  17. Belmonte C, Gallar J. Cold thermoreceptors, unexpected players in tear production and ocular dryness sensations. Invest Ophthalmol Vis Sci. 2011;52(6):3888–92. https://doi.org/10.1167/iovs.09-5119 PubMed PMID: 21632706.

    Article  Google Scholar 

Download references

Acknowledgments

We would like to appreciate department of medical statistics, the Catholic university of Korea for assistance with statistical analysis.

Funding

This research was supported by a grant from the Korea Health Technology R&D Project through the Korea Health Industry Development Institute, funded by the Ministry of Health & Welfare, Republic of Korea (grant number: HI17C0659), the Basic Science Research Program through the National Research Foundation of Korea, funded by the Ministry of Education, Republic of Korea (grant number: 2017R1A1A2A10000681), and National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT & Future Planning (2017R1C1B5017408). This research was also supported by a Grant of Translational R&D Project through the Institute for Bio-Medical Convergence, Incheon St. Mary’s Hospital, the Catholic University of Korea.

Author information

Authors and Affiliations

Authors

Contributions

HSH designed the study. YHK and ECK collected the data. HBH and HSK analyzed the data, and HBH and HSK drafted the manuscript. HBH, YHK, HYP, ECK, HSK and MSK was all involved in the interpretation of the results and was major contributions to the writing of the manuscript. All authors approved the final version of the paper for submission.

Corresponding author

Correspondence to Ho Sik Hwang.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the institutional review board of Chuncheon Sacred Hospital and Incheon St. Mary’s Hospital. It also strictly adhered to the principles of the Declaration of Helsinki. And all participants provided written consent to participate in this study.

Consent for publication

Not applicable.

Competing interests

The authors have no proprietary or commercial interest in any of the materials discussed in this study. No author has a financial or proprietary interest in any material or method mentioned. The authors alone are responsible for the content and writing of the paper. Also, as a member of the editorial board (Ho Sik Hwang) of BMC Ophthalmology, in order to transparency, all the authors want to strictly follow the review process of this journal.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hwang, H., Ku, Y., Kim, E. et al. Easy and effective test to evaluate tear-film stability for self-diagnosis of dry eye syndrome: blinking tolerance time (BTT). BMC Ophthalmol 20, 438 (2020). https://doi.org/10.1186/s12886-020-01686-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12886-020-01686-5

Keywords